Bivona ZZ19JN80ZSC098NRecalls & reported problems
Icu Medical, Inc. · Class 2
All models of Bivona (469)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15019517209099), received between Feb 28, 2020 and Feb 28, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Leak/Splash1
- Material Integrity Problem1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 28, 2020Burst Container or Vessel; Material Integrity Problem; Burst Container or Vessel; Material Integrity Problem
IT WAS REPORTED THE PATIENT'S CUFF WAS TESTED BEFORE INSERTION AND NO LEAKAGE WAS NOTICED AT THE TIME. DURING FILLING IN THE EVENING, A LEAKAGE WAS NOTICED FOR THE FIRST TIME AND WATER LEAKED INTO THE TRACHEA. THIS WAS NOTICED ON (B)(6) 2020. PER THE REPORTER, A TUBE CHANGE OUT WAS NOT COMPLETED AT THAT TIME (THIS REPORTING DECISION WAS CAPTURED UNDER THE FIRST RELATED FILE, 3012307300-2020-01603). THEN ON (B)(6) 2020, THE CUFF OF THE SAME TRACHEOSTOMY (TRACH) TUBE AS BEFORE BURST WITHOUT ANY MANIPULATION. AT THIS POINT, AN EMERGENCY TRACH CHANGE WAS DONE. DUE TO THE PATIENT BEING UNDER VENTILATED AND DUE TO THE REPORTED LEAKAGE, THE REPORTER STATED THE PATIENT HAD TO GO TO THE EMERGENCY ROOM. IT WAS ALSO NOTED THAT 12 ML OF WATER ESCAPED FROM THE CUFF AND WENT INTO THE PATIENT. FURTHERMORE, THE REPORTER ADDED THAT THE PATIENT WAS UNABLE TO USE COUGH ASSIST, SO THE HEALTH CONDITION OF THE PATIENT DETERIORATED. ADDITIONALLY, THE REPORTER STATED AN INFUSION THERAPY AND FURTHER DIAGNOSTIC MEASURES HAD TO BE TAKEN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9771683 ↗InjuryFeb 28, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED THE PATIENT'S CUFF WAS TESTED BEFORE INSERTION AND NO LEAKAGE WAS NOTICED AT THE TIME. DURING FILLING IN THE EVENING, A LEAKAGE WAS NOTICED FOR THE FIRST TIME AND WATER LEAKED INTO THE TRACHEA. THIS WAS NOTICED ON (B)(6) 2020. PER THE REPORTER, A TUBE CHANGE OUT WAS NOT COMPLETED AT THAT TIME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9771678 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.
- Z-0260-2007 · Nov 17, 2006 · TerminatedCustom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026