Skip to content

CADD 21-2112-0402-51LRecalls & reported problems

Icu Medical, Inc. · Class 2

All models of CADD (123)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This ICU Medical CADD neuraxial administration set has 122 adverse event reports, with no deaths or injuries. Many reports describe cracks in the membrane found during inspection, along with sensing and infusion failures. Three open recalls name this device: two about false upstream occlusion alarms that can interrupt an active infusion, and one tied to software. Reports rose from 38 in the previous year to 64 in the latest year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
129
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 129 adverse event reports that list this exact device (DI 15019517155006), received between Nov 2, 2023 and Sep 30, 2026 — classified as 129 malfunctions.

The most frequently reported specific device problem is “Crack” (53 reports).

3 FDA recall records name this device's identifier, 3 still open.

4 other recalls cover devices cleared under the same FDA submission (K170982); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 5 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 6 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 3 reportsMar 2025: 6 reportsApr 2025: 1 reportsMay 2025: 5 reportsJun 2025: 6 reportsJul 2025: 5 reportsAug 2025: 6 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 3 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 6 reportsMar 2026: 9 reportsApr 2026: 1 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 37 reportsAug 2026: 2 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202637
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 10, 2025Class IOpen, Classified

    Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

    Z-1668-2025 · Root cause: Labeling design · Recall details · FDA record ↗

  2. Apr 10, 2025Class IOpen, Classified

    Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

    Z-1667-2025 · Root cause: Labeling design · Recall details · FDA record ↗

  3. Feb 27, 2024Class IOpen, Classified

    The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.

    Z-2420-2024 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗

Most reported problems

Device problems

  • Crack53
  • Device Sensing Problem15
  • Failure to Infuse12
  • Failure to Sense11
  • Device Alarm System7
  • Detachment of Device or Device Component6
  • Degraded5
  • Air/Gas in Device4

Patient problems

  • No Clinical Signs, Symptoms or Conditions123
  • Insufficient Information6

Recent adverse event reports

The 25 most recent of 129 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Device Sensing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953123 ↗
  • MalfunctionSep 25, 2026Defective Alarm; Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50892620 ↗
  • MalfunctionSep 25, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50892619 ↗
  • MalfunctionSep 25, 2026Defective Alarm; Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50892617 ↗
  • MalfunctionSep 22, 2026Crack

    It was reported that the downstream occlusion was damaged(cracked). It occurred during preventive maintenance. There was no patient involvement and harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50855524 ↗
  • MalfunctionSep 21, 2026Pumping Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50845540 ↗
  • MalfunctionSep 9, 2026Failure to Infuse

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50680850 ↗
  • MalfunctionSep 7, 2026Excess Flow or Over-Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50658653 ↗
  • MalfunctionAug 13, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50240281 ↗
  • MalfunctionAug 11, 2026Crack; Calibration Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50201868 ↗
  • MalfunctionJul 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25816092 ↗
  • MalfunctionJul 14, 2026Crack

    IT WAS REPORTED THAT CRACKS FOUND IN THE MEMBRANE OF THE CADD PUMP DURING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT. HOWEVER, IF THIS FAILURE MODE REOCCURS DURING INFUSION IT MAY LEAD TO FLUID INGRESS INTO THE PUMP WHICH WILL INTERRUPT THERAPY AND POTENTIALLY IMPACT PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25815848 ↗
  • MalfunctionJul 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25815791 ↗
  • MalfunctionJul 14, 2026Crack

    IT WAS STATED THAT THERE WAS CRACKS IN THE MEMBRANE. THE ISSUE WAS FOUND DURING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25815633 ↗
  • MalfunctionJul 14, 2026Crack

    IT WAS REPORTED THAT CRACKS IN THE MEMBRANE OF THE INFUSION PUMP WERE IDENTIFIED DURING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25815615 ↗
  • MalfunctionJul 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25815255 ↗
  • MalfunctionJul 14, 2026Crack

    IT WAS REPORTED THAT CRACKS IN THE MEMBRANE OF THE INFUSION PUMP WERE IDENTIFIED DURING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25815190 ↗
  • MalfunctionJul 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25815135 ↗
  • MalfunctionJul 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25814991 ↗
  • MalfunctionJul 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25814937 ↗
  • MalfunctionJul 14, 2026Crack

    IT WAS REPORTED THAT CRACKS FOUND IN THE MEMBRANE OF THE CADD PUMP DURING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT. HOWEVER, IF THIS FAILURE MODE REOCCURS DURING INFUSION IT MAY LEAD TO FLUID INGRESS INTO THE PUMP WHICH WILL INTERRUPT THERAPY AND POTENTIALLY IMPACT PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25814826 ↗
  • MalfunctionJul 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25814809 ↗
  • MalfunctionJul 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25814482 ↗
  • MalfunctionJul 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25813657 ↗
  • MalfunctionJul 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25813520 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170982. One submission often covers many models, so check the product and lot details in each record.

  • Z-1662-2025 · Apr 10, 2025 · Open, Classified
    CADD Solis HSPCA Pump ***Updated 5/27/26*** Scope updated to apply specifically to customers using the wireless system. Affected products associated with the change in scope: PUM
  • Z-1681-2025 · Apr 10, 2025 · Open, Classified
    CADD-Solis Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-51; 2) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW,
  • Z-0466-2025 · Jun 26, 2024 · Open, Classified
    smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps
  • Z-0963-2023 · Dec 9, 2022 · Open, Classified
    CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24 3) REF 21-7302-24, 4) REF 21-7321-24, 5) REF 21-7322-24, 6) REF 21-7323-24, 7) REF 21-7324-24, 8)

Other Neuraxial Administration Set - Intrathecal delivery devices

All Neuraxial Administration Set - Intrathecal delivery devices →

Sources and updates