Bivona SU15HN55NSA145NRecalls & reported problems
Icu Medical, Inc. · Class 2
All models of Bivona (469)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 15019517142358), received between Mar 5, 2019 and Jan 3, 2020 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Damaged by Another Device” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Damaged by Another Device3
- Material Fragmentation1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 3, 2020Material Fragmentation; Material Fragmentation
INFORMATION WAS RECEIVED INDICATING THAT A PATIENT WITH A SMITHS MEDICAL PORTEX® BIVONA CUSTOMIZED TRACHEOSTOMY TUBE WAS CONNECTED TO A VENTILATOR IN WHICH THE RING WAS OBSERVED TO BREAK OFF. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9549555 ↗InjuryMar 5, 2019Device Damaged by Another Device; Device Damaged by Another Device
INFORMATION WAS RECEIVED THAT WHILE IN USE A SMITHS MEDICAL TRACHEOSTOMY SILICONE BIVONA TUBE "BROKE AT THE NECK WHEN THEY ATTACHED AN HME". "TRACED THE HME ON TRACH TUBE AND GETS STUCK. THEY BROKE IT WHILE ATTEMPTING TO REMOVE IT." OF AN UNKNOWN BRAND. SUBSEQUENTLY, A TRACHEOSTOMY TUBE CHANGE OUT WAS PERFORMED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8392253 ↗InjuryMar 5, 2019Device Damaged by Another Device; Device Damaged by Another Device
INFORMATION WAS RECEIVED THAT WHILE IN USE A SMITHS MEDICAL TRACHEOSTOMY SILICONE BIVONA TUBE "BROKE AT THE NECK WHEN THEY ATTACHED AN HME". "TRACED THE HME ON TRACH TUBE AND GETS STUCK. THEY BROKE IT WHILE ATTEMPTING TO REMOVE IT." OF AN UNKNOWN BRAND. SUBSEQUENTLY, A TRACHEOSTOMY TUBE CHANGE OUT WAS PERFORMED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8392070 ↗InjuryMar 5, 2019Device Damaged by Another Device; Device Damaged by Another Device
INFORMATION WAS RECEIVED THAT WHILE IN USE A SMITHS MEDICAL TRACHEOSTOMY SILICONE BIVONA TUBE "BROKE AT THE NECK WHEN THEY ATTACHED AN HME". "TRACED THE HME ON TRACH TUBE AND GETS STUCK. THEY BROKE IT WHILE ATTEMPTING TO REMOVE IT." OF AN UNKNOWN BRAND. SUBSEQUENTLY, A TRACHEOSTOMY TUBE CHANGE OUT WAS PERFORMED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8391662 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.
- Z-0260-2007 · Nov 17, 2006 · TerminatedCustom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page