SPROTTE® NRFit 021163-29A (UDI 14048223026950)Recalls & reported problems
Pajunk GmbH Medizintechnologie · SPROTTE® NRFit · Class 2
All models of SPROTTE NRFit (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 14048223026950), received between Jan 19, 2018 and Aug 5, 2020 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Loss of or Failure to Bond2
- Break1
- Failure To Adhere Or Bond1
- Fracture1
Patient problems
- No Known Impact Or Consequence To Patient5
- Foreign Body In Patient2
- Inadequate Pain Relief1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2020Loss of or Failure to Bond; Loss of or Failure to Bond
IRN#:(B)(4). WHILST PERFORMING A SPINAL ANAESTHETIC THE PLASTIC HUB OF THE NEEDLE BROKE AWAY FROM THE METAL NEEDLE, LEAVING THE METAL NEEDLE STICKING OUT OF THE PATIENTS BACK. NO EXCESSIVE FORCE WAS USED. THIS IS THE ONLY REPORT OF SUCH AN INCIDENT IN OUR ANAESTHETIC DEPARTMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10369803 ↗MalfunctionJun 19, 2019Break; Break
(B)(4). NEEDLE BREAKAGE WHILE A SPINAL WAS IN PROCESS. IT BENT WHERE THE NEEDLE LEFT THE INTRODOCER. AFTER STRAIGHTENED OUT THE NEEDLE BROKE. BROKEN END REMOVED SUCCESSFULLY.
Patient problems: Foreign Body In Patient; No Known Impact Or Consequence To Patient
MAUDE report 8710526 ↗MalfunctionApr 11, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
IRN# 620_109-19. INITIAL REPORTER´S NARRATIVE: PATIENT ON ELECTIVE CAESAREAN SECTION LIST. STRAIGHTFORWARD SPINAL USING THE NR FIT PAJUNK SPROTTE 25G 90MM NEEDLE VIA THE SUPPLIED 30MM INTRODUCER NEEDLE. INJECTED SPINAL DRUGS UNEVENTFULLY. ATTEMPTED TO REMOVE THE INTRODUCER AND SPINAL NEEDLE AS A SINGLE UNIT WHEREUPON THE HUB OF THE SPINAL NEEDLE DETACHED FROM THE NEEDLE LEAVING THE NEEDLE PROTRUDING FROM THE PATIENT'S BACK. THE NEEDLE WAS SUBSEQUENTLY REMOVED MANUALLY AND WAS INTACT.
Patient problems: Inadequate Pain Relief; No Known Impact Or Consequence To Patient
MAUDE report 8502928 ↗MalfunctionFeb 13, 2019Loss of or Failure to Bond; Detachment of Device or Device Component; Loss of or Failure to Bond; Detachment of Device or Device Component
IRN# 610_048-19 . INITIAL REPORTERS NARRATIVE: WHILST PERFORMING A SPINAL THE CLINICIAN EXPERIENCED THE HUB BREAKING/COMING AWAY FROM THE NEEDLE LEAVING THE DISTAL PART IN THE PATIENT WHICH WAS RETRIEVED.
Patient problems: Foreign Body In Patient; No Known Impact Or Consequence To Patient
MAUDE report 8334602 ↗MalfunctionJan 19, 2018Failure To Adhere Or Bond; Fracture; Failure To Adhere Or Bond; Fracture
(B)(4). INITIAL REPORTER´S NARRATIVE: DISTAL END CAME OFF.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7204920 ↗
Other NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER) devices
- ARROW IPN922632Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)2 recalls0 reports
- ARROW IPN922603Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922604Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN923597Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922631Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- ARROW IPN922605Teleflex Incorporated · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- 22G 1 1/2IN Regional Block Needle 408348Becton, Dickinson And Company · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)1 recall0 reports
- BD QUINCKE Spinal Needles 405180Becton, Dickinson And Company · NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)23 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page