Bivona 67SP055Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Bivona (469)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 15021312000757), received between Dec 9, 2018 and Dec 8, 2025 — classified as 5 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K083641, K914088); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 29, 2024Open, Classified
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Z-2615-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Inflation Problem7
- Leak/Splash3
- Failure to Deliver2
- Fitting Problem2
- Air/Gas in Device1
- Break1
- Complete Blockage1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Known Impact Or Consequence To Patient3
- Insufficient Information2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 8, 2025Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED THAT THE BALLON WOULD NOT INFLATE AND ALSO STUCK ON THE SHAFT. THE EVENT OCCURRED UPON OPENING AT PATIENT'S HOME. THE OUTCOME OF THE EVENT WAS ON GOING AND AWAITING FOR REPLACEMENT. THE CHILD WAS VENT DEPENDENT AND GOT THE SAME RESULT WITH THIS TRACH AS THE PREVIOUS ONE WHICH WAS REGISTERED UNDER (B)(4) BUT STOPPED BEFORE INSTILLING MORE AIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23745619 ↗MalfunctionOct 31, 2023Complete Blockage; Inflation Problem; Complete Blockage; Inflation Problem
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE DID NOT INFLATE DURING TESTING. THERE WAS A BLOCKAGE BETWEEN THE OUTSIDE BALLOON AND THE INSIDE CUFF PREVENTING THE INFLATION ON THE INSIDE CUFF. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18042987 ↗MalfunctionFeb 14, 2023Failure to Deliver; Failure to Deliver
IT WAS REPORTED THAT THE TRACH TUBE WAS DEFECTIVE. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT, NO OBSERVABLE CLINICAL SYMPTOMS, OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16371746 ↗MalfunctionFeb 13, 2023Failure to Deliver; Failure to Deliver
IT WAS REPORTED THAT THE TRACH TUBE WAS DEFECTIVE. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16362999 ↗MalfunctionSep 19, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE CUFF WAS BEING TESTED WITH STERILE WATER IN CUFF PRIOR TO INSERTION AND ONLY ONE SIDE OF THE CUFF INFLATED. SURGEON PRESENT AT THE TIME AS WELL AND WITNESSED DEFECTIVE CUFF. CUFF WAS MANIPULATED, BUT THE OTHER SIDE STILL DID NOT OPEN UP. A NEW CUFF TRACH WAS OBTAINED WHICH WORKED WELL. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15446697 ↗MalfunctionSep 19, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING PRE-TEST THE CUFF INFLATED ASYMMETRICALLY. NO PATIENT INJURY OR INVOLVEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15446525 ↗MalfunctionSep 19, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE CUFF NOT EVENLY INFLATING. CUFF STICKING TO THE PLASTIC MOLDED CONTAINER. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15446447 ↗MalfunctionSep 15, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING THE PRE-TEST, THE CUFF INFLATED ASYMMETRICALLY. NO PATIENT INJURY OR INVOLVEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15427175 ↗MalfunctionAug 23, 2022Fitting Problem; Fitting Problem
IT WAS REPORTED THAT THE CUFF WAS STICKING TO THE SHAFT OF THE DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15280307 ↗MalfunctionAug 23, 2022Fitting Problem; Fitting Problem
IT WAS REPORTED THAT THE CUFF WAS STICKING TO THE SHAFT OF THE DEVICE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15280289 ↗MalfunctionAug 23, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT DURING PRE-TEST OF A ROUTINE TRACH CHANGE, TWO CUFFS INFLATED ASYMMETRICALLY. A THIRD TRACH WAS USED TO COMPLETED THE TRACH CHANGE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15277446 ↗InjurySep 23, 2019Inflation Problem; Inflation Problem
INFORMATION WAS RECEIVED INDICATING THAT THE CUFF TO A SMITHS MEDICAL PORTEX BIVONA FLEXTEND¿ TTS¿ TRACHEOSTOMY TUBE WAS OBSERVED TO BE DEFLATING. THE DEFECTIVE TRACH TUBES WERE REPORTED TO BE BROUGHT IN BY THE RESPIRATORY THERAPIST; INDICATING THEY HAD BEEN REMOVED. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9106919 ↗InjuryJul 2, 2019Break; Break
IT WAS REPORTED THAT PRIOR TO A ROUTINE TRACHEOSTOMY TUBE CHANGE OUT, THE CUFF HAD BURST, CAUSING THE PATIENT TO LOSE VENTILATION. THE PATIENT WAS TAKEN TO A HOSPITAL FOR A TUBE CHANGE OUT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8755896 ↗InjuryDec 9, 2018Leak/Splash; Leak/Splash
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PORTEX BIVONA TTS (TIGHT-TO-SHAFT) TRACHEOSTOMY TUBE CUFF LEAKED IMMEDIATELY UPON USE. SUBSEQUENTLY, A TRACHEOSTOMY (TRACH) CHANGE OUT WAS REQUIRED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8145356 ↗InjuryDec 9, 2018Leak/Splash; Leak/Splash
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PORTEX BIVONA TTS (TIGHT-TO-SHAFT) TRACHEOSTOMY TUBE CUFF LEAKED IMMEDIATELY UPON USE. SUBSEQUENTLY, A TRACHEOSTOMY (TRACH) CHANGE OUT WAS REQUIRED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8145353 ↗InjuryDec 9, 2018Leak/Splash; Leak/Splash
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PORTEX BIVONA TTS (TIGHT-TO-SHAFT) TRACHEOSTOMY TUBE CUFF LEAKED IMMEDIATELY UPON USE. SUBSEQUENTLY, A TRACHEOSTOMY (TRACH) CHANGE OUT WAS REQUIRED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8145350 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083641, K914088. One submission often covers many models, so check the product and lot details in each record.
- Z-0205-2024 · Aug 22, 2023 · Open, ClassifiedSmiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
- Z-2946-2020 · Aug 28, 2020 · TerminatedBivona Mid-Range Aire-Cuf Adult Tracheostomy Tube This tube is intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed up to
- Z-1465-2017 · Feb 9, 2017 · TerminatedBivona TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.
- Z-0930-05 · Apr 26, 2005 · TerminatedBivona brand Pediatric TTS Tracheostomy Tube, 3.5 mm tracheostomy tubes with a 3.5mm obturator ; product code 67P035.
- Z-0302-06 · Feb 17, 2005 · TerminatedBivona brand Neonatal TTS Tracheostomy Tubes, sterile, product code 67N040.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page