BCI 3311Recalls & reported problems
Icu Medical, Inc. · Class 2
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 15019517058253), received between Oct 25, 2021 and Aug 30, 2022 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K970209); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Failure to Power Up3
- Component Missing1
- Defective Device1
- Electrical /Electronic Property Problem1
- Fracture1
- Loose or Intermittent Connection1
- Power Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 30, 2022Defective Device; Defective Device
IT WAS REPORTED THAT THE CABLE DID NOT METER CORRECTLY ANYMORE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15320515 ↗MalfunctionAug 30, 2022Component Missing; Component Missing
IT WAS REPORTED THAT THE CONNECTION CORD IS MISSING A PIN. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15320494 ↗MalfunctionJul 26, 2022Fracture; Loose or Intermittent Connection; Fracture; Loose or Intermittent Connection
IT WAS REPORTED THAT THE CABLES WERE FOUND TO HAVE SUSPECTED FRACTURES WITHIN THE PLUG/STRAIN RELIEF AT THE SENSOR END OF THE CABLE WHICH CAUSED INTERMITTENT SIGNAL TRANSMISSION WHEN GENTLY FLEXED. NO PATIENT INJURY WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 15104598 ↗MalfunctionOct 26, 2021Failure to Power Up; Failure to Power Up
IT WAS REPORTED THAT A SMITHS MEDICAL DEVICE NEVER WORKED OUT OF THE PACKAGE. NO PATIENT INVOLVEMENT.
Patient problems: Insufficient Information
MAUDE report 12700747 ↗MalfunctionOct 25, 2021Electrical /Electronic Property Problem; Failure to Power Up; Electrical /Electronic Property Problem; Failure to Power Up
IT WAS REPORTED THAT UPON TESTING BY PULMOLINK, BOTH CABLES WERE FOUND TO HAVE SUSPECTED FRACTURES WITHIN THE PLUG/STRAIN RELIEF AT THE SENSOR END OF THE CABLE WHICH CAUSED INTERMITTENT SIGNAL TRANSMISSION WHEN GENTLY FLEXED. NO PATIENT INVOLVEMENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12693838 ↗MalfunctionOct 25, 2021Failure to Power Up; Power Problem; Failure to Power Up; Power Problem
IT WAS REPORTED THAT UPON TESTING BY PULMOLINK, BOTH CABLES WERE FOUND TO HAVE SUSPECTED FRACTURES WITHIN THE PLUG/STRAIN RELIEF AT THE SENSOR END OF THE CABLE WHICH CAUSED INTERMITTENT SIGNAL TRANSMISSION WHEN GENTLY FLEXED. NO PATIENT INVOLVEMENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12693644 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K970209. One submission often covers many models, so check the product and lot details in each record.
- Z-1014-2013 · Feb 6, 2013 · TerminatedSmiths Medical, BCI Remote Alarm Cable (Normally Open & Normally Closed), Reorder Numbers: WW3398NO & WW3398NC. For use with respiration carbon dioxide measurement equipment. Pr
Other ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE devices
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 deaths8,444 reports
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall607 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE485 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE471 reports
- NA GF-210RNihon Kohden Corporation · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE386 reports
- Trilogy IT2110X21BRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls30 reports
- Trilogy Trilogy Evo Ventilator, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls26 reports
- Trilogy Evo Trilogy Evo Ventilator with oxygen blending model, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls9 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.