Olympus MAJ-1606Recalls & reported problems

Keymed (Medical & Industrial Equipment) Limited · Class 2

Quick summary

This Olympus flushing pump is used with endoscopes to pump air and fluid during exams. The 26 reports mostly describe the pump failing to pump or prime, stalling, and leaking or dripping hoses, with no deaths and no injuries mentioned. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
26
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 15019778003528), received between Mar 8, 2024 and Mar 25, 2026 — classified as 26 malfunctions.

The most frequently reported specific device problem is “Pumping Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 4 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Pumping Problem5
  • Intermittent Loss of Power4
  • Crack3
  • Leak/Splash3
  • Failure to Pump2
  • Free or Unrestricted Flow2
  • Activation Failure1
  • Circuit Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions26

Recent adverse event reports

The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2026Failure to Prime

    IT WAS REPORTED THAT THE FLUSHING PUMP FAILED TO PRIME DURING A DIAGNOSTIC ENDOSCOPY PROCEDURE WHILE THE PATIENT WAS UNDER SEDATION AND THE DEVICE WAS INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24692794 ↗
  • MalfunctionMar 2, 2026Mechanical Jam

    IT WAS REPORTED THAT THE OLYMPUS FLUSHING PUMP HAD A MOTOR STALLED. THE EVENT OCCURRED DURING A PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24489300 ↗
  • MalfunctionJan 28, 2026Gas/Air Leak

    IT WAS REPORTED THE OLYMPUS FLUSHING PUMP HAD DRIBBLING FROM THE DISTAL END DURING A DIAGNOSTIC COLONOSCOPY. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24203829 ↗
  • MalfunctionJan 9, 2026Pumping Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24019091 ↗
  • MalfunctionNov 21, 2025Pumping Problem; Pumping Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23617332 ↗
  • MalfunctionOct 27, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THE SUBJECT DEVICE EXHIBITED THAT IRRIGATION HOSE HAS A LEAK IN IT. THE ISSUE OCCURRED DURING SET UP/ INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23394484 ↗
  • MalfunctionMay 16, 2025Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THE SUBJECT DEVICE PRESENTED A FLUSHING PUMP NOT FLUSHING. THE ISSUE OCCURRED DURING A DIAGNOSTIC COLONOSCOPY PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22039706 ↗
  • MalfunctionMay 15, 2025Pumping Problem; Pumping Problem

    THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22026629 ↗
  • MalfunctionOct 29, 2024Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THE IRRIGATION PUMP WAS NOT WORKING. THE ISSUE OCCURRED DURING A DIAGNOSTIC (COLONOSCOPY) PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20556296 ↗
  • MalfunctionOct 7, 2024Crack; Pumping Problem; Crack; Pumping Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED FAULTY PUMP HEAD AND FOUND FLOW CONTROL KNOB FRONT WAS DAMAGED WITH A CRACK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20391596 ↗
  • MalfunctionOct 7, 2024Degraded; Electrical /Electronic Property Problem; Difficult to Flush; Degraded; Electrical /Electronic Property Problem; Difficult to Flush

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLUSHING PUMP EXHIBITED FAULTY PUMP HEAD ASSY, ITS PLASTIC CAP WORN OUT AND POWER SWITCH AT FRONT FLICKERED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20388058 ↗
  • MalfunctionSep 5, 2024Crack; Activation Failure; Material Split, Cut or Torn; Intermittent Loss of Power; Crack; Activation Failure; Material Split, Cut or Torn; Intermittent Loss of Power

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE OLYMPUS FLUSHING PUMP EXHIBITED THAT THE UNIT FAILED TO ACTIVATE THE PUMP MOTOR WHEN FOOTSWITCH WAS PRESSED DUE FAULTY MAIN BOARD, THE POWER SWITCH LED FLICKERED, THE LED PLASTIC CAP HAS A CUT MARK, THE FRONT PANEL WAS DAMAGED WITH CRACK ON TOP LEFT CORNER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20149114 ↗
  • MalfunctionSep 5, 2024Intermittent Loss of Power; Intermittent Loss of Power

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE OLYMPUS FLUSHING PUMP EXHIBITED THAT THE POWER SWITCH LED FLICKERED AND FOUND WITH PLASTIC CAP WORN OUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20148117 ↗
  • MalfunctionSep 3, 2024Crack; Crack

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED THAT THE BOTTLE HOLDER WAS DAMAGED WITH A CRACK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20125432 ↗
  • MalfunctionAug 13, 2024Material Integrity Problem; Intermittent Loss of Power; Material Integrity Problem; Intermittent Loss of Power

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE OLYMPUS FLUSHING PUMP POWER SWITCH'S LIGHT EMITTING DIODE (LED) FLICKERED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19977934 ↗
  • MalfunctionJul 10, 2024Pumping Problem; Pumping Problem

    IT WAS REPORTED THAT, THE SUBJECT DEVICE HAD LACK OF PERFORMANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19711249 ↗
  • MalfunctionJun 26, 2024Intermittent Loss of Power; Intermittent Loss of Power

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLUSHING PUMP EXHIBITED POWER SWITCH LED FLICKERING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19615313 ↗
  • MalfunctionJun 14, 2024Failure to Pump; Failure to Pump

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLUSHING PUMP EXHIBITED A FAULTY PUMP HEAD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19537997 ↗
  • MalfunctionJun 10, 2024Free or Unrestricted Flow; Free or Unrestricted Flow

    IT WAS REPORTED THAT THE FLUSHING PUMP'S IRRIGATION CONTINUED WHEN NO BUTTON WAS PRESSED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19505029 ↗
  • MalfunctionJun 7, 2024Circuit Failure; No Flow; Power Problem; Circuit Failure; No Flow; Power Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLUSHING PUMP EXHIBITED A FAULTY PUMP HEAD ASSEMBLY AND THE POWER SWITCH'S LED FLICKERED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19492394 ↗
  • MalfunctionMay 13, 2024Failure to Pump; Failure to Pump

    IT WAS REPORTED, DURING PREPARATION FOR USE OF AN UNSPECIFIED PROCEDURE THE SUBJECT DEVICE WAS NOT PUSHING THE WATER OUT AND IT JUST MAKES NOISE LIKE IT IS WORKING BUT IT IS NOT WORKING. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19299667 ↗
  • MalfunctionApr 25, 2024Defective Device; Defective Device

    IT WAS REPORTED, THE FLUSHING PUMP HEAD WASN'T CLAMPING DOWN ON THE TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19188530 ↗
  • MalfunctionApr 8, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE FLUSHING PUMP HAD WATER INFILTRATION. THE EVENT WAS FOUND DURING AN UNKNOWN DIAGNOSTIC PROCEDURE. THERE ARE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19060060 ↗
  • MalfunctionMar 29, 2024Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THE SUBJECT DEVICE WAS NOT ACTIVATING WHEN PRESSING ON THE FOOT SWITCH. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19006223 ↗
  • MalfunctionMar 28, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED THE FLUSHING PUMP HAD A LEAK UNDER THE PUMP HEAD LID. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18994628 ↗

Other Pump, Air, Non-Manual, For Endoscope devices