HUDSON RCI IPN048896Recalls & reported problems
Teleflex Incorporated · ET TUBE, HVT, 7.5 · Class 2
All models of HUDSON RCI (152)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Hudson RCI endotracheal tube is a breathing tube used to secure the airway during intubation. The 32 reports mostly describe the cuff or hub coming loose or not holding air properly, so the tube disconnected, leaked air, or deflated on its own — in some cases the patient had to be extubated and reintubated. Reports also mention low oxygen levels and 2 deaths; no injuries were reported. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 14026704616753), received between May 17, 2021 and Jul 31, 2023 — classified as 2 deaths, 30 malfunctions.
The most frequently reported specific device problem is “Connection Problem” (8 reports).
27 other recalls cover devices cleared under the same FDA submission (K822082); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Connection Problem8
- Deflation Problem6
- Material Rupture6
- Inflation Problem4
- Gas/Air Leak2
- Leak/Splash2
- Break1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions20
- Insufficient Information10
- Low Oxygen Saturation3
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 31, 2023
A POWDER LIKE SUBSTANCE WAS FOUND AT THE TOP OF THE TUBE AFTER OPENING THE STERILE PACKAGE & PRIOR TO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17423685 ↗MalfunctionApr 18, 2023Connection Problem; Connection Problem
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HUB DISCONNECTED. AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. THE COMPLAINT FILE WILL BE UPDATED IF FURTHER INFORMATION IS RECEIVED.
Patient problems: Insufficient Information
MAUDE report 16763570 ↗MalfunctionMar 24, 2023Leak/Splash; Leak/Splash
REPORTED ISSUE: THE TUBE PASSED THE PRE-TEST, BUT AIR LEAKING ALARM OCCURRED AFTER INTUBATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16616167 ↗MalfunctionMar 14, 2023Leak/Splash; Leak/Splash
REPORTED ISSUE: THE CUFF DEFLATED AND THE PATIENT HAD TO BE EXTUBATED AND THEN REINTUBATED. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16539685 ↗MalfunctionMar 14, 2023Fluid/Blood Leak; Fluid/Blood Leak
REPORTED ISSUE: THE CUFF DEFLATED AND THE PATIENT HAD TO BE EXTUBATED AND THEN REINTUBATED. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16539649 ↗MalfunctionMar 6, 2023Connection Problem; Connection Problem
REPORTED ISSUE: THE ET TUBE DISCONNECTS.
Patient problems: Insufficient Information
MAUDE report 16487409 ↗MalfunctionMar 6, 2023Connection Problem; Connection Problem
REPORTED ISSUE: THE ET TUBE DISCONNECTS.
Patient problems: Insufficient Information
MAUDE report 16487361 ↗MalfunctionMar 6, 2023Connection Problem; Connection Problem
REPORTED ISSUE: ET TUBE DISCONNECTS.
Patient problems: Insufficient Information
MAUDE report 16487329 ↗MalfunctionMar 6, 2023Connection Problem; Connection Problem
REPORTED ISSUE: THE ET TUBE CUFF DEFLATES.
Patient problems: Insufficient Information
MAUDE report 16487170 ↗MalfunctionMar 6, 2023Failure to Disconnect; Failure to Disconnect
REPORTED ISSUE: ET TUBE DISCONNECTS.
Patient problems: Insufficient Information
MAUDE report 16486879 ↗MalfunctionMar 6, 2023Deflation Problem; Deflation Problem
REPORTED ISSUE: THE CUFF DEFLATED. THE PATIENT WAS REINTUBATED; THERE WAS NO DESATURATION OR INJURY.
