SPROTTE® NRFit 121163-30ARecalls & reported problems
Pajunk GmbH Medizintechnologie · SPROTTE® NRFit · Class 2
All models of SPROTTE NRFit (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 14048223026905), received between Oct 2, 2019 and Dec 8, 2020 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Crack” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Crack3
- Material Deformation3
- No Flow3
- Nonstandard Device3
- Component Misassembled2
- Loss of or Failure to Bond2
- Material Separation1
Patient problems
- No Consequences Or Impact To Patient6
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 8, 2020Nonstandard Device; Material Deformation; No Flow; Nonstandard Device; Material Deformation; No Flow
(B)(4). CATEGORY ONE LSCS, SPINAL ANAESTHETIC. CLINICALLY STRAIGHTFORWARD SPINAL ANAESTHETIC, NO DIFFICULTY OBTAINING CLINICAL END POINT OF SPINAL NEEDLE INSERTION. ON REMOVING CENTRE PORTION OF SPINAL NEEDLE IT FELT VERY GRATEY, IT WAS NOT SMOOTH AS YOU WOULD EXPECT. NO CSF, SEEN FROM NEEDLE, WHICH WAS SURPRISING AS CLINICALLY I FELT I WAS IN THE CORRECT PLACE. SWITCHED TO NEW NEEDLE, WHICH I INSERTED THROUGH SAME INTRODUCER WITH NO OTHER ADJUSTMENTS. CENTRE PART REMOVED SMOOTHLY AS YOU WOULD EXPECT AND FREE FLOWING CSF WAS SEEN. SPINAL ANAESTHETIC SUCCESSFULLY. ON EXAMINING THE FIRST NEEDLE AFTER THE EVENT IT APPEARED BENT. THE EXACT SAME THING HAPPENED TO ME ON (B)(6) 2020, WITH A GRATEY SENSATION ON REMOVING CENTRE PORTION OF NEEDLE, NO CSF DESPITE GOOD CLINICAL END POINT, NEEDLE APPEARED BENT ON REMOVAL DESPITE NON DIFFICULT INSERTION. THE LOT NUMBERS (1324) ETC ARE IDENTICAL ON BOTH NEEDLES BUT THE HUB ON THE GRATEY ONE WAS A PALER PURPLE. ON EXAMINING THE SELECTION OF NEEDLES THERE WAS A MIXTURE OF ONES WITH A PALE AND DARKER PURPLE HUB ALL WITH IDENTICAL LOT NUMBERS. ON OPENING ANOTHER PALE PURPLE HUBBED NEEDLE, AGAIN THE CENTRE BIT WAS GRATEY TO REMOVE, NOT SMOOTH AS IT SH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10964104 ↗MalfunctionDec 8, 2020Nonstandard Device; Material Deformation; No Flow; Nonstandard Device; Material Deformation; No Flow
(B)(4). CATEGORY ONE LSCS, SPINAL ANAESTHETIC. CLINICALLY STRAIGHTFORWARD SPINAL ANAESTHETIC, NO DIFFICULTY OBTAINING CLINICAL END POINT OF SPINAL NEEDLE INSERTION. ON REMOVING CENTRE PORTION OF SPINAL NEEDLE IT FELT VERY GRATEY, IT WAS NOT SMOOTH AS YOU WOULD EXPECT. NO CSF, SEEN FROM NEEDLE, WHICH WAS SURPRISING AS CLINICALLY I FELT I WAS IN THE CORRECT PLACE. SWITCHED TO NEW NEEDLE, WHICH I INSERTED THROUGH SAME INTRODUCER WITH NO OTHER ADJUSTMENTS., CENTRE PART REMOVED SMOOTHLY AS YOU WOULD EXPECT AND FREE FLOWING CSF WAS SEEN. SPINAL ANAESTHETIC SUCCESSFULLY. ON EXAMINING THE FIRST NEEDLE AFTER THE EVENT IT APPEARED BENT. THE EXACT SAME THING HAPPENED TO ME ON (B)(6) 2020, WITH A GRATEY SENSATION ON REMOVING CENTRE PORTION OF NEEDLE, NO CSF DESPITE GOOD CLINICAL END POINT, NEEDLE APPEARED BENT ON REMOVAL DESPITE NON DIFFICULT INSERTION. THE LOT NUMBERS (1324) ETC ARE IDENTICAL ON BOTH NEEDLES BUT THE HUB ON THE GRATEY ONE WAS A PALER PURPLE. ON EXAMINING THE SELECTION OF NEEDLES THERE WAS A MIXTURE OF ONES WITH A PALE AND DARKER PURPLE HUB ALL WITH IDENTICAL LOT NUMBERS. ON OPENING ANOTHER PALE PURPLE HUBBED NEEDLE, AGAIN THE CENTRE BIT WAS GRATEY TO REMOVE, NOT SMOOTH AS IT S
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10964103 ↗MalfunctionDec 8, 2020Nonstandard Device; Material Deformation; No Flow; Nonstandard Device; Material Deformation; No Flow
