Portex 100/199/085Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Portex (409)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Portex endotracheal tube is a breathing tube placed in the airway during surgery or ventilation. The 31 reports mostly describe air leaking from the cuff and pilot balloon, the tube not inflating properly, and tears or punctures in the material or packaging, sometimes leading to the patient needing to be reintubated. Four reports mention injuries and no deaths. There are no recalls for this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 15019315022357), received between Sep 14, 2018 and Apr 29, 2026 — classified as 4 injuries, 27 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak10
- Leak/Splash6
- Inflation Problem5
- Tear, Rip or Hole in Device Packaging3
- Material Integrity Problem2
- Material Puncture/Hole2
- Connection Problem1
- Contamination /Decontamination Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions26
- No Consequences Or Impact To Patient4
- Insufficient Information2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 29, 2026Gas/Air Leak
IT WAS REPORTED THAT AIR LEAK SOUND HEARD AND TV (TIDAL VOLUME) DECREASED BY 100-200ML. THE INFLATION LINE TRACING THE LEAK WAS TORN OFF AT THE ROOT DURING THE PATIENT USE. THERE WAS NO ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25028465 ↗MalfunctionFeb 26, 2026Inflation Problem
IT WAS REPORTED THAT THE ETT (ENDOTRACHEAL TUBE) NOTED TO NOT INFLATE UPON INSERTION, ETT (ENDOTRACHEAL TUBE) REMOVED AND PATIENT WAS REINTUBATED WITHOUT FURTHER ISSUE. THEATRE INTUBATION UNDER GENERAL ANAESTHESIA WAS BEING ADMINISTERED OR PERFORMED. THERE WAS DELAY IN THERAPY. THE EVENT DID NOT INVOLVE DEATH OR SERIOUS DETERIORATION OF A PERSON. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24462437 ↗MalfunctionFeb 2, 2026Inflation Problem; Gas/Air Leak
IT WAS REPORTED THERE WAS REINTUBATION DUE TO PERSISTENT CUFF LEAK AND CONCERN OVER CUFF INTEGRITY, REPEATED INFLATION OF PILOT BALLOON AND SUBSEQUENT LOSS OF PRESSURE. THERE WAS PATIENT INVOLVEMENT. THERE NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24235236 ↗MalfunctionFeb 2, 2026Gas/Air Leak
IT WAS REPORTED THERE WAS REINTUBATION DUE TO A PERSISTENT CUFF LEAK AND CONCERN OVER CUFF INTEGRITY, REPEATED INFLATION OF PILOT BALLOON AND SUBSEQUENT LOSS OF PRESSURE. THERE WAS PATIENT INVOLVEMENT. THERE NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24234716 ↗MalfunctionMar 17, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE BALLOON IN THE TUBE WAS PUNCTURED. THERE WAS NO PATIENT INVOLVEMENT AND ACCORDING TO CUSTOMER'S REPORT, THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21619952 ↗MalfunctionMar 17, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE BALLOON IN THE TUBE WAS PUNCTURED. THERE WAS NO PATIENT INVOLVEMENT AND ACCORDING TO CUSTOMER'S REPORT, THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21619858 ↗MalfunctionFeb 26, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THERE WAS AIR LEAK IN THE CUFF. THERE WAS NO DISINFECTION OR STERILIZATION, AND NO INFECTIOUS DISEASE OCCURRED. THERE WERE NO PATIENT HARM OR ADVERSE EVENTS REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21474480 ↗MalfunctionNov 12, 2024Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT LEAKAGE WAS FOUND AT THE BLUE PILOT BALLOON, AND IT SHOWED MANY BUBBLES UNDER THE WATER. THE EVENT OCCURRED DURING PRE-TEST. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20664671 ↗InjuryJun 24, 2024Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING USE, THE TRACHEOSTOMY TUBE FAILED TO INFLATE. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19595638 ↗InjuryJun 24, 2024Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING USE, THE TRACHEOSTOMY TUBE FAILED TO INFLATE. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19595594 ↗MalfunctionSep 19, 2023Inflation Problem; Connection Problem; Inflation Problem; Connection Problem
IT WAS REPORTED THAT DURING PRE-TESTING, THE CUSTOMER TRIED TO INFLATE THE CUFF, BUT THE CONNECTOR WAS DEFORMED AND THE SYRINGE COULD NOT BE ATTACHED. NO ADVERSE EFFECTS HAVE BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17777199 ↗MalfunctionJun 26, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT BEFORE OPENING THE PACKAGE, THE CUSTOMER NOTICED THE PRODUCT LABEL ON THE PACKAGE READ 100/199/085, BUT THE INTERNAL DIAMETER OF THE ACTUAL PRODUCT IN THE PACKAGE WAS 8.0 MM. NO ADVERSE PATIENT EFFECTS WERE REPORTED BY THE CUSTOMER. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17204064 ↗MalfunctionMay 1, 2023Product Quality Problem; Product Quality Problem
