Skip to content

Tuohy 1150-4M200Recalls & reported problems

Pajunk GmbH Medizintechnologie · Tuohy with spec. Point 17G x 8", metal stylet · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 14048223022648), received between Jan 7, 2020 and Jan 8, 2020 — classified as 2 deaths.

The most frequently reported specific device problem is “Off-Label Use” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Off-Label Use2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Death2
  • Cardiac Arrest1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • DeathJan 8, 2020Off-Label Use; Off-Label Use

    THE FOLLOWING REPORT OF AN ADVERSE EVENT WAS PROVIDED TO PAJUNK MEDICAL SYSTEMS ON JANUARY 3, 2020: AN ADVERSE EVENT REPORT WAS RECEIVED, IN WHICH YOUR DEVICE WAS INVOLVED. EVENT DESCRIPTION: DURING A VENTRICULAR TACHYCARDIA ABLATION PROCEDURE, A PLEURAL EFFUSION, CARDIAC ARREST AND SUBSEQUENT DEATH OCCURRED. WHILE ATTEMPTING EPICARDIAL ACCESS, A PLEURAL EFFUSION WAS CONFIRMED ON ICE AT THE RV APEX, SUGGESTIVE PERFORATION FROM THE PAJUNK NEEDLE. EPICARDIAL ACCESS WAS CONFIRMED WITH DYE VISUALIZED LAYERING AROUND THE PERICARDIAL SPACE. A WIRE WAS THEN ADVANCED AND OBSERVED TO CURL IN THE EPICARDIAL SPACE WITHOUT ADVANCING BEYOND THE CARDIAC SILHOUETTE. AN AGILIS SHEATH WAS THEN ADVANCED OVER THE WIRE INTO THE PERICARDIAL SPACE. A SECOND WIRE WAS ADVANCED INTO THE EPICARDIUM. THE AGILIS WAS REMOVED AND ADVANCED OVER ONE OF THE WIRES. THE WIRE AND DILATOR WERE FROM THE AGILIS AND THE REMAINING WIRE WAS SECURED TO THE DRAPE. APPROXIMATELY 290 CC OF BLOOD WAS REMOVED FROM THE PERICARDIAL SPACE. THE EFFUSION DID NOT REACCUMULATE AND WAS CLOSELY MONITORED DURING THE PROCEDURE. A CORONARY ANGIOGRAM WAS COMPLETED AFTERWARDS. FOLLOWING THE ABLATION, THE PATIENT BECAME HYPOTENSIVE. BLOOD WAS

    Patient problems: Death

    MAUDE report 9566604 ↗
  • DeathJan 7, 2020Off-Label Use; Adverse Event Without Identified Device or Use Problem; Off-Label Use; Adverse Event Without Identified Device or Use Problem

    (B)(4). DESCRIPTION PROVIDED BY DISTRIBUTOR: AN ADVERSE EVENT REPORT WAS RECEIVED, IN WHICH YOUR DEVICE WAS INVOLVED. EVENT DESCRIPTION: DURING A VENTRICULAR TACHYCARDIA ABLATION PROCEDURE, A PLEURAL EFFUSION, CARDIAC ARREST AND SUBSEQUENT DEATH OCCURRED. WHILE ATTEMPTING EPICARDIAL ACCESS, A PLEURAL EFFUSION WAS CONFIRMED ON ICE AT THE RV APEX, SUGGESTIVE PERFORATION FROM THE PAJUNK NEEDLE. EPICARDIAL ACCESS WAS CONFIRMED WITH DYE VISUALIZED LAYERING AROUND THE PERICARDIAL SPACE. A WIRE WAS THEN ADVANCED AND OBSERVED TO CURL IN THE EPICARDIAL SPACE WITHOUT ADVANCING BEYOND THE CARDIAC SILHOUETTE. AN AGILIS SHEATH WAS THEN ADVANCED OVER THE WIRE INTO THE PERICARDIAL SPACE. A SECOND WIRE WAS ADVANCED INTO THE EPICARDIUM. THE AGILIS WAS REMOVED AND ADVANCED OVER ONE OF THE WIRES. THE WIRE AND DILATOR WERE FROM THE AGILIS AND THE REMAINING WIRE WAS SECURED TO THE DRAPE. APPROXIMATELY 290 CC OF BLOOD WAS REMOVED FROM THE PERICARDIAL SPACE. THE EFFUSION DID NOT REACCUMULATE AND WAS CLOSELY MONITORED DURING THE PROCEDURE. A CORONARY ANGIOGRAM WAS COMPLETED AFTERWARDS. FOLLOWING THE ABLATION, THE PATIENT BECAME HYPOTENSIVE. BLOOD WAS VISUALIZED IN THE PERICARDIAL SPACE. CHEST COMPRESSIONS

    Patient problems: Cardiac Arrest; Death

    MAUDE report 9556364 ↗

FDA clearance and classification

Cleared through 510(k) K040965 on Sep 7, 2004 (decision: Substantially Equivalent). Submitted by Pajunk GmbH as “Pajunk Tuohy Needles, Quincke Needles, Chiba Needles & Crawford Needles”. The FDA review took 146 days. FDA record ↗

The FDA regulates product code BSP (needle, conduction, anesthetic (w/wo introducer)) under 21 CFR 868.5150 as a Class II device. 21 CFR 868.5150 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Tuohy 1150-4M200

Has the Tuohy 1150-4M200 been recalled?

No FDA recall record in our data names this exact device identifier (DI 14048223022648), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Tuohy 1150-4M200?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jan 7, 2020 and Jan 8, 2020. In those reports, the most common specific device problem is “Off-Label Use” (2); the most common patient problems are “Death” (2) and “Cardiac Arrest” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Tuohy 1150-4M200?

2 reports are classified as deaths, 0 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Tuohy 1150-4M200 FDA approved?

It was cleared through 510(k) K040965 on Sep 7, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Tuohy 1150-4M200 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Tuohy 1150-4M200, and is it still sold?

The FDA's device database lists Pajunk GmbH Medizintechnologie as the company responsible for it (brand Tuohy, model 1150-4M200). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER) devices

All NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page