Skip to content

Portex 24-1302-22Recalls & reported problems

Icu Medical, Inc. · Class 2

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 15019517075342), received between Dec 2, 2024 and Aug 26, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

7 other recalls cover devices cleared under the same FDA submission (K965017); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Complete Blockage1
  • Detachment of Device or Device Component1
  • Device Damaged by Another Device1
  • Loose or Intermittent Connection1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50461571 ↗
  • MalfunctionJul 14, 2025Break; Break

    IT WAS REPORTED THAT THE DEVICE SNAPPED DURING REMOVAL AT ABOUT 1CM FROM THE BLUE TIP. A COMPUTED TOMOGRAPHY (CT) SCAN OF PATIENT¿S LUMBAR SPINE DIDN'T SHOW THE RETAINED SECTION. A MAGNETIC RESONANCE IMAGING (MRI) COULD NOT BE UNDERTAKEN AS THE PATIENT HAD JUST HAD A KNEE PROSTHESIS IMPLANTED. PATIENT WAS ASYMPTOMATIC UNTIL THE MOMENT THE ANESTHETIST REPORTED THE ISSUE. THE PATIENT WAS AN 81Y OLD, 112KG, MALE. THERE WAS PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 22495369 ↗
  • MalfunctionMay 21, 2025Break; Break

    IT WAS REPORTED THAT THE PATIENT WAS IN LABOR AND UPON REMOVAL OF EPIDURAL CATHETER, THE TIP OF THE EPIDURAL CATHETER WAS MISSING. PER REPORTER, THE CATHETER WAS REMOVED WITH EASE, AND THERE WAS NO DIFFICULTY OR TENSION. THE PATIENT UNDERWENT AN MRI AND NO FOREIGN BODY WAS REPORTED. A CT SCAN WAS ALSO PERFORMED, AND NO FOREIGN BODY WAS IDENTIFIED. THE PATIENT WAS DISCHARGED. REPORTER STATED THAT UPON ADDITIONAL REVIEW OF THE MRI AND CT RESULTS, THERE WAS THEN AN ADDENDUM WHICH STATED THAT THE TIP OF THE EPIDURAL CATHETER WAS PRESENT ON THE MRI. THE CUSTOMER STATED THAT THIS ADDENDUM HAD NOT BEEN COMMUNICATED WITH THE ANESTHETIC OR OBSTETRIC TEAMS AND THAT THE CLINICAL TEAMS WERE UNAWARE OF THIS CHANGE IN FINDINGS.

    Patient problems: Insufficient Information

    MAUDE report 22061143 ↗
  • MalfunctionFeb 13, 2025Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT TWO SYRINGES HAVE ENTRAINED AIR VIA THE PLUNGER RATHER THAN ASPIRATING THE SALINE. THE PLUNGER ELEMENT WAS TOO LOOSE INSIDE THE BARREL WHEN PULLED BACK. THE USER EXPLAINED THAT PROBABLY IT WOULDN¿T HAVE BEEN AN ISSUE IF USING SALINE FOR LOSS OF RESISTANCE, BUT IF USING AIR, THEN THE AIR MIGHT HAVE EXPELLED ¿VIS¿ THE PLUNGER ELEMENT, GIVING A FALSE END POINT FOR LOSS OF RESISTANCE, AS ONE CAN¿T CHECK IF THE SYRINGE IS WORKING BEFOREHAND. THERE WAS PATIENT INVOLVEMENT, BUT NO HUMAN HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21379081 ↗
  • MalfunctionDec 31, 2024Break; Device Damaged by Another Device; Break; Device Damaged by Another Device

    IT WAS REPORTED THAT IT WAS DIFFICULT FOR THE OPERATOR OF THE DEVICE TO FEED THE CATHETER THROUGH TOUEY EVEN AFTER THEY TRIED TO ROTATE IT. BOTH TOUEY AND CATHETER WERE EVENTUALLY REMOVED BUT IT APPEARED TO BE BROKEN. THERE WERE FOUR PATIENTS INVOLVED IN THIS EVENT. UNKNOWN PATIENT HARM WAS REPORTED DUE TO THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21048371 ↗
  • MalfunctionDec 30, 2024Break; Break

    IT WAS REPORTED THAT WHILE SITING THE EPIDURAL FOR A WOMAN IN LABOR, THE HEALTHCARE PROFESSIONAL WAS UNABLE TO FEED THE CATHETER THROUGH THE TOUEY. TRIED TO ROTATE 180 DEGREES BUT STILL UNABLE TO FEED CATHETER. TOUEY AND CATHETER WERE REMOVED, BUT WHILE PULLING BACK, A SNAP WAS FELT. TIP OF CATHETER HAD BEEN SHEARED OFF. THERE WAS PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21035947 ↗
  • MalfunctionDec 2, 2024Complete Blockage; Material Rupture; Complete Blockage; Material Rupture

    IT WAS REPORTED THAT EPIDURAL FILTER SPLIT AT THE SEAM DURING TEST DOSE INJECTION AT LOW PRESSURE, FLUID DID NOT GO DOWN THE CATHETER. NO ADVERSE EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20827979 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K965017. One submission often covers many models, so check the product and lot details in each record.

  • Z-1101-2026 · Dec 19, 2025 · Open, Classified
    Portex Spinal Tray, Item No. A3729-24 A3595
  • Z-0211-2007 · Oct 27, 2006 · Terminated
    Portex Continuous Epidural tray Ref 4948-17
  • Z-0244-06 · Nov 17, 2005 · Terminated
    4904 Glass L.O.R. Syringe
  • Z-0243-06 · Nov 17, 2005 · Terminated
    4903 Glass L.O.R. Syringe
  • Z-0242-06 · Nov 17, 2005 · Terminated
    4902 Glass L.O.R. Syringe
  • Z-0245-06 · Nov 17, 2005 · Terminated
    4950-16 Portex Epidural Mini-pack
  • Z-0166-04 · Oct 10, 2003 · Terminated
    Portex Spinal Tray with 25G Whitacre Needle. Kit contains Bupivacaine in Dextrose, Epinephrine,Lidocaine, Lidocaine in Dextrose and Tetracaine Catalog Number: 15545-20

Other Anesthesia conduction kit devices

All Anesthesia conduction kit devices →