IT WAS REPORTED THAT A WHITE LOOKING POWDER FOUND ON THE BALLOON CUFF OF THE TUBES. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16847606 ↗MalfunctionMay 1, 2023Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT BEFORE OPENING THE PACKAGE, THE CUSTOMER FOUND A HOLE IN IT. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16842982 ↗MalfunctionApr 30, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT A WHITE LOOKING POWDER WAS FOUND ON THE BALLOON CUFF OF THE TUBES. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16841287 ↗MalfunctionApr 29, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT A FOREIGN WHITE POWDER WAS FOUND ON THE BALLON CUFF. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16841257 ↗MalfunctionApr 27, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT A WHITE LOOKING POWDER FOUND ON THE BALLOON CUFF OF THE TUBES. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16823125 ↗InjuryApr 25, 2023Leak/Splash; Leak/Splash
ADDITIONAL INFORMATION: THE CUFF WAS NOT INFLATABLE. THE TUBE WAS CHANGED TO RESOLVE THE ISSUE. NO INJURY TO THE PATIENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16811595 ↗InjuryApr 25, 2023Leak/Splash; Leak/Splash
ADDITIONAL INFORMATION: THE CUFF WAS NOT INFLATABLE. THE TUBE WAS CHANGED TO RESOLVE THE ISSUE. NO INJURY TO THE PATIENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16811470 ↗MalfunctionApr 4, 2023Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT AFTER THE CUSTOMER INSERTED THE PRODUCT INTO THE PATIENT, LEAKAGE OF AIR FROM THE CUFF WAS OBSERVED. NO PATIENT INJURY WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE FOR THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16678257 ↗MalfunctionDec 28, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT DURING THE USE OF THE PRODUCT, LEAKAGE OF AIR FROM IT WAS OBSERVED. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16062518 ↗MalfunctionNov 8, 2022Failure to Deflate; Failure to Deflate
IT WAS REPORTED THAT DURING THE PRE-USE CHECK, THE CUFF WAS UNABLE TO BE DEFLATED. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15752994 ↗MalfunctionJun 21, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THE HOSPITAL EXPERIMENTED ON THE PRODUCT AFTER CASES OF MICRO ASPIRATIONS, AND FOUND THAT CUFFS WERE LEAKING. THE FUNCTIONALITY TESTS WERE PERFORMED IN THE PRESENCE OF A SMITHS REP AND IT WAS OBSERVED THAT INFLATED CUFFS SHOWED LEAKS AND BUBBLES WHEN IMMERSED IN TRANSPARENT CONTAINER FILLED WITH WATER. ADDITIONALLY, IT WAS ALLEGED THE SPRING VALVE OF THE PILOT BALLOON WASN?T WORKING RIGHT. THE TEST WAS DONE IN 189 AND 199. THEY BLEW THE CUFF WITH AIR AT 30 CM AS RECOMMENDED ON THE TUBE, THEN DETACHED THE MANOMETER. THEY RECONNECTED THE MANOMETER AND AFTER A SECOND AND OBSERVED THAT THE READING DROPPED 26CM IN SIZE 7 AND 8 ET. THE PRESSURE WAS NOT GETTING STABILIZED AND WAS CONTINUOUSLY DROPPING. THIS IS THE SAME THING THEY ARE EXPERIENCING AT PATIENT BEDSIDE THOUGHT THEY WERE MAINTAINING THE CUFF PRESSURE FOR 3 TIMES A DAY AND STILL HAVING VAP CASE. ALSO TRIED TO USE 199 PLAIN ET AND FOUND THE PRESSURE WAS NOT GETTING STABLE, CONTINUOUSLY DROPPING. THEY USED 10 UNITS TO ASSESS THE FUNCTIONALITY WITH THE SAME OUTCOME. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14761037 ↗MalfunctionJun 21, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THE HOSPITAL EXPERIMENTED ON THE PRODUCT AFTER CASES OF MICRO ASPIRATIONS, AND FOUND THAT CUFFS WERE LEAKING. THE FUNCTIONALITY TESTS WERE PERFORMED IN THE PRESENCE OF A SMITHS REP AND IT WAS OBSERVED THAT INFLATED CUFFS SHOWED LEAKS AND BUBBLES WHEN IMMERSED IN TRANSPARENT CONTAINER FILLED WITH WATER. ADDITIONALLY, IT WAS ALLEGED THE SPRING VALVE OF THE PILOT BALLOON WASN'T WORKING RIGHT. THE TEST WAS DONE IN 189 AND 199. THEY BLEW THE CUFF WITH AIR AT 30 CM AS RECOMMENDED ON THE TUBE, THEN DETACHED THE MANOMETER. THEY RECONNECTED THE MANOMETER AND AFTER A SECOND AND OBSERVED THAT THE READING DROPPED 26CM IN SIZE 7 AND 8 ET. THE PRESSURE WAS NOT GETTING STABILIZED AND WAS CONTINUOUSLY DROPPING. THIS IS THE SAME THING THEY ARE EXPERIENCING AT PATIENT BEDSIDE THOUGHT THEY WERE MAINTAINING THE CUFF PRESSURE FOR 3 TIMES A DAY AND STILL HAVING VAP CASE. ALSO TRIED TO USE 199 PLAIN ET AND FOUND THE PRESSURE WAS NOT GETTING STABLE, CONTINUOUSLY DROPPING. THEY USED 10 UNITS TO ASSESS THE FUNCTIONALITY WITH THE SAME OUTCOME. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14760625 ↗MalfunctionJul 22, 2021Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
FOLLLOW UP REDACTION REQUIRED AND CORRECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12214935 ↗
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page