HUDSON RCI IPN048872Recalls & reported problems
Teleflex Incorporated · ET TUBE,CF,7.0 · Class 2
All models of HUDSON RCI (152)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 14026704616517), received between Aug 26, 2021 and Jun 27, 2023 — classified as 1 death, 9 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (3 reports).
27 other recalls cover devices cleared under the same FDA submission (K822082); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Deflation Problem3
- Connection Problem2
- Gas/Air Leak2
- Deformation Due to Compressive Stress1
- Inflation Problem1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Insufficient Information2
- Cardiac Arrest1
- Respiratory Arrest1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 27, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT "THE BALLOON DID NOT INFLATE PROPERLY". AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR MULTIPLE REQUESTS FOR ADDITIONAL INFORMATION. IF FURTHER INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.
Patient problems: Insufficient Information
MAUDE report 17213209 ↗MalfunctionApr 3, 2023Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT"THE DOCTOR FOUND CUFF LEAKAGE WHEN USING ON PATIENT. THEN CHANGED NEW ONE, NO IMPACT ON PATIENT. INVOLVED 1 PC".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16665654 ↗MalfunctionJan 20, 2023Connection Problem; Connection Problem
REPORTED EVENT: THE PATIENT'S ET TUBE BECAME DISCONNECTED OR CAME APART WHERE THE TUBE CONNECTS TO THE 15 MM CONNECTOR WHILE AN MRI WAS BEING PERFORMED. THE PATIENT WAS EXTUBATED AND AFTER A FEW MORE DAYS IN THE HOSPITAL WAS DISCHARGED TO HOME. (ASSOCIATED REPORT #3003898360-2023-00015).
Patient problems: Insufficient Information
MAUDE report 16215416 ↗DeathJan 20, 2023Connection Problem; Connection Problem
REPORTED EVENT: ALL OF THE FOLLOWING WAS REPORTED BY THE CUSTOMER REPRESENTATIVE. DURING USE, A PATIENT'S ET TUBE BECAME DISCONNECTED WHERE THE TUBE CONNECTS TO THE 15 MM CONNECTOR. THE PATIENT CODED. THE PATIENT HAD A HISTORY OF COPD AND CHF AND PRESENTED IN THE EMERGENCY DEPARTMENT VIA EMS ON (B)(6) DUE TO SHORTNESS OF BREATH. THE PATIENT DID NOT HAVE AN ET TUBE ON ARRIVAL TO THE EMERGENCY DEPARTMENT. THE ET TUBE WAS FOUND DISCONNECTED AT THE CONNECTION POINT ON THE ET TUBE AT APPROXIMATELY 1354. AT THIS POINT THE PATIENT WAS MOVING HIS HEAD BACK AND FORTH AND THE OXYGEN SATURATION HAD DROPPED. THE VENTILATOR WAS ALARMING. THE ICU TEAM ARRIVED AT BEDSIDE AND THE PATIENT DEVELOPED CARDIAC ARREST WITH PEA. AS PER ACLS PROTOCOL, THE PATIENT RECEIVED TWO ROUNDS OF CPR AND TWO ROUNDS OF EPINEPHRINE AND SUBSEQUENTLY ACHIEVED ROSC. THE RHYTHM WAS NOTED TO BE ATRIAL FIBRILLATION WITH HEART RATE BETWEEN 90-100 BEATS/MIN. THE PATIENT SUBSEQUENTLY DEVELOPED SHOCK FOR WHICH THE PATIENT WAS ADMINISTERED PRESSOR SUPPORT WITH LEVOPHED, VASOPRESSIN AND STRESS DOSAGE STEROIDS. THE FAMILY WAS UPDATED AT THE BEDSIDE AND THE CODE STATUS UPDATED TO NO CPR. THE PATIENT WAS TERMINALLY EXTUBATED ON (B)(
Patient problems: Cardiac Arrest; Respiratory Arrest
MAUDE report 16215362 ↗MalfunctionDec 20, 2022Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
IT WAS REPORTED THAT THE DOCTOR FOUND THAT THE CUFF WAS DEFORMED AND CAUSED OCCLUSION WHEN DOING CLINICALS BEFORE USING ON A PATIENT. A NEW DEVICE WAS USED AND THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16015429 ↗MalfunctionNov 4, 2022Leak/Splash; Leak/Splash
THE INFORMATION STATES THAT THE CLINICIAN TESTED THE CUFF PRIOR TO INTUBATION, AND IT INFLATED NORMALLY. ONCE THEY INTUBATED THE PATIENT, THE CUFF DEFLATED. THE CLINICIAN HAD TO ADD 60-70MMHG TO INFLATE THE CUFF WHICH COULD POTENTIALLY HARM THE PATIENT. THE CUSTOMER CONFIRMED "THE PATIENT WAS DOING PERFECTLY FINE, AND THE DEVICE DID NOT HARM THE PATIENT IN ANY WAY". THE PATIENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15736734 ↗MalfunctionOct 15, 2021Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT: "THE CUFF DEFLATED WHEN PATIENTS WERE POSITIONED IN PRONE POSITION. NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12638964 ↗MalfunctionSep 23, 2021Deflation Problem; Deflation Problem
IT WAS REPORTED THAT: "DURING THE USE OF THESE ET TUBES, THE USERS OBSERVED LEAKS, REQUIRING AN "OVER-INFLATION" OF THE CUFF OVER 30CM H2O. MONITERED WITH THE MANOMETER". NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12519040 ↗MalfunctionSep 23, 2021Deflation Problem; Deflation Problem
IT WAS REPORTED THAT: "DURING THE USE OF THESE ET TUBES, THE USERS OBSERVED LEAKS, REQUIRING AN "OVER-INFLATION" OF THE CUFF OVER 30CM H2O. MONITERED WITH THE MANOMETER". NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12518903 ↗MalfunctionAug 26, 2021Deflation Problem; Deflation Problem
IT WAS REPORTED "THE STAFF REPORTED THAT THE CUFF OF THE ET TUBE DEFLATED DURING USE". NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12373834 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K822082. One submission often covers many models, so check the product and lot details in each record.
- Z-2313-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 mm, REF 5-22515 e) 8.0 mm, REF 5-22516 f) 8.5 mm, REF 5-22517 g) 9.0
- Z-2316-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REF 5-10313 e) 7.0 mm, REF 5-10314 f) 7.5 mm, REF 5-10315 g) 8.0 mm,
- Z-2312-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan/CF: a) 3.0 mm, REF 5-10106, REF 5-10206 b) 3.5 mm, REF 5-10107 c) 4.0 mm, REF 5-10108, REF 5-10208 d) 4.5 mm, REF 5-10109 e) 5.0 mm, REF 5-10110, REF 5-10
- Z-2311-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Tracheal tube/airway management
- Z-2315-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan Uncuffed: a) 2.0 mm, REF 5-10404 b) 2.5 mm, REF 5-10405 c) 3.0 mm, REF 5-10406 d) 3.5 mm, REF 5-10407 e) 4.0 mm, REF 5-10408 f) 4.5 mm, REF 5-10409 g)
- Z-2314-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan Preformed: a) 3.0 mm, REF 5-22006, REF 5-22106 b) 3.5 mm, REF 5-22007, REF 5-22107 c) 4.0 mm, REF 5-22008, REF 5-22108, REF 5-22208 d) 4.5 mm, REF 5-22009,
- Z-2317-2019 · Jul 30, 2019 · TerminatedSheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management
- Z-2318-2019 · Jul 30, 2019 · TerminatedSheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10317 Product Usage: Tracheal tube/airway management
- Z-1956-2019 · Jun 7, 2019 · TerminatedSheridan Endotracheal Tube contained inside Centurion kit code TC7855.
- Z-1764-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) HVT 6.0 mm, Product Code 5-10312 2) HVT 6.5 mm, Product Code 5-10313 3) HVT 7.0 mm, Product Code 5-10314 4) HVT 7.5 mm, Product Code 5-10315 5) HVT 8.0
- Z-1760-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) CF 6.0 mm, Product Codes: 5-10112, 5-10212 2) CF 6.5 mm, Product Codes: 5-10113, 5-10213 3) CF 7.0 mm, Product Codes: 5-10114, 5-10214 4) CF 7.5 mm,
- Z-1763-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6.5 mm, Product Code 5-10413 3) Uncuffed 7.0 mm, Product Code 5-10414 Product Usage: Tracheal tube/
- Z-1765-2019 · May 24, 2019 · TerminatedSheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Tracheal tube/airway management
- Z-1766-2019 · May 24, 2019 · TerminatedSheridan/HVT: 1) Novaplus 7.0 mm, Product Code V5-10314 2) Novaplus 7.5 mm, Product Code V5-10315 3) Novaplus 8.0 mm, Product Code V5-10316 4) Novaplus 8.5 mm, Product Code V5-
- Z-1761-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 mm, Product Code 5-22514 4) EZ-ENDO 7.5 mm, Product Code 5-2
…and 12 more.
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page