Bivona 750170Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Bivona (469)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 15021312006414), received between Aug 22, 2018 and Apr 4, 2025 — classified as 9 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K083641, K914088); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 29, 2024Open, Classified
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Z-2589-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Inflation Problem7
- Defective Device3
- Defective Component2
- Fluid/Blood Leak2
- Gas/Air Leak2
- Break1
- Deflation Problem1
- Failure to Deflate1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Respiratory Insufficiency4
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- Unspecified Respiratory Problem2
- Airway Obstruction1
- Ventilator Dependent1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 4, 2025Deflation Problem; Gas/Air Leak; Deflation Problem; Gas/Air Leak
IT WAS REPORTED THAT THE CUFF HAD AIR LEAKING ISSUES, AND MORE AIR WAS SUPPLIED. THERE WAS PATIENT INVOLVEMENT BUT UNKNOWN PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21774602 ↗MalfunctionSep 3, 2024Leak/Splash; Gas/Air Leak; Leak/Splash; Gas/Air Leak
IT WAS REPORTED THAT THE DEVICE HAD A CUFF LEAK AFTER FOUR DAYS OF USE. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20127789 ↗MalfunctionMay 11, 2023Defective Device; Defective Device
IT WAS REPORTED, THAT THERE WAS A RESPIRATORY ISSUE. TUBE INSERTED ON (B)(6) 2022 AND REMOVED ON (B)(6) 2022. SAME PATIENT AS COMPLAINT FILES (B)(6). NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16915817 ↗InjuryApr 19, 2023Defective Device; Defective Device
IT WAS REPORTED THAT THE PATIENT COULD NOT PROPERLY BREATH WITH THE DEVICE. NO MEDICAL OR SURGICAL INTERVENTION WAS REPORTED.
Patient problems: Airway Obstruction
MAUDE report 16766517 ↗MalfunctionApr 17, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE CUFF WAS NOT INFLATING DURING PRE-TEST. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16752326 ↗MalfunctionApr 14, 2023Defective Device; Defective Device
IT WAS REPORTED THAT THERE WAS A RESPIRATORY ISSUE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16742421 ↗MalfunctionOct 17, 2022Defective Component; Defective Component
IT WAS REPORTED THAT THE DEVICE CUFF IS FAULTY IN THE SET. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15620459 ↗MalfunctionSep 26, 2022Failure to Deflate; Failure to Deflate
IT WAS REPORTED THAT THERE WAS NO DEFECTION THAT WAS DETECTED DURING PRE-TEST, HOWEVER, CUFF CONSTANTLY REQUIRED PUMPING WHEN PLACED INSIDE PATIENT. PATIENT REQUIRES VENTILATOR. DEFECTIVE PRODUCT WAS REPLACED.
Patient problems: Ventilator Dependent
MAUDE report 15487957 ↗MalfunctionSep 13, 2022Volume Accuracy Problem; Volume Accuracy Problem
IT WAS REPORTED THAT DEVICE'S BALLOON WAS INFLATING ASYMMETRICALLY. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15415070 ↗InjuryMar 9, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT WHILE THE PATIENT'S SPOUSE WAS ATTEMPTING TO CHANGE A CANNULA BUT, THE SPOUSE NOTED THE REPLACEMENT CANNULA HAD A DYSFUNCTIONAL CUFF (ASYMMETRIC CUFF AND NOT ALLOWING ADEQUATE VENTILATION).THEY THEREFORE TOOK THE LAST CANNULA (WITH NO ANOMALY) AND WAS ABLE TO MAKE A CHANGE OF CANNULA, RESOLVING THE REPORTED EVENT.
Patient problems: Unspecified Respiratory Problem
MAUDE report 13710677 ↗InjuryMar 9, 2022Break; Break
IT WAS REPORTED THAT DURING THE NIGHT, THE TRACHEAL CANNULA CUFF SUDDENLY RUPTURED ALONG ITS ENTIRE LENGTH (THIS PATIENT HAVING NO POSSIBLE MOBILITY AND NO RESPIRATORY CAPACITY TO ACHIEVE HYPERPRESSURE). THE PATIENT'S SPOUSE WAS ABLE TO EVENTUALLY REPLACE THE CANNULA.
Patient problems: Unspecified Respiratory Problem
MAUDE report 13710649 ↗InjuryFeb 9, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT AN ASYMMETRIC CUFF WAS OBSERVED WITH THE TRACHEOSTOMY TUBE ALONG WITH WATER BEING PRESENT. ISSUE OCCURRED WHILE IN USE WITH PATIENT. PATIENT HAD RESPIRATORY DIFFICULTIES. AIRWAY WAS MAINTAINED BY CHANGING THE TRACHEOSTOMY WHICH RESOLVED THE ISSUE.
