Portex 100/126/025Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Portex (409)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 15019315056758), received between Oct 31, 2017 and Jul 16, 2026 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K790312); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 13, 2025Class IOpen, Classified
Affected devices have a smaller diameter than expected and may potentially result in inadequate ventilation to the patient. Patient may experience hypoxia, underdose, or cardiopulmonary collapse.
Z-1289-2025 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
- Inadequacy of Device Shape and/or Size2
- Component Incompatible1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
- Insufficient Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25840750 ↗InjurySep 9, 2021Break; Break
IT WAS REPORTED THAT THE TUBE BROKE OFF NEAR THE HUB AFTER TRYING TO LOOSEN THE BLUE HUB FROM THE TUBE. NO PATIENT INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12450531 ↗MalfunctionJan 15, 2020Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE CUSTOMER REPORTED THE FOLLOWING PRODUCT PROBLEM. THE ISSUE WAS RELATED TO THE ENDOTRACHEAL TUBES NOT ALL BEING THE SAME LENGTH. THE CUSTOMER NOTICED THAT THEY WERE CUT AT DIFFERENT POSITIONS WHICH CAUSED THE INLINE SUCTION CATHETERS TO NOT BE INSERTED TO THE DESIRED DEPTH. THE CUSTOMER BELIEVED THAT IT WAS NOT A TRUE PRODUCT FAILURE, AS THE ENDOTRACHEAL TUBE STILL PROVIDED A PATENT AIRWAY TO THE PATIENT FOR INTUBATION. THE CUSTOMER NOTED THAT IT WAS JUST NOT A STANDARD CUT WHERE THE TUBE WAS ATTACHED TO THE HUB OF THE CONNECTOR. THE CUSTOMER ADJUSTED THE POSITION OF THE INLINE SUCTION CATHETER INSERTION SO THAT THEY COULD CLEAR THE PATIENT'S AIRWAY, AND HAVE THE CATHETER GO TO THE PROPER DEPTH. NO PATIENT INJURY OR COMPLICATIONS WERE REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9593817 ↗MalfunctionJan 15, 2020Component Incompatible; Inadequacy of Device Shape and/or Size; Component Incompatible; Inadequacy of Device Shape and/or Size
THE CUSTOMER REPORTED THE FOLLOWING PRODUCT PROBLEM. THE ISSUE WAS RELATED TO THE ENDOTRACHEAL TUBES NOT ALL BEING THE SAME LENGTH. THE CUSTOMER NOTICED THAT THEY WERE CUT AT DIFFERENT POSITIONS WHICH CAUSED THE INLINE SUCTION CATHETERS TO NOT BE INSERTED TO THE DESIRED DEPTH. THE CUSTOMER BELIEVED THAT IT WAS NOT A TRUE PRODUCT FAILURE, AS THE ENDOTRACHEAL TUBE STILL PROVIDED A PATENT AIRWAY TO THE PATIENT FOR INTUBATION. THE CUSTOMER NOTED THAT IT WAS JUST NOT A STANDARD CUT WHERE THE TUBE WAS ATTACHED TO THE HUB OF THE CONNECTOR. THE CUSTOMER ADJUSTED THE POSITION OF THE INLINE SUCTION CATHETER INSERTION SO THAT THEY COULD CLEAR THE PATIENT'S AIRWAY, AND HAVE THE CATHETER GO TO THE PROPER DEPTH. NO PATIENT INJURY OR COMPLICATIONS WERE REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 9593787 ↗InjuryOct 31, 2017Separation Failure; Separation Failure
IT WAS REPORTED THAT A SMITHS MEDICAL CUFFED BLUE LINE ULTRA® TRACHEOSTOMY TUBE ADAPTER WAS REQUIRED TO BE REMOVED WITH CONSIDERABLE EFFORT OR WAS CUT OFF IN ORDER TO PLACE IN-LINE SUCTION. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6991240 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K790312. One submission often covers many models, so check the product and lot details in each record.
- Z-2539-2014 · Jul 7, 2014 · TerminatedH4002S Neonatal/Paed.Intubation Kit 3.0mm Murphy, 15mm Conn, Holder Pediatric Endotracheal Intubation Kit
- Z-2536-2014 · Jul 7, 2014 · Terminated4001PS Neonatal/Paed.Intubation Kit 2.5mm Murphy, LDS 15mm Conn Pediatric Endotracheal Intubation Kit
- Z-2537-2014 · Jul 7, 2014 · TerminatedH4001S Neonatal/Paed.Intubation Kit 2.5mm Murphy,15mm Conn, Holder Pediatric Endotracheal Intubation Kit
- Z-2538-2014 · Jul 7, 2014 · TerminatedH4002PS Neonatal/Paed.Intubation Kit 3.0mm Murphy, LDS 15mm Conn Pediatric Endotracheal Intubation Kit
- Z-1641-2014 · Apr 10, 2014 · TerminatedPortex Low Dead Space Connector with Sideport, 3.5mm Product Usage: The connector is used to attach the endotracheal tube to the patient breathing circuit. The low dead space con
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page