Bivona CMZ3289NRecalls & reported problems
Icu Medical, Inc. · Class 2
All models of Bivona (469)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15019517148688), received between May 22, 2019 and Oct 13, 2022 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Connection Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Connection Problem2
- Device Markings/Labelling Problem1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 13, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT THE CUSTOM-MADE TRACHEOSTOMY TUBE IS NOT ACCORDING TO SPECIFICATIONS. THE FIXED FLANGE IS NOT FIXED AND A RING WAS ADDED TOO CLOSE TO THE PROXIMAL FOME-CUFF. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15598755 ↗InjuryJul 30, 2019Connection Problem; Connection Problem
INFORMATION WAS RECEIVED THAT WHEN DISCONNECTING A SMITHS MEDICAL (B)(4) CUSTOM TUBE FROM THE ISO CONNECTOR, THE ISO CONNECTOR DETACHED FROM THE REST OF THE CANNULA LEAVING NO CONNECTION FOR THE VENTILATOR. A REPLACEMENT CANNULA HAD TO BE INSERTED PROMPTLY. NO FURTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8843385 ↗InjuryMay 22, 2019Connection Problem; Connection Problem
IT WAS REPORTED THAT TRACHEOSTOMY TUBE WAS DISCONNECTED FROM THE ISO CONNECTOR, THE ISO CONNECTOR DETACHED FROM THE REST OF THE CANNULA LEAVING NO CONNECTION TO THE VENTILATOR. AN UNPLANNED REPLACEMENT CANNULA HAD TO BE INSERTED PROMPTLY. IN ADDITION, A RESUSCITATOR WAS ALSO USED ON THE PATIENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8634220 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.
- Z-0260-2007 · Nov 17, 2006 · TerminatedCustom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page