Bivona 60PFS50Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Bivona (469)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
6
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 15021312005882), received between Jul 10, 2017 and Dec 30, 2024 — classified as 6 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K083641, K944178); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2024Class IOpen, Classified

    Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

    Z-2599-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break4
  • Material Integrity Problem3
  • Material Split, Cut or Torn3
  • Gel Leak2
  • Device Damaged Prior to Use1
  • Insufficient Device Problem Information1
  • Labelling, Instructions for Use or Training Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient4
  • Insufficient Information1
  • No Information1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 30, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT WHILE THE TRACHEOSTOMY TUBE WAS IN THE CHILD'S TRACHEA, A SPLIT WAS NOTICED WHERE THE FLEXTEND PART OF THE TUBE MET THE UNIVERSAL CONNECTOR. THIS RISKED THE TRACHEOTOMY TUBE BECOMING RIPPED FROM THE UNIVERSAL CONNECTOR. THE OPERATOR OF THE DEVICE WAS A HEALTHCARE PROFESSIONAL. THE DEFECT WAS NOTICED DURING A ROUTINE DAILY CHECK OF THE CHILD'S TUBE WHILE IT WAS IN USE. UNPLANNED NON-EMERGENCY TUBE CHANGE WAS UNDERTAKEN TO REPLACE AFFECTED TUBE, SO THE EVENT IS RESOLVED AS A NEW TUBE WAS INSERTED. THIS INCIDENT OCCURRED IN THE COMMUNITY IN THE PATIENT'S HOME. THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21037119 ↗
  • InjuryNov 30, 2023Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information

    IT WAS REPORTED THAT DAMAGE OCCURRED TO A BIVONA TUBE WHILE INSERTED IN A CHILD. THE FAMILY HAD STERILISED AND REUSED THE TUBE AS PER MANUFACTURER GUIDANCE. THIS DAMAGE OCCURRED ON ITS THIRD USE. NO PHYSICAL HARM WAS CAUSED TO THE CHILD HOWEVER HE HAD TO UNDERGO AN UNPLANNED TUBE CHANGE OUTSIDE OF HIS USUAL TWENTY-EIGHT (28) DAY CHANGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18240741 ↗
  • MalfunctionApr 10, 2023Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE EYELETS AND THAT THE TWILL TAPES HAVE CUT THROUGH THE SILICONE. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 16710435 ↗
  • MalfunctionDec 30, 2022Device Damaged Prior to Use; Device Damaged Prior to Use

    IT WAS REPORTED THAT THE WIRE AND OR SPRING IS STICKING OUT ON THE END OF THE TRACHEOSTOMY. THE CUSTOMER DID NOT PLACE THE ITEM IN THEIR BODY BECAUSE THEY WERE CONCERNED ABOUT THE WIRE THAT WAS HANGING OUT. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16077735 ↗
  • MalfunctionOct 18, 2022Labelling, Instructions for Use or Training Problem; Labelling, Instructions for Use or Training Problem

    CUSTOMER REPORTS TWO PRODUCTS LABELED WITH NONEXISTENT EXPIRATION DATES. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15626933 ↗
  • InjuryMay 9, 2022Break; Break

    IT WAS REPORTED THAT WHILST IN USE IN A YOUNG PERSON, THE EYELET WAS DAMAGED, CLOSE TO FULLY SPLITTING WHILST ON THE TUBE'S 4TH USE. NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14342975 ↗
  • InjuryDec 17, 2020Break; Break

    INFORMATION WAS RECEIVED INDICATING THAT THE BIVONA TUBE SPLIT. THE ISSUE WAS DISCOVERED AFTER TEN DAYS IN USE AND A TUBE CHANGE OUT PROCEDURE WAS PERFORMED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11030404 ↗
  • MalfunctionOct 8, 2019Material Integrity Problem; Material Integrity Problem

    THE SURFACE ON A CANNULA OF A SMITHS MEDICAL TRACHEOSTOMY WAS REPORTED TO BE UNEVEN. NO ADVERSE EFFECTS REPORTED.

    Patient problems: No Information

    MAUDE report 9168554 ↗
  • InjurySep 19, 2018Gel Leak; Material Integrity Problem; Material Split, Cut or Torn; Gel Leak; Material Integrity Problem; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CARETAKER, HEARD AN AUDIBLE LEAK, FOLLOWED BY AN ALARM FROM THE VENTILATOR. THE CARETAKER IMMEDIATELY REMOVED THE FAULTY TRACH TUBE AND INSERTED A NEW ONE, ON THE VENTILATOR DEPENDENT CHILD. IT WAS REVEALED THAT THE SILICONE ON THE FAULTY TUBE HAD A SPLIT AT THE TOP OF THE SHAFT, EXPOSING THE COILED WIRE. THE TRACH TUBE HAD ONLY BEEN USED TWICE AT THIS POINT. NO FURTHER COMPLICATIONS WERE REPORTED IN CONNECTION WITH THIS INCIDENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7891752 ↗
  • InjurySep 19, 2018Gel Leak; Material Integrity Problem; Material Split, Cut or Torn; Gel Leak; Material Integrity Problem; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CARETAKER, HEARD AN AUDIBLE LEAK, FOLLOWED BY AN ALARM FROM THE VENTILATOR. THE CARETAKER IMMEDIATELY REMOVED THE FAULTY TRACH TUBE AND INSERTED A NEW ONE, ON THE VENTILATOR DEPENDENT CHILD. IT WAS REVEALED THAT THE SILICONE ON THE FAULTY TUBE HAD A SPLIT AT THE TOP OF THE SHAFT, EXPOSING THE COILED WIRE. THE TRACH TUBE HAD ONLY BEEN USED TWICE AT THIS POINT. NO FURTHER COMPLICATIONS WERE REPORTED IN CONNECTION WITH THIS INCIDENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7891751 ↗
  • MalfunctionJul 10, 2017Leak/Splash; Split; Leak/Splash; Split

    IT WAS REPORTED THAT THE SILICONE ON A PORTEX BIVONA UNCUFFED PEDIATRIC FLEXTEND PLUS SILICONE TRACHEOSTOMY TUBE SPLIT AT THE TOP OF THE SHAFT, AND THE COIL WIRE WAS EXPOSED. THE TRACHEOSTOMY TUBE WAS IN USE WHEN A STAFF MEMBER HEARD AN AUDIBLE LEAK AND THE VENTILATOR ALARMED. THE TRACHEOSTOMY TUBE WAS REMOVED AND A NEW ONE WAS PLACED IMMEDIATELY. IT WAS AFTER REMOVAL THAT THE SPLIT IN THE SILICONE WAS OBSERVED. IT WAS NOTED THAT THE TRACHEOSTOMY TUBE HAD "ONLY BEEN USED TWICE". NO INJURY WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6697687 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083641, K944178. One submission often covers many models, so check the product and lot details in each record.

  • Z-0205-2024 · Aug 22, 2023 · Open, Classified
    Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
  • Z-1145-2021 · Jan 18, 2021 · Terminated
    Tube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck Flange 4.5mm - Product Us
  • Z-0726-2021 · Dec 15, 2020 · Terminated
    PORTEX Bivona Cuffless Neonatal/Pediatric FlexTend Tracheostomy Tube, REF 60PFSS45
  • Z-2946-2020 · Aug 28, 2020 · Terminated
    Bivona Mid-Range Aire-Cuf Adult Tracheostomy Tube This tube is intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed up to
  • Z-1465-2017 · Feb 9, 2017 · Terminated
    Bivona TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →