CADD recallZ-1667-2025 · Smiths Medical ASD, Inc. · initiated Apr 10, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
FDA root cause category: Labeling design
What the company is doing
On April 10, 2025, the firm notified affected consignees via letters titled "Urgent Medical Device Correction." Customers were informed of conditions under which the false Upstream Occlusion Alarm may occur. The USO alarm can be cleared during these conditions by removing the administration from the pump, re-attaching the administration set, and restarting delivery. Customers should verify there is no upstream occlusion present. The firm will be making an update to the Infusion Pump Operator's Manual to include these conditions.
Recall details
- Recall number
- Z-1667-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 10, 2025
- Posted by the FDA
- May 7, 2025
- Recalling firm
- Smiths Medical ASD, Inc.
- Quantity
- 185,975 US; 127,701 OUS Total
- Product code
- MEA
- FDA submissions
- K072144, K130394, K170982
- Distribution
- Worldwide - US nationwide distribution.
Product description
CADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR CMSA 21-2111-0300-09_CMSA CADD Solis PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR FUKUOKA 21-2111-0300-09_FUK CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR KYOTO 21-2111-0300-09_KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OKAYAMA 21-2111-0300-09_OKA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OOMIYA 21-2111-0300-09_OMI CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OSAKA 21-2111-0300-09_OSA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR SAPPORO 21-2111-0300-09_SAP CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR TOKYO 21-2111-0300-09_TOK CADD_SOLIS_MODEL 2110_ENGLISH_1/EA 21-2111-0200-50JP CADD-Solis VIP CADD-SOLIS V4.2 LOAN FUK 21-2111-0402-09-FUK CADD-SOLIS V4.2 LOAN NAG 21-2111-0402-09-NAG CADD-SOLIS V4.2 LOAN OKA 21-2111-0402-09-OKA CADD-SOLIS V4.2 LOAN OSA 21-2111-0402-09-OSA CADD-SOLIS V4.2 LOAN SAI 21-2111-0402-09-SAI CADD-SOLIS V4.2 LOAN SAP 21-2111-0402-09-SAP CADD-SOLIS V4.2 LOAN TOKE 21-2111-0402-09-TOKE CADD-SOLIS V4.2 LOAN TOKW 21-2111-0402-09-TOKW CADD-SOLIS V4.2 LOAN YELLOW KYO 21-2112-0402-09-KYO CADD-SOLIS V4.2 LOAN YELLOW NAG 21-2112-0402-09-NAG CADD-SOLIS V4.2 LOAN YELLOW YOK 21-2112-0402-09-YOK CADD-SOLIS V4.2 LOAN YOK 21-2111-0402-09-YOK Number:21-2111-0403-02Part Type:List Number/Finished GoodLifecycle Phase:03-ReleasedDescription:PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, FRENCH, 1/EA 21-2111-0403-02 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, YELLOW, ENGLISH, 1/EA 21-2112-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA 21-2111-0300-00 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DANISH 1/EA 21-2111-0300-12 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DUTCH 1/EA 21-2111-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, FRENCH 1/EA 21-2111-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, GERMAN 1/EA 21-2111-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, ITALIAN 1/EA 21-2111-0300-07 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, JAPANESE 21-2111-0300-09JP PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, FRENCH 1/EA 21-2111-0300-232 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA 21-2111-0300-231 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, SWEDISH 1/EA 21-2111-0300-244 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA 21-2111-0300-50 Not PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, PORTUGUESE 1/EA 21-2111-0300-17 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SPANISH 1/EA 21-2111-0300-08 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SWEDISH 1/EA 21-2111-0300-14 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, DUTCH 1/EA 21-2112-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, GERMAN 1/EA 21-2112-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA 21-2112-0300-00 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, DANISH 1/EA 21-2101-0200-12 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, FRENCH 1/EA 21-2101-0200-02 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, GERMAN 1/EA 21-2101-0200-03 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, ITALIAN 1/EA 21-2101-02
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