Patient problems: Insufficient Information
MAUDE report 16486765 ↗MalfunctionFeb 27, 2023Inflation Problem; Inflation Problem
REPORTED ISSUE: THE ENDOTRACHEAL TUBE CUFF WOULD NOT HOLD PRESSURE. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16447130 ↗MalfunctionFeb 2, 2023Connection Problem; Connection Problem
REPORTED EVENT: CONNECTOR OF ETT COULD NOT STAY ON, KEPT POPPING OFF. SO IT HAD TO BE REMOVED AND THE PATIENT WAS REINTUBATED WITH ANOTHER ETT. THIS OCCURRED WITH 2 DIFFERENT PATIENTS ON THE SAME DAY, SAME PROBLEM, SAME SOLUTION TO REINTUBATE. NO INJURY REPORTED. (ASSOCIATED COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16283502 ↗MalfunctionFeb 2, 2023Connection Problem; Connection Problem
REPORTED EVENT: CONNECTOR OF ETT COULD NOT STAY ON, KEPT POPPING OFF. SO IT HAD TO BE REMOVED AND THE PATIENT WAS REINTUBATED WITH ANOTHER ETT. THIS OCCURRED WITH 2 DIFFERENT PATIENTS ON THE SAME DAY, SAME PROBLEM, SAME SOLUTION TO REINTUBATE. NO INJURY REPORTED. (ASSOCIATED COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16283463 ↗MalfunctionJan 4, 2023Connection Problem; Connection Problem
REPORTED ISSUE: THE PATIENT WAS INTUBATED WITH TELEFLEX SHERIDAN 7.5 ETT ON 16 DECEMBER. THE TUBE'S HUB CONNECTOR WAS GETTING DISCONNECTED FROM THE ETT AND HAD TO BE REPLACED. THIS WAS REPORTED BY THE UNIT RESPIRATORY THERAPISTS ALONG WITH SIMILAR CONCERNS WITH THE TELEFLEX SHERIDAN ETTS IN THE PAST. NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16096473 ↗MalfunctionDec 14, 2022Inflation Problem; Inflation Problem
REPORTED ISSUE: PRIOR TO INTUBATING THE PATIENT THEY NOTICED THERE WAS A LEAK IN THE CUFF AND WERE ABLE TO GRAB A DIFFERENT TUBE. THE REPORTED DEFECT WAS DETECTED PRIOR TO USE IN A CLINICAL SETTING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15982086 ↗MalfunctionDec 12, 2022Inflation Problem; Inflation Problem
REPORTED ISSUE: DURING CASE WITH PATIENT INTUBATED THEY NOTICED THAT THERE WAS A LEAK IN THE CUFF AND THEY NEEDED TO CHANGE IT OUT FOR ANOTHER TUBE. THERE WAS NO REPORTED INJURY AND THE PATIENT'S CONDITION WAS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15960178 ↗MalfunctionFeb 7, 2022Break; Break
IT WAS REPORTED "THE ISSUE IS THAT THE CONNECTOR FROM OUR STANDARD TUBE IS UNCOUPLING ON ITS OWN". NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION UNKNOWN AT TIME OF REPORT. SPECIFIC INFORMATION REQUESTED TO THE CUSTOMER REGARDING THE EVENT HAS BEEN UNSUCCESSFUL.
Patient problems: Insufficient Information
MAUDE report 13475711 ↗MalfunctionJan 28, 2022Disconnection; Disconnection
IT WAS REPORTED THE CONNECTOR IS UNCOUPLING ON ITS OWN. NO REPORT OF PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION UNKNOWN AT TIME OF REPORT.
Patient problems: Insufficient Information
MAUDE report 13385141 ↗MalfunctionNov 11, 2021Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED "WE JUST HAD ANOTHER ETT CUFF FAIL. IT APPEARS TO HAVE LEAKED AND WAS NOT ABLE TO HOLD AIR". PATIENT CONDITION REPORTED AS "FINE". MULTIPLE REQUESTS FOR ADDITIONAL INFORMATION TO THE CUSTOMER HAS BEEN UNSUCCESSFUL.
Patient problems: Insufficient Information
MAUDE report 12800394 ↗MalfunctionOct 28, 2021Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED "THE NURSE JUST HEARD AIR COMING FROM PATIENT, RT ASSESSED, NOTIFIED THE PROVIDER, TUBE WAS EXCHANGED OUT DUE TO CUFF LEAK". NO PATIENT HARM OR INJURY REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12712358 ↗MalfunctionOct 18, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT THE ET TUBES HAD RUPTURED CUFFS AND REQUIRED TO BE EXCHANGED. NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12647232 ↗MalfunctionOct 18, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT THE ET TUBES HAD RUPTURED CUFFS AND REQUIRED TO BE EXCHANGED. NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12647128 ↗MalfunctionOct 18, 2021Material Rupture; Material Rupture
