Skip to content

Portex 100/127/030Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Portex (409)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15019315019814), received between Nov 22, 2023 and Nov 22, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Complete Blockage” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K790312); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 13, 2025Class IOpen, Classified

    Affected devices have a smaller diameter than expected and may potentially result in inadequate ventilation to the patient. Patient may experience hypoxia, underdose, or cardiopulmonary collapse.

    Z-1289-2025 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Complete Blockage1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 22, 2023Complete Blockage; Material Deformation; Complete Blockage; Material Deformation

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH THE PRODUCT. THE "8FR" INLINE SUCTION CATHETER COULD NOT PASS THROUGH THE TUBE. AFTER DISCOVERING THE PROBLEM, NICU REPLACED THE CATHETER IT WITH A "7FR" CATHETER WITH "NO ISSUES". FURTHER INVESTIGATION REVEALED THAT THE AFFECTED ETT HAD A FLAT SIDE. ADVERSE EFFECTS WERE NOT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18193642 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K790312. One submission often covers many models, so check the product and lot details in each record.

  • Z-2539-2014 · Jul 7, 2014 · Terminated
    H4002S Neonatal/Paed.Intubation Kit 3.0mm Murphy, 15mm Conn, Holder Pediatric Endotracheal Intubation Kit
  • Z-2536-2014 · Jul 7, 2014 · Terminated
    4001PS Neonatal/Paed.Intubation Kit 2.5mm Murphy, LDS 15mm Conn Pediatric Endotracheal Intubation Kit
  • Z-2538-2014 · Jul 7, 2014 · Terminated
    H4002PS Neonatal/Paed.Intubation Kit 3.0mm Murphy, LDS 15mm Conn Pediatric Endotracheal Intubation Kit
  • Z-2537-2014 · Jul 7, 2014 · Terminated
    H4001S Neonatal/Paed.Intubation Kit 2.5mm Murphy,15mm Conn, Holder Pediatric Endotracheal Intubation Kit