All CADD-Solis and CADD-Solis VIP Ambulatory Infusion Pumps UDI-DI: 10610586034787 10610586038778 10610586038785 10610586038808 10610586038822 10610586039232 10610586041037 10610586041051 10610586042591 10610586042669 10610586042829 10610586042836 15019517084368 15019517096378 15019517096385 15019517096392 15019517099690 15019517101003 15019517101010 15019517101034 15019517101041 15019517126587 15019517126594 15019517126600 15019517126617 15019517128697 15019517135343 15019517149999 15019517150001 15019517150292 15019517154290 15019517154887 15019517154924 15019517154979 15019517154986 15019517154993 15019517155006 15019517155013 15019517155044 15019517246735 15019517247008 15019517247169 15019517247176 15019517247183 15019517247206 00610586038900 00610586039310 04546027095620 04546027095637 04546027095644 04546027095668 04546027095675 04546027095682 04546027095699 04546027095705 04546027095712 04546027095729 04546027095736 04546027095743 04546027095897 04546027095903 04546027095910 04546027422983 04546027424116 04546027424130 04546027424147 04546027424178 04546027424185 04546027424192 04546027424208 04546027424598 05019517154934 05019517155108 10610586031854 10610586033933 10610586033940 10610586035067 10610586035357 10610586035395 10610586035401 10610586035418 10610586035432 10610586035456 10610586035494 10610586038501 10610586038518 10610586038792 10610586038839 10610586038860 10610586038891 10610586038914 10610586038921 10610586038938 10610586038952 10610586038969 10610586038983 10610586038990 10610586039010 10610586039041 10610586039065 10610586039072 10610586039089 10610586039195 10610586039256 10610586039270 10610586039409 10610586039416 10610586039430 10610586039447 10610586039454 10610586039461 10610586039478 10610586039485 10610586039492 10610586040429 10610586040443 10610586040528 10610586041822 10610586042607 10610586042614 10610586042843 15019517084382 15019517084399 15019517084405 15019517084412 15019517084429 15019517084443 15019517084450 15019517084467 15019517086720 15019517149517 15019517149975 15019517150025 15019517150032 15019517150049 15019517150056 15019517150063 15019517150070 15019517150087 15019517150094 15019517150100 15019517150117 15019517150124¿ 15019517150131 15019517150179 15019517150193 15019517150216 15019517150278 15019517150308 15019517150315 15019517150322 15019517150391 30610586035344 30610586035368 30610586035511 30610586039205 30610586039397 30610586040430 30610586040539 35019517128677 35019517128684 35019517128691 35019517128721 35019517128738 35019517128745 35019517128769 35019517128783 35019517128837 35019517135316 35019517135323 35019517135330 35019517135354 35019517135361 35019517154850 35019517154867 35019517154874 35019517154904 35019517154911 35019517154942 35019517154959 35019517154966 35019517155031 35019517155055 35019517155062 35019517155079 35019517155086 35019517155130
Devices named in this recall
16 devices in the FDA device registry whose identifier appears in the recall's code information.
- CADD 21-2120-0105-01Icu Medical, Inc. · Pump, infusion4 recalls3 deaths5,914 reports
- CADD 21-2120-0102-51Icu Medical, Inc. · Pump, infusion4 recalls2 deaths3,555 reports
- CADD 21-2111-0402-51Icu Medical, Inc. · Pump, infusion, pca4 recalls1,535 reports
- CADD 21-2120-0103-01Icu Medical, Inc. · Pump, infusion4 recalls585 reports
- CADD 21-2111-0401-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION4 recalls461 reports
- CADD 21-2120-0102-02Icu Medical, Inc. · Pump, infusion3 recalls381 reports
- CADD 21-2111-0400-51Icu Medical, Inc. · Pump, infusion4 recalls124 reports
- CADD 21-2112-0402-51LIcu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery3 recalls129 reports
- CADD 21-2111-0402-51LIcu Medical, Inc. · Pump, infusion, pca3 recalls127 reports
- CADD 21-2112-0300-00Icu Medical, Inc. · Pump, infusion, pca4 recalls6 reports
- CADD 2110Icu Medical, Inc. · Pump, infusion, pca4 recalls4 reports
- CADD 2120Icu Medical, Inc. · Pump, infusion4 recalls2 reports
- CADD 21-2111-0300-231Icu Medical, Inc. · Pump, infusion, pca4 recalls0 reports
- CADD 2120Icu Medical, Inc. · Pump, infusion4 recalls0 reports
- CADD 2110Icu Medical, Inc. · Pump, infusion, pca4 recalls0 reports
- CADD 2120Icu Medical, Inc. · Pump, infusion4 recalls0 reports
Product lines
Other devices under the same FDA submission
16 devices cleared under K072144, K130394, K170982 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CADD 21-2111-0300-01Icu Medical, Inc. · Pump, infusion, pca1 recall1,116 reports
- CADD 21-2112-0402-51Icu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery3 recalls469 reports
- CADD 21-2112-0300-01Icu Medical, Inc. · Pump, infusion, pca1 recall406 reports
- CADD 21-2112-0100-51Icu Medical, Inc. · Pump, infusion, pca2 recalls59 reports
- CADD 21-2111-0402-78Icu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery3 recalls5 reports
- CADD 21-2112-0402-78Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION3 recalls0 reports
- CADD 21-2112-0401-51Icu Medical, Inc. · Pump, infusion, pca1 recall97 reports
- CADD 2110Icu Medical, Inc. · Pump, infusion, pca2 recalls1 report
- CADD 2110Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall2 reports
- CADD 21-2111-0402-249Icu Medical, Inc. · Pump, infusion, pca1 recall0 reports
- CADD 21-2112-0402-249Icu Medical, Inc. · Set, administration, intravascular1 recall0 reports
- CADD 21-2131-25Icu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery33 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96620).
- Z-1668-2025 · Open, ClassifiedCADD-Solis VIP Ambulatory Infusion Pump; Model/Number: 21-2120-0100-02 21-2120-0100-03 21-2120-0100-06 21-2120-0100-232 21-2120-0100-233 21-2120-0100-50 21-2120-0100-51 21-2120-0…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page