(B)(4). CATEGORY ONE LSCS, SPINAL ANAESTHETIC. CLINICALLY STRAIGHTFORWARD SPINAL ANAESTHETIC, NO DIFFICULTY OBTAINING CLINICAL END POINT OF SPINAL NEEDLE INSERTION. ON REMOVING CENTRE PORTION OF SPINAL NEEDLE IT FELT VERY GRATEY, IT WAS NOT SMOOTH AS YOU WOULD EXPECT. NO CSF, SEEN FROM NEEDLE, WHICH WAS SUPRISING AS CLINICALLY I FELT I WAS IN THE CORRECT PLACE. SWITCHED TO NEW NEEDLE, WHICH I INSERTED THROUGH SAME INTRODUCER WITH NO OTHER ADJUSTMENTS., CENTRE PART REMOVED SMOOTHLY AS YOU WOULD EXPECT AND FREE FLOWING CSF WAS SEEN. SPINAL ANAESTHETIC SUCCESSFULLY. ON EXAMINING THE FIRST NEEDLE AFTER THE EVENT IT APPEARED BENT. THE EXACT SAME THING HAPPENED TO ME ON (B)(6) 2020, WITH A GRATEY SENSATION ON REMOVING CENTRE PORTION OF NEEDLE, NO CSF DESPITE GOOD CLINICAL END POINT, NEEDLE APPEARED BENT ON REMOVAL DESPITE NON DIFFICULT INSERETION. THE LOT NUMBERS (1324) ETC ARE IDENTICAL ON BOTH NEEDLES BUT THE HUB ON THE GRATEY ONE WAS A PALER PURPLE. ON EXAMINING THE SELECTION OF NEEDLES THERE WAS A MIXTURE OF ONES WITH A PALE AND DARKER PURPLE HUB ALL WITH IDENTICAL LOT NUMBERS. ON OPENING ANOTHER PALE PURPLE HUBBED NEEDLE, AGAIN THE CENTRE BIT WAS GRATEY TO REMOVE, NOT SMOOTH AS IT S
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10964100 ↗MalfunctionOct 31, 2019Component Misassembled; Component Misassembled
IRN# (B)(4). INITIAL REPORTER´S NARRATIVE: "PATIENT REQUIRING SPINAL ANAESTHETIC. DIFFICULT TO DO, MULTIPLE NEEDLES USED. ONE NEEDLE MANAGED TO GET INTO CSF UNFORTUNATELY HAD A LEUR LOCK CONNECTION THEREFORE IT WAS NOT POSSIBLE TO INFILTRATE ANY MEDICATION. SUBSEQUENT NEEDLE HAD TO BE PASSED AND PATIENT SUFFERED AN UNNECESSARY DURAL PUNCTURE. ON INVESTIGATION AFTERWARDS IT APPEARED THE LEUR LOCK NEEDLE HAD BEEN IN AN NRFIT PACKAGE."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9260280 ↗MalfunctionOct 31, 2019Crack; Component Misassembled; Crack; Component Misassembled
(B)(4). INITIAL REPORTER´S NARRATIVE: "UPON INJECTION OF THE SPINAL ANAESTHETIC IT WAS OBSERVED THAT THE HUB OF THE NEEDLE WAS CRACKED & SUBSEQUENTLY THE ANAESTHETIC WAS LEAKING & RUNNING DOWN THE HUB & NEEDLE."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9259874 ↗MalfunctionOct 24, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
IRN# (B)(4). INITIAL REPORTER´S NARRATIVE: CER: PERFORMED A SPINAL INJECTION AND ON THE REMOVAL OF THE NEEDLE FROM THE PATIENT THE NEEDLE HUB CAME AWAY FROM THE NEEDLE ITS SELF. THIS WAS A STRAIGHTFORWARD PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9231205 ↗MalfunctionOct 2, 2019Crack; Crack
IRN# (B)(4). INITIAL REPORTER´S NARRATIVE: AN INJECTION LEAKAGE OCCURED AROUND THE HUB OF THE 24G X 90MM SPROTTE NEEDLE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9143778 ↗MalfunctionOct 2, 2019Loss of or Failure to Bond; Material Separation; Loss of or Failure to Bond; Material Separation
IRN# (B)(4). INITIAL REPORTER´S NARRATIVE: PERFORMED SPINAL AS USUAL ALTHOUGH LIGAMNET WAS TOUGH, THE SPINAL WORKED FINE. ON REMOVAL OF THE SPINAL NEEDLE THE HUB WAS DETACHED FROM THE METAL PART. SURGICAL NEEDLE HOLDERS WERE REQUIRED TO REMOVE THE METAL PART OF THE NEEDLE FROM THE PATIENTS BACK.
Patient problems: Device Embedded In Tissue or Plaque; No Consequences Or Impact To Patient
MAUDE report 9143775 ↗MalfunctionOct 2, 2019Crack; Crack
IRN# (B)(4). INITIAL REPORTER´S NARRATIVE: LEAKAGE AT THE HUB NEEDLE CONNECTION UPON PRIMING OF NEEDLE. ALL OUTSTANDING NEEDLES OF AFFECTED LOT NUMBER HAVE BEEN COLLECTED FROM ALL CLINICAL AREAS AND STORES.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9143772 ↗
Other NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER) devices
- ARROW IPN922632Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)2 recalls0 reports
- ARROW IPN922603Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922604Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN923597Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922631Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922605Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- 22G 1 1/2IN Regional Block Needle 408348Becton, Dickinson And Company · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- BD QUINCKE Spinal Needles 405180Becton, Dickinson And Company · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)23 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page