Portex 101/563/090Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Portex (409)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 15019517103427), received between Jan 19, 2024 and Nov 4, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Crack” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Crack4
- Incomplete or Inadequate Connection2
- Break1
- Decrease in Suction1
- Device Markings/Labelling Problem1
- Fracture1
- Gas/Air Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 4, 2025Crack; Crack
IT WAS REPORTED THAT AN INCIDENT OCCURRED IN THE INTENSIVE CARE UNIT REGARDING THE PRODUCT, WHERE THE BLUE CONNECTOR ON THE SUBGLOTTIC CANAL WHERE THE SYRINGE IS ATTACHED FOR ASPIRATION IS CRACKED, AND AIR FROM THE ROOM IS ENTERING THE SYRINGE INSTEAD OF MUCUS FROM THE AIRWAY. THERE WAS PATIENT INVOLVEMENT, WHICH LED TO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23470284 ↗MalfunctionOct 21, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE SUCTIONAID TRACHEOSTOMY TUBE TAKEN FROM THE SIZE 9 BLUPERC KIT COULD NOT MAINTAIN CUFF PRESSURE. CUFF PRESSURE MANOMETER OBSERVATIONS CONFIRMED LOSS OF PRESSURE WHILE IN-USE WITH THE PATIENT. THE DEVICE WAS RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23352708 ↗MalfunctionAug 25, 2025Fracture; Fracture
IT WAS REPORTED THAT THE ABOVE-CUFF SUCTION AID BLUE PORT WAS CRACKED, AND THE TRACHEOSTOMY WAS SUCCESSFULLY CHANGED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22883951 ↗MalfunctionJun 14, 2025Decrease in Suction; Decrease in Suction
IT WAS REPORTED THAT THE HEALTHCARE PROVIDER WAS UNABLE TO PASS WHITE SLIM DILATOR THROUGH PURPLE INTRODUCER DUE TO NARROWING OF DILATOR. THE EVENT DID PROLONG THE TRACHEOSTOMY. THERE WAS REPORTED PATIENT INVOLVEMENT, BUT NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22208137 ↗MalfunctionJul 2, 2024Break; Break
IT WAS REPORTED THAT WITH THE DEVICE THE PATIENT WAS UNABLE TO OR HAVING DIFFICULTIES IN ASPIRATING FROM SUBGLOTTIC PORT DUE TO BROKEN PLASTIC, MAKING WEANING OF PATIENT CHALLENGING DUE TO A POTENTIAL ASPIRATION RISK. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19661772 ↗MalfunctionMay 1, 2024Crack; Crack
IT WAS REPORTED THAT THE SUCTION AID PORT CRACKED, MAKING IT DIFFICULT TO ASPIRATE SECRETIONS ON THE PATIENT. THIS POTENTIALLY CAUSED AN INCREASED RISK IN ASPIRATION PNEUMONIA. "AS THIS IS SOMEONE¿S AIRWAY THE DEVICE CAN NOT BE CHANGED EASILY. THESE ARE FOR THE TUBES THAT HAD BEEN PERCUTANEOUSLY INSERTED ON DIFFERENT SIZE TUBES." THE DEVICE WAS ALREADY INSERTED IN THE PATIENT. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT HARM/ADVERSE EVENT REPORTED. THE EVENT OCCURRED IN THE HOSPITAL DURING. "STEPS TO RESOLVE THE ISSUE: NEED TO MARK TUBES AND HIGHLIGHT THE PROBLEM. HAD THE WHOLE TUBE CHANGED WHEN PATIENT WAS ABLE TO. ISSUE RESOLVED: NO."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19222029 ↗MalfunctionMay 1, 2024Crack; Incomplete or Inadequate Connection; Crack; Incomplete or Inadequate Connection
IT WAS REPORTED THAT THE SUCTIONAID PORT CRACKED, MAKING IT DIFFICULT TO ASPIRATE SECRETIONS ON PATIENTS WITH THE POTENTIAL TO CAUSE INCREASED RISK IN ASPIRATION PNEUMONIA. OCCURRING IN MULTIPLE TUBES NOW ON THE BLU SELECT PORTEX SUCTIONAID TUBE. THE EVENT OCCURRED IN THE HOSPITAL WHILE IN USE WITH PATIENT AND THE DEVICE WAS OPERATED BY A HEALTHCARE PROFESSIONAL. THE ISSUE WAS NOT REVOLVED. THERE WAS A NEED TO MARK THE TUBES AND HIGHLIGHT THE PROBLEM AND HAVE THE WHOLE TUBE CHANGED WHEN THE PATIENT WAS ABLE TO. THERE WAS PATIENT INVOLVEMENT AND NO HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19220826 ↗MalfunctionMay 1, 2024Crack; Incomplete or Inadequate Connection; Crack; Incomplete or Inadequate Connection
IT WAS REPORTED THAT THE SUCTIONAID PORT CRACKED, MAKING IT DIFFICULT TO ASPIRATE SECRETIONS ON PATIENTS WITH THE POTENTIAL TO CAUSE INCREASED RISK IN ASPIRATION PNEUMONIA. OCCURRING IN MULTIPLE TUBES NOW ON THE BLU SELECT PORTEX SUCTIONAID TUBE. THE EVENT OCCURRED IN THE HOSPITAL WHILE IN USE WITH PATIENT AND THE DEVICE WAS OPERATED BY A HEALTHCARE PROFESSIONAL. THE ISSUE WAS NOT REVOLVED. THERE WAS A NEED TO MARK THE TUBES AND HIGHLIGHT THE PROBLEM AND HAVE THE WHOLE TUBE CHANGED WHEN THE PATIENT WAS ABLE TO. THERE WAS PATIENT INVOLVEMENT AND NO HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19220807 ↗MalfunctionJan 19, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT WHEN THE PURCHASED PRODUCT ARRIVED, THE OPERATOR CHECKED IT AND FOUND A SAMPLE STICKER ON THE BOX. NO PATIENT INVOLVEMENT REPORTED. THERE WAS NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18545686 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.