Skip to content

Bivona 67SP050Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Bivona (469)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
19
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 15021312000740), received between Mar 1, 2019 and Jul 6, 2026 — classified as 1 injury, 18 malfunctions.

The most frequently reported specific device problem is “Contamination” (9 reports).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K083641, K914088); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2024Open, Classified

    Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

    Z-2615-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Contamination9
  • Inflation Problem4
  • Break1
  • Crack1
  • Defective Device1
  • Gas/Air Leak1
  • Material Separation1
  • Short Fill1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • Insufficient Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 6, 2026Gas/Air Leak

    IT WAS REPORTED THAT THE BIVONA TRACHEOSTOMY TUBE CUFF WAS LEAKING. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25716664 ↗
  • MalfunctionMay 10, 2023Contamination; Contamination

    IT WAS REPORTED THAT A PATIENT REPORTED A BLACK RESIDUE ON THE SEAM AT THE END OF THE CUFF FOR 8 CUFFS. THIS WAS FOUND WHEN THE TRACH WAS REMOVED FROM THE AIRWAY. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16901432 ↗
  • MalfunctionMar 28, 2023Contamination; Contamination

    IT WAS REPORTED THAT THE PATIENT REPORTED BLACK RESIDUE ON THE SEAM AT THE END OF THE CUFF FOR 8 CUFFS. THIS WAS FOUND WHEN THE TRACH WAS REMOVED FROM THE AIRWAY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16631630 ↗
  • MalfunctionMar 28, 2023Contamination; Contamination

    IT WAS REPORTED THAT THE PATIENT REPORTED BLACK RESIDUE ON THE SEAM AT THE END OF THE CUFF FOR 8 CUFFS. THIS WAS FOUND WHEN THE TRACH WAS REMOVED FROM THE AIRWAY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16630188 ↗
  • MalfunctionMar 28, 2023Contamination; Contamination

    IT WAS REPORTED THAT THE PATIENT REPORTED BLACK RESIDUE ON THE SEAM AT THE END OF THE CUFF FOR 8 CUFFS. THIS WAS FOUND WHEN THE TRACH WAS REMOVED FROM THE AIRWAY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16630137 ↗
  • MalfunctionMar 28, 2023Contamination; Contamination

    IT WAS REPORTED THAT THE PATIENT REPORTED BLACK RESIDUE ON THE SEAM AT THE END OF THE CUFF FOR 8 CUFFS. THIS WAS FOUND WHEN THE TRACH WAS REMOVED FROM THE AIRWAY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16630123 ↗
  • MalfunctionMar 28, 2023Contamination; Contamination

    IT WAS REPORTED THAT THE PATIENT REPORTED BLACK RESIDUE ON THE SEAM AT THE END OF THE CUFF FOR 8 CUFFS. THIS WAS FOUND WHEN THE TRACH WAS REMOVED FROM THE AIRWAY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16630006 ↗
  • MalfunctionMar 28, 2023Contamination; Contamination

    IT WAS REPORTED THAT THE. PATIENT REPORTED BLACK RESIDUE ON THE SEAM AT THE END OF THE CUFF FOR 8 CUFFS. THIS WAS FOUND WHEN THE TRACH WAS REMOVED FROM THE AIRWAY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16629955 ↗
  • MalfunctionMar 28, 2023Contamination; Contamination

    IT WAS REPORTED THAT THE PATIENT REPORTED BLACK RESIDUE ON THE SEAM AT THE END OF THE CUFF FOR 8 CUFFS. THIS WAS FOUND WHEN THE TRACH WAS REMOVED FROM THE AIRWAY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16629918 ↗
  • MalfunctionMar 28, 2023Contamination; Contamination

    IT WAS REPORTED THAT THE.PATIENT REPORTED BLACK RESIDUE ON THE SEAM AT THE END OF THE CUFF FOR 8 CUFFS. THIS WAS FOUND WHEN THE TRACH WAS REMOVED FROM THE AIRWAY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16629866 ↗
  • MalfunctionMar 14, 2023Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT DURING THE PRE-USE CHECK, THE TUBE DID NOT EXPAND PROPERLY DESPITE 10-15 ATTEMPTS WITH STERILE H20. ISSUE WAS DISCOVERED DURING PRE-TEST AND NOT USED ON PATIENT. ANOTHER TRACH SAME SIZE AND MAKE WAS USED AND TESTED OUT NORMALLY PRIOR TO USE.

    Patient problems: Insufficient Information

    MAUDE report 16539933 ↗
  • InjuryFeb 2, 2023Defective Device; Defective Device

    IT WAS REPORTED THAT THE TRACH TUBE WAS DEFECTIVE. THIS WAS FOUND AFTER A DAY OF USE. NO PATIENT HARM REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 16286045 ↗
  • MalfunctionSep 13, 2022Short Fill; Short Fill

    IT WAS REPORTED THAT DURING THE PRE-TEST OF THE DEVICE, THE CUFF WOULD NOT INFLATE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15410021 ↗
  • MalfunctionMar 4, 2022Crack; Crack

    IT WAS REPORTED THAT THE FLANGE INSPECTED ON SECOND TRACH AND A CRACK WAS NOTED ON IT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13667093 ↗
  • MalfunctionMar 4, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT WAS FOUND WITH TRACH IN HAND, CUFF SIDE FLANGE ATTACHED TO TRACH, TIES SEPARATED FROM STEM/BASE OF TRACH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13667035 ↗
  • MalfunctionFeb 8, 2022Break; Break

    IT WAS REPORTED THAT ONE OF THE PATIENTS TRACH TUBES BROKE AT THE SHAFT. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13480714 ↗
  • MalfunctionFeb 8, 2022Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT THE TRACHEOSTOMY DEVICE ONLY INFLATES ON ONE SIDE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13480547 ↗
  • MalfunctionMay 24, 2021Inflation Problem; Inflation Problem

    INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL TRACH WAS NOT INFLATING UPON OPENING THE PACKAGE. THERE WERE NO REPORTED ADVERSE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11873720 ↗
  • MalfunctionMar 1, 2019Inflation Problem; Inflation Problem

    INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL BIVONA TTS CUFFED PEDIATRIC WITH V NECK FLANGE TRACHEOSTOMY TUBE CUFF WOULD NOT INFLATE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8384138 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083641, K914088. One submission often covers many models, so check the product and lot details in each record.

  • Z-0205-2024 · Aug 22, 2023 · Open, Classified
    Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
  • Z-2946-2020 · Aug 28, 2020 · Terminated
    Bivona Mid-Range Aire-Cuf Adult Tracheostomy Tube This tube is intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed up to
  • Z-1465-2017 · Feb 9, 2017 · Terminated
    Bivona TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.
  • Z-0930-05 · Apr 26, 2005 · Terminated
    Bivona brand Pediatric TTS Tracheostomy Tube, 3.5 mm tracheostomy tubes with a 3.5mm obturator ; product code 67P035.
  • Z-0302-06 · Feb 17, 2005 · Terminated
    Bivona brand Neonatal TTS Tracheostomy Tubes, sterile, product code 67N040.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates