Bivona XT18ES50NSC980NRecalls & reported problems
Icu Medical, Inc. · Class 2
All models of Bivona (469)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15019517150650), received between Mar 17, 2020 and Apr 3, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak2
- Material Fragmentation1
- Material Puncture/Hole1
Patient problems
- Low Oxygen Saturation2
- Tachycardia2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 3, 2020Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
INFORMATION WAS RECEIVED INDICATING THAT THE CUFF TO A SMITHS MEDICAL PORTEX BIVONA FLEXTEND¿ TTS¿ TRACHEOSTOMY TUBE WAS FOUND LEAKING UPON ASSESSING THE PATIENT'S DROP-IN OXYGEN SATURATION LEVEL. THE CUFF WAS RE-INFLATED WITH 3MLS OF WATER AND WAS LATER CHECKED TO FIND ONLY 0.5MLS OF WATER LEFT. MOTHER STATED THAT THE ENT AND RESPIRATORY THERAPY WERE NOTIFIED WITH INSTRUCTION TO RE-INFLATE THE CUFF WITH WATER WHEN NEEDED UNTIL THE REPLACEMENT TRACH TUBE ARRIVED. THE TRACH TUBE WAS EVENTUALLY CHANGED OUT. IT WAS REPORTED THAT PART OF THE WEDGE USED TO REMOVE THE TUBE HAD CAUSED A MUCOUS PLUG WHICH WAS RELIEVED BY SUCTIONING. THE PATIENT'S MOTHER REPORTED AN ELEVATED HEART RATE AT THAT TIME BUT THEN RESOLVED. THERE WERE NO FURTHER REPORTED ADVERSE EFFECTS.
Patient problems: Low Oxygen Saturation; Tachycardia
MAUDE report 9925316 ↗InjuryMar 17, 2020Fluid/Blood Leak; Material Fragmentation; Fluid/Blood Leak; Material Fragmentation
INFORMATION WAS RECEIVED INDICATING THAT THE CUFF TO A SMITHS MEDICAL PORTEX BIVONA FLEXTEND¿ TTS¿ TRACHEOSTOMY TUBE WAS FOUND LEAKING UPON ASSESSING THE PATIENT'S DROP-IN OXYGEN SATURATION LEVEL. THE CUFF WAS RE-INFLATED WITH 3MLS OF WATER AND WAS LATER CHECKED TO FIND ONLY 0.5MLS OF WATER LEFT. MOTHER STATED THAT THE ENT AND RESPIRATORY THERAPY WERE NOTIFIED WITH INSTRUCTION TO RE-INFLATE THE CUFF WITH WATER WHEN NEEDED UNTIL THE REPLACEMENT TRACH TUBE ARRIVED. THE TRACH TUBE WAS EVENTUALLY CHANGED OUT. IT WAS REPORTED THAT PART OF THE WEDGE USED TO REMOVE THE TUBE HAD CAUSED A MUCOUS PLUG WHICH WAS RELIEVED BY SUCTIONING. THE PATIENT'S MOTHER REPORTED AN ELEVATED HEART RATE AT THAT TIME BUT THEN RESOLVED. THERE WERE NO FURTHER REPORTED ADVERSE EFFECTS.
Patient problems: Low Oxygen Saturation; Tachycardia
MAUDE report 9845351 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.
- Z-0260-2007 · Nov 17, 2006 · TerminatedCustom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page