Skip to content

Bivona ZZ19KN70ZZZ006NRecalls & reported problems

Icu Medical, Inc. · Class 2

All models of Bivona (469)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 15019517209983), received between Aug 10, 2022 and May 20, 2024 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Component Missing” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Missing2
  • Inflation Problem2
  • Defective Device1
  • Detachment of Device or Device Component1
  • Fitting Problem1
  • Loss of or Failure to Bond1
  • Material Puncture/Hole1
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Insufficient Information1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2024Detachment of Device or Device Component; Separation Problem; Detachment of Device or Device Component; Separation Problem

    IT WAS REPORTED THAT THE PILOT-BALLOON CAME LOSE DURING TRACH-CHANGE. THE RED PILOT-BALLON CAME LOOSE DURING UNCUFFING FOR CANNULA REPLACEMENT. "THE SURFACE OF THE END OF INFLATION LINE IS SLIPPERY AND THE SUPER SLICK IS NOT REMOVED IN THE ADHESIVE AREA. THERE ARE REMAINS OF GLUE ON/IN THE PILOT BALLOON.¿ ADVERSE PATIENT EFFECTS ARE UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 19355178 ↗
  • MalfunctionSep 21, 2022Component Missing; Component Missing

    IT WAS REPORTED THAT THE CUSTOMER PLACED A PURCHASE ORDER OF A PRODUCT WITH AN ANESTHESIA BAG. HOWEVER, IMMEDIATELY AFTER OPENING THE PACKAGE, THEY NOTICED NO BAG WAS ENCLOSED IN IT. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15462132 ↗
  • MalfunctionSep 21, 2022Component Missing; Component Missing

    IT WAS REPORTED THAT DURING A PRE-USE CHECK, THE ANESTHESIA BAG GOT DISCONNECTED FROM THE CONNECTOR. NO WHITE TAPE WAS WOUND OVER THE CONNECTOR. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15462126 ↗
  • MalfunctionSep 21, 2022Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT DURING A PRE-USE CHECK, THE CUSTOMER NOTICED THE AIR PRESSURE WITHIN THE CIRCUIT DID NOT REACH THE SPECIFIED VALUE. THEY SUSPECTED AIR LEAK OCCURRED IN THE PRODUCT. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15461080 ↗
  • MalfunctionSep 21, 2022Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT DURING A PRE-USE CHECK, THE CUSTOMER NOTICED AIR WAS LEAKING FROM THE PRODUCT. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15461073 ↗
  • MalfunctionSep 21, 2022Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT DURING A PRE-USE CHECK, THE CUSTOMER FOUND A TORN HOLE IN THE PRODUCT. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15461057 ↗
  • MalfunctionSep 13, 2022Fitting Problem; Fitting Problem

    IT WAS REPORTED THAT THE EYELET IS NOT REINFORCED AND THERE IS NO SUPERSLICK SURFACE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15410144 ↗
  • MalfunctionSep 13, 2022Defective Device; Defective Device

    IT WAS REPORTED THAT DURING A QUALITY CHECK, THE RED WING PILOT PORT DETACHED FROM THE HOSE. THE DEVICE WAS REPAIRED BY REMOVING THE SUPERSLICK AT THE GLUING AREA AND GLUING IT BACK ON WITH SILICONE DUE TO THE PATIENT NEEDING THE TUBES. NO PATIENT INJURY WAS REPORTED PHOTOS ATTACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15409675 ↗
  • MalfunctionAug 10, 2022Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT PRIOR TO USE, IT WAS NOTICED THAT THE NECK FLANGE IS NOT PROPERLY GLUED AND GAPS ARE ABLE TO BE VISUALIZED. CUSTOMER REPORTED THAT THE GAPS SEEM TO BE FILLED WITH "SUPERSLICK". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15204566 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.

  • Z-0260-2007 · Nov 17, 2006 · Terminated
    Custom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates