Bivona ZZ19KN70ZZZ006NRecalls & reported problems
Icu Medical, Inc. · Class 2
All models of Bivona (469)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 15019517209983), received between Aug 10, 2022 and May 20, 2024 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Component Missing” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K923878); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Component Missing2
- Inflation Problem2
- Defective Device1
- Detachment of Device or Device Component1
- Fitting Problem1
- Loss of or Failure to Bond1
- Material Puncture/Hole1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2024Detachment of Device or Device Component; Separation Problem; Detachment of Device or Device Component; Separation Problem
IT WAS REPORTED THAT THE PILOT-BALLOON CAME LOSE DURING TRACH-CHANGE. THE RED PILOT-BALLON CAME LOOSE DURING UNCUFFING FOR CANNULA REPLACEMENT. "THE SURFACE OF THE END OF INFLATION LINE IS SLIPPERY AND THE SUPER SLICK IS NOT REMOVED IN THE ADHESIVE AREA. THERE ARE REMAINS OF GLUE ON/IN THE PILOT BALLOON.¿ ADVERSE PATIENT EFFECTS ARE UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 19355178 ↗MalfunctionSep 21, 2022Component Missing; Component Missing
IT WAS REPORTED THAT THE CUSTOMER PLACED A PURCHASE ORDER OF A PRODUCT WITH AN ANESTHESIA BAG. HOWEVER, IMMEDIATELY AFTER OPENING THE PACKAGE, THEY NOTICED NO BAG WAS ENCLOSED IN IT. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15462132 ↗MalfunctionSep 21, 2022Component Missing; Component Missing
IT WAS REPORTED THAT DURING A PRE-USE CHECK, THE ANESTHESIA BAG GOT DISCONNECTED FROM THE CONNECTOR. NO WHITE TAPE WAS WOUND OVER THE CONNECTOR. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15462126 ↗MalfunctionSep 21, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING A PRE-USE CHECK, THE CUSTOMER NOTICED THE AIR PRESSURE WITHIN THE CIRCUIT DID NOT REACH THE SPECIFIED VALUE. THEY SUSPECTED AIR LEAK OCCURRED IN THE PRODUCT. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15461080 ↗MalfunctionSep 21, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING A PRE-USE CHECK, THE CUSTOMER NOTICED AIR WAS LEAKING FROM THE PRODUCT. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15461073 ↗MalfunctionSep 21, 2022Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT DURING A PRE-USE CHECK, THE CUSTOMER FOUND A TORN HOLE IN THE PRODUCT. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15461057 ↗MalfunctionSep 13, 2022Fitting Problem; Fitting Problem
IT WAS REPORTED THAT THE EYELET IS NOT REINFORCED AND THERE IS NO SUPERSLICK SURFACE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15410144 ↗MalfunctionSep 13, 2022Defective Device; Defective Device
IT WAS REPORTED THAT DURING A QUALITY CHECK, THE RED WING PILOT PORT DETACHED FROM THE HOSE. THE DEVICE WAS REPAIRED BY REMOVING THE SUPERSLICK AT THE GLUING AREA AND GLUING IT BACK ON WITH SILICONE DUE TO THE PATIENT NEEDING THE TUBES. NO PATIENT INJURY WAS REPORTED PHOTOS ATTACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15409675 ↗MalfunctionAug 10, 2022Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED THAT PRIOR TO USE, IT WAS NOTICED THAT THE NECK FLANGE IS NOT PROPERLY GLUED AND GAPS ARE ABLE TO BE VISUALIZED. CUSTOMER REPORTED THAT THE GAPS SEEM TO BE FILLED WITH "SUPERSLICK". NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15204566 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K923878. One submission often covers many models, so check the product and lot details in each record.
- Z-0260-2007 · Nov 17, 2006 · TerminatedCustom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page