IT WAS REPORTED "PT INTUBATED BY DR. PANCHAL USING GLIDESCOPE AND 8.0 ETT PLACED SECURED AT 26CM LIP. ETT SOUNDED LIKE THERE WAS A LEAK- ADDED 1CC OF AIR, DID NOT CHANGE LEAK. ADVANCED TUBE TO 28 CM AT LIP PER DR. PANCHAL. PT ACHIEVING VOLUMES ON VENT, PT AND SPO2 IMPRICING. LARGE AIR LEAK, VENT LOSING VOLUMES AND PT DESATURATING. BRONCHOSCOPE OBTAINED FOR QUICK LOOK. BOUGIE PLACED AND ETT EXCHANGED 7.5 PLACED. BRONCHOSCOPE INSERTED. ETT SECURED AT 23CM LIP. OLD ETT INSPECTED AND CUFF BLOWN". NO PATIENT INJURY OR HARM REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12646713 ↗DeathOct 18, 2021Material Rupture; Material Rupture
IT WAS REPORTED "PATIENT WAS TRANSFERRED EMERGENTLY TO THE ICU VIA RAPID RESPONSE TEAM AND EMERGENTLY INTUBATED DUE TO SPO2 60% DESPITE MAX BIPAP THERAPY. UPON INTUBATION THE ETT PLACEMENT WAS CONFIRMED WITH BILATERAL AUSCULTATION AND RADIOGRAPH, LOW TIDAL VOLUMES AND EXPIRATORY VOLUMES WERE NOTED WITH SIGNIFICANT CUFF LEAK. CUFF WAS DETERMINED TO BE BLOWN. ETT INSERTED WAS TELEFLEX HUDSON RCI SHERIDAN HVT 7.5MM. PATIENT SUBSEQUENTLY RESPIRATORY ARRESTED REQUIRING ACLS MEASURES AND EXCHANGE OF ETT BY MD WITH ROSC. HOWEVER, SHORTLY AFTER PATIENT SUBSEQUENTLY ARRESTED AGAIN AND DESPITE MEDICAL INTERVENTIONS EXPIRED".
Patient problems: Low Oxygen Saturation
MAUDE report 12646675 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K822082. One submission often covers many models, so check the product and lot details in each record.
- Z-2312-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan/CF: a) 3.0 mm, REF 5-10106, REF 5-10206 b) 3.5 mm, REF 5-10107 c) 4.0 mm, REF 5-10108, REF 5-10208 d) 4.5 mm, REF 5-10109 e) 5.0 mm, REF 5-10110, REF 5-10
- Z-2311-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Tracheal tube/airway management
- Z-2315-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan Uncuffed: a) 2.0 mm, REF 5-10404 b) 2.5 mm, REF 5-10405 c) 3.0 mm, REF 5-10406 d) 3.5 mm, REF 5-10407 e) 4.0 mm, REF 5-10408 f) 4.5 mm, REF 5-10409 g)
- Z-2318-2019 · Jul 30, 2019 · TerminatedSheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10317 Product Usage: Tracheal tube/airway management
- Z-2316-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REF 5-10313 e) 7.0 mm, REF 5-10314 f) 7.5 mm, REF 5-10315 g) 8.0 mm,
- Z-2317-2019 · Jul 30, 2019 · TerminatedSheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management
- Z-2313-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 mm, REF 5-22515 e) 8.0 mm, REF 5-22516 f) 8.5 mm, REF 5-22517 g) 9.0
- Z-2314-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan Preformed: a) 3.0 mm, REF 5-22006, REF 5-22106 b) 3.5 mm, REF 5-22007, REF 5-22107 c) 4.0 mm, REF 5-22008, REF 5-22108, REF 5-22208 d) 4.5 mm, REF 5-22009,
- Z-1956-2019 · Jun 7, 2019 · TerminatedSheridan Endotracheal Tube contained inside Centurion kit code TC7855.
- Z-1761-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 mm, Product Code 5-22514 4) EZ-ENDO 7.5 mm, Product Code 5-2
- Z-1763-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6.5 mm, Product Code 5-10413 3) Uncuffed 7.0 mm, Product Code 5-10414 Product Usage: Tracheal tube/
- Z-1765-2019 · May 24, 2019 · TerminatedSheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Tracheal tube/airway management
- Z-1760-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) CF 6.0 mm, Product Codes: 5-10112, 5-10212 2) CF 6.5 mm, Product Codes: 5-10113, 5-10213 3) CF 7.0 mm, Product Codes: 5-10114, 5-10214 4) CF 7.5 mm,
- Z-1766-2019 · May 24, 2019 · TerminatedSheridan/HVT: 1) Novaplus 7.0 mm, Product Code V5-10314 2) Novaplus 7.5 mm, Product Code V5-10315 3) Novaplus 8.0 mm, Product Code V5-10316 4) Novaplus 8.5 mm, Product Code V5-
- Z-1764-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) HVT 6.0 mm, Product Code 5-10312 2) HVT 6.5 mm, Product Code 5-10313 3) HVT 7.0 mm, Product Code 5-10314 4) HVT 7.5 mm, Product Code 5-10315 5) HVT 8.0
…and 12 more.
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page