Patient problems: Respiratory Insufficiency
MAUDE report 13501530 ↗InjuryFeb 9, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT AN ASYMMETRIC CUFF WAS OBSERVED WITH THE TRACHEOSTOMY TUBE ALONG WITH WATER BEING PRESENT. ISSUE OCCURRED WHILE IN USE WITH PATIENT. PATIENT HAD RESPIRATORY DIFFICULTIES. AIRWAY WAS MAINTAINED BY CHANGING THE TRACHEOSTOMY WHICH RESOLVED THE ISSUE.
Patient problems: Respiratory Insufficiency
MAUDE report 13501276 ↗InjuryFeb 9, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT AN ASYMMETRIC CUFF WAS OBSERVED WITH THE TRACHEOSTOMY TUBE ALONG WITH WATER BEING PRESENT. ISSUE OCCURRED WHILE IN USE WITH PATIENT. PATIENT HAD RESPIRATORY DIFFICULTIES. AIRWAY WAS MAINTAINED BY CHANGING THE TRACHEOSTOMY WHICH RESOLVED THE ISSUE.
Patient problems: Respiratory Insufficiency
MAUDE report 13500854 ↗InjuryFeb 9, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT AN ASYMMETRIC CUFF WAS OBSERVED WITH THE TRACHEOSTOMY TUBE ALONG WITH WATER BEING PRESENT. ISSUE OCCURRED WHILE IN USE WITH PATIENT. PATIENT HAD RESPIRATORY DIFFICULTIES. AIRWAY WAS MAINTAINED BY CHANGING THE TRACHEOSTOMY WHICH RESOLVED THE ISSUE.
Patient problems: Respiratory Insufficiency
MAUDE report 13500579 ↗InjuryAug 24, 2020Fluid/Blood Leak; Fluid/Blood Leak
INFORMATION RECEIVED A SMITHS MEDICAL TRACHEOSTOMY/SILICONE - BIVONA TUBES ADULT AIRE-CUFF MALFUNCTIONED WHEN PLACED IN THE PATIENT, TWO TRACH CUFFS WOULD NOT INFLATE. NO PATIENT ADVERSE EVENTS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10442967 ↗InjuryAug 24, 2020Fluid/Blood Leak; Fluid/Blood Leak
INFORMATION RECEIVED A SMITHS MEDICAL TRACHEOSTOMY/SILICONE - BIVONA TUBES ADULT AIRE-CUFF MALFUNCTIONED WHEN PLACED IN THE PATIENT, TWO TRACH CUFFS WOULD NOT INFLATE. NO PATIENT ADVERSE EVENTS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10442941 ↗MalfunctionAug 23, 2018Inflation Problem; Inflation Problem
INFORMATION WAS RECEIVED INDICATING THAT THE BALLOON TO A SMITHS MEDICAL PORTEX® BIVONA® AIRE-CUF® ADULT TRACHEOSTOMY TUBE WOULD DEFLATE. THE BALLOON WAS REQUIRED TO BE INFLATED EVERY TWO HOURS WITH 1-2MLS OF AIR. IT WAS REPORTED THAT THE PATIENT WAS UNDER VENTILATED FOR 22 HOURS. THE CANNULA WAS REPORTED TO BE CHANGED EVERY 15 DAYS. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7808917 ↗MalfunctionAug 22, 2018Fitting Problem; Defective Component; Material Deformation; Fitting Problem; Defective Component; Material Deformation
INFORMATION WAS RECEIVED INDICATING THAT THE SPEAKING VALVE WAS UNABLE TO CONNECT TO A SMITHS MEDICAL PORTEX® BIVONA® AIRE-CUF® TRACHEOSTOMY TUBE. IT WAS NOTED THAT THE 15MM CONNECTOR IS OUT OF SPECIFICATION. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7805216 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083641, K914088. One submission often covers many models, so check the product and lot details in each record.
- Z-0205-2024 · Aug 22, 2023 · Open, ClassifiedSmiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
- Z-2946-2020 · Aug 28, 2020 · TerminatedBivona Mid-Range Aire-Cuf Adult Tracheostomy Tube This tube is intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed up to
- Z-1465-2017 · Feb 9, 2017 · TerminatedBivona TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.
- Z-0930-05 · Apr 26, 2005 · TerminatedBivona brand Pediatric TTS Tracheostomy Tube, 3.5 mm tracheostomy tubes with a 3.5mm obturator ; product code 67P035.
- Z-0302-06 · Feb 17, 2005 · TerminatedBivona brand Neonatal TTS Tracheostomy Tubes, sterile, product code 67N040.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page