Skip to content

Bivona 65P035Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Bivona (469)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15021312005967), received between Nov 14, 2023 and Nov 14, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2024Open, Classified

    Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

    Z-2604-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn1
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 14, 2023Material Split, Cut or Torn; Separation Problem; Material Split, Cut or Torn; Separation Problem

    IT WAS REPORTED THAT DURING PATIENT USE THE TRACHEOSTOMY TUBE SPLIT. IT WAS REMOVED FROM THE PATIENT AND NEW TRACHEOSTOMY INSERTED. THERE WAS NO PATIENT OR CLINICAL INJURY AND NO MEDICAL INTERVENTION OCCURRED AS THE PARENT CHANGED THE TRACHEOSTOMY. THE OUTCOME OF EVENT WAS NEW TRACHEOSTOMY TUBE IN FULL WORKING ORDER. THE PRE-EXISTING CONDITIONS WAS PIERRE ROBIN SYNDROME. HOME VENTILATED VIA TRACHEOSTOMY 24 HOURS/DAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18136152 ↗

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page