HUDSON RCI IPN048873Recalls & reported problems
Teleflex Incorporated · ET TUBE,CF,7.5 · Class 2
All models of HUDSON RCI (152)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Hudson RCI endotracheal tube is a breathing tube placed in the windpipe. The 25 reports mostly describe air leaking from the cuff or the cuff not inflating, with the tube replaced and no patient impact mentioned. There are no deaths and no injuries in the reports. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 14026704616524), received between Feb 8, 2021 and Apr 12, 2023 — classified as 25 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (16 reports).
27 other recalls cover devices cleared under the same FDA submission (K822082); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Gas/Air Leak16
- Deflation Problem4
- Break1
- Burst Container or Vessel1
- Fluid/Blood Leak1
- Inflation Problem1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- Insufficient Information4
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 12, 2023Fluid/Blood Leak; Fluid/Blood Leak
REPORTED ISSUE: THE DOCTOR FOUND CUFF LEAKAGE WHEN USING IT ON THE PATIENT. THE DEVICE WAS REPLACED WITH A NEW ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16726442 ↗MalfunctionApr 3, 2023Gas/Air Leak; Gas/Air Leak
REPORTED ISSUE: THE DOCTOR FOUND CUFF LEAKAGE BEFORE USING THE DEVICE ON A PATIENT. THIS OCCURRED IN A CLINICAL SETTING. THE DEVICE WAS REPLACED WITH A NEW ONE AND THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16666389 ↗MalfunctionMar 27, 2023Break; Break
REPORTED ISSUE: THE CUFF WOULD NOT INFLATE. THE CUSTOMER WAS UNABLE TO ANSWER MULTIPLE INQUIRIES REGARDING IF THE EVENT HAPPENED WHILE IN USE ON A PATIENT, HOW LONG THE DEVICE WAS IN USE, IF MEDICAL INTERVENTION WAS REQUIRED, IF REINTUBATION WAS REQUIRED, IF DESATURATION OCCURRED, AND THE CURRENT HEALTH STATUS OF THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16622558 ↗MalfunctionMar 15, 2023Gas/Air Leak; Gas/Air Leak
REPORTED ISSUE: THE DOCTOR FOUND THE CUFF LEAKAGE WHEN USING IT ON A PATIENT. A NEW TUBE WAS USED AND THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16550897 ↗MalfunctionFeb 14, 2023Gas/Air Leak; Gas/Air Leak
REPORTED EVENT: THE DOCTOR FOUND CUFF LEAKAGE WHEN USING IT ON THE PATIENT. IT WAS CHANGED TO A NEW ONE. NO IMPACT ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16374255 ↗MalfunctionFeb 14, 2023Gas/Air Leak; Gas/Air Leak
REPORTED EVENT: THE DOCTOR FOUND CUFF LEAKAGE WHEN USING IT ON THE PATIENT. IT WAS CHANGED TO A NEW ONE. NO IMPACT ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16374200 ↗MalfunctionFeb 14, 2023Gas/Air Leak; Gas/Air Leak
REPORTED EVENT: THE DOCTOR FOUND CUFF LEAKAGE WHEN USING IT ON THE PATIENT. IT WAS CHANGED TO A NEW ONE. NO IMPACT ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16374038 ↗MalfunctionFeb 14, 2023Gas/Air Leak; Gas/Air Leak
REPORTED EVENT: THE DOCTOR FOUND CUFF LEAKAGE WHEN USING IT ON THE PATIENT. IT WAS CHANGED TO A NEW ONE. NO IMPACT ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16373922 ↗MalfunctionJan 9, 2023Deflation Problem; Deflation Problem
REPORTED ISSUE: THE ED PHYSICIAN INTUBATED A PATIENT AND HAD DIFFICULTY WITH THE TUBE. RESPIRATORY CHECKED THE BALLOON PRIOR TO INTUBATION AND IT INFLATED PROPERLY. AFTER INTRODUCING THE TUBE, IT WOULD NOT HOLD AIR. THIS HAPPENED ON 2 DIFFERENT ET TUBES, 3RD HELD AIR. THE TUBE IS A 7.5MM HUDSON RCI TUBE, LOT 73D220280 WITH A 04/10/2027 EXPIRATION. THIS SAME SITUATION HAPPENED WITH TWO OTHER 8.0 TUBE ET TUBES. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT.
Patient problems: Insufficient Information
MAUDE report 16125076 ↗MalfunctionJan 9, 2023Deflation Problem; Deflation Problem
REPORTED ISSUE: THE ED PHYSICIAN INTUBATED A PATIENT AND HAD DIFFICULTY WITH THE TUBE. RESPIRATORY CHECKED THE BALLOON PRIOR TO INTUBATION AND IT INFLATED PROPERLY. AFTER INTRODUCING THE TUBE, IT WOULD NOT HOLD AIR. THIS HAPPENED ON 2 DIFFERENT ET TUBES, 3RD HELD AIR. THE TUBE IS A 7.5MM HUDSON RCI TUBE, LOT 73D220280 WITH A 04/10/2027 EXPIRATION. THIS SAME SITUATION HAPPENED WITH TWO OTHER 8.0 TUBE ET TUBES. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT.
Patient problems: Insufficient Information
MAUDE report 16123796 ↗MalfunctionDec 22, 2022Deflation Problem; Deflation Problem
IT WAS REPORTED THAT THE DOCTOR FOUND THAT THE CUFF CANNOT DEFLATE WHEN USING IT ON A PATIENT. A NEW ONE WAS USED AND THERE WAS NO DESATURATION OR OTHER PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16038044 ↗MalfunctionNov 17, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE DOCTOR FOUND CUFF LEAKAGE WHEN USING ON PATIENT. IT WAS CHANGED TO A NEW ONE. THERE WAS NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15814583 ↗MalfunctionNov 2, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED ON (B)(6)2022 THAT THE DOCTOR FOUND THE CUFF LEAKING WHILE USING IT ON A PATIENT. THE DEVICE WAS CHANGED TO A NEW ONE. THERE WAS NO IMPACT ON THE PATIENT. NO MEDICAL INTERVENTION WAS REPORTED. THE CURRENT PATIENT'S CONDITION WAS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15717632 ↗MalfunctionOct 13, 2022Material Rupture; Material Rupture
THE REPORTED COMPLAINT STATES THAT: "THE DOCTOR FOUND CUFF RUPTURED WHEN USING ON PATIENT. THEN CHANGED TO A NEW ONE. NO IMPACT TO PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15596335 ↗MalfunctionOct 7, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT: "THE DOCTOR FOUND CUFF RUPTURED WHEN USING ON PATIENT. THEN CHANGED TO A NEW ONE. NO IMPACT ON PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15563146 ↗MalfunctionMar 17, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT "THE DOCTOR FOUND CUFF LEAKAGE WHEN USING ON PATIENT AND CHANGED NEW ONE". NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13794678 ↗MalfunctionMar 17, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT "THE DOCTOR FOUND CUFF LEAKAGE WHEN USING ON PATIENT AND CHANGED NEW ONE". NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13794624 ↗MalfunctionDec 23, 2021Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT "THE DOCTOR FOUND CUFF LEAKAGE WHEN DOING CLINICAL SETTING BEFORE USING ON PATIENT". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13080397 ↗MalfunctionDec 17, 2021Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT: "THE CUFF IS LEAKING. THE CUFF'S WALLS STICK TOGETHER AFTER INTUBATION". UNKNOWN WHEN ISSUE WAS FIRST IDENTIFIED. PATIENT CONDITION UNKNOWN AT TIME OF REPORT. ATTEMPTS TO CUSTOMER FOR ADDITIONAL INFORMATION HAS BEEN UNSUCCESSFUL.
Patient problems: Insufficient Information
MAUDE report 13026239 ↗MalfunctionSep 14, 2021Inflation Problem; Inflation Problem
IT WAS REPORTED THAT "THE CUFF DID NOT INFLATE DURING A LEAK TEST BEFORE USE ON A PATIENT. THEREFORE, A NEW UNIT WAS OPENED INSTEAD". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12470621 ↗MalfunctionAug 26, 2021Deflation Problem; Deflation Problem
IT WAS REPORTED THAT "ON (B)(6) 2021, THE ANAESTHESISTS OBSERVED THAT THE CUFF OF THE ET TUBE DEFLATES. IT IS IMPOSSIBLE TO MAINTAIN A CONSTANT VOLUME AND WHEN CHECKING WITH THE MANOMETER IT SHOWED A DECREASE IN PRESSURE. CLINICAL CONSEQUENCES: THE ET TUBE WAS REPLACED WITHOUT ANY CONSEQUENCES FOR THE PATIENT". NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12373209 ↗MalfunctionMar 22, 2021Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED "THE DOCTOR FOUND CUFF LEAKAGE IN THE CLINICAL SETTING BEFORE USING ON PATIENT, NO IMPACT ON THE OPERATION". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11545960 ↗MalfunctionMar 22, 2021Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED "THE DOCTOR FOUND CUFF LEAKAGE IN THE CLINICAL SETTING BEFORE USING ON PATIENT, NO IMPACT ON THE OPERATION". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11545949 ↗MalfunctionFeb 11, 2021Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT "A DISPOSABLE TRACHEAL TUBE WAS USED FOR THE TREATMENT OF THE PATIENT. DURING THE USE, IT WAS FOUND THAT THE BALLOON WAS LEAKING. THEN REPLACED NEW ONE WHICH HAS NO EFFECT ON THE PATIENT". NO PATIENT HARM OR INJURY REPORTED. NO DESATURATION REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11316209 ↗MalfunctionFeb 8, 2021Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT "DOCTOR FOUND THE CUFF LEAKAGE DURING FUNCTION TEST PRIOR TO THE PATIENT". DEVICE REPLACED. NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11298235 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K822082. One submission often covers many models, so check the product and lot details in each record.
- Z-2311-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Tracheal tube/airway management
- Z-2317-2019 · Jul 30, 2019 · TerminatedSheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management
- Z-2316-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REF 5-10313 e) 7.0 mm, REF 5-10314 f) 7.5 mm, REF 5-10315 g) 8.0 mm,
- Z-2318-2019 · Jul 30, 2019 · TerminatedSheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10317 Product Usage: Tracheal tube/airway management
- Z-2313-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 mm, REF 5-22515 e) 8.0 mm, REF 5-22516 f) 8.5 mm, REF 5-22517 g) 9.0
- Z-2312-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan/CF: a) 3.0 mm, REF 5-10106, REF 5-10206 b) 3.5 mm, REF 5-10107 c) 4.0 mm, REF 5-10108, REF 5-10208 d) 4.5 mm, REF 5-10109 e) 5.0 mm, REF 5-10110, REF 5-10
- Z-2314-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan Preformed: a) 3.0 mm, REF 5-22006, REF 5-22106 b) 3.5 mm, REF 5-22007, REF 5-22107 c) 4.0 mm, REF 5-22008, REF 5-22108, REF 5-22208 d) 4.5 mm, REF 5-22009,
- Z-2315-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan Uncuffed: a) 2.0 mm, REF 5-10404 b) 2.5 mm, REF 5-10405 c) 3.0 mm, REF 5-10406 d) 3.5 mm, REF 5-10407 e) 4.0 mm, REF 5-10408 f) 4.5 mm, REF 5-10409 g)
- Z-1956-2019 · Jun 7, 2019 · TerminatedSheridan Endotracheal Tube contained inside Centurion kit code TC7855.
- Z-1760-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) CF 6.0 mm, Product Codes: 5-10112, 5-10212 2) CF 6.5 mm, Product Codes: 5-10113, 5-10213 3) CF 7.0 mm, Product Codes: 5-10114, 5-10214 4) CF 7.5 mm,
- Z-1759-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management
- Z-1763-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6.5 mm, Product Code 5-10413 3) Uncuffed 7.0 mm, Product Code 5-10414 Product Usage: Tracheal tube/
- Z-1761-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 mm, Product Code 5-22514 4) EZ-ENDO 7.5 mm, Product Code 5-2
- Z-1766-2019 · May 24, 2019 · TerminatedSheridan/HVT: 1) Novaplus 7.0 mm, Product Code V5-10314 2) Novaplus 7.5 mm, Product Code V5-10315 3) Novaplus 8.0 mm, Product Code V5-10316 4) Novaplus 8.5 mm, Product Code V5-
- Z-1765-2019 · May 24, 2019 · TerminatedSheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Tracheal tube/airway management
…and 12 more.
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page