Skip to content

Bivona 755190Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Bivona (469)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 15021312006490), received between Apr 9, 2021 and Apr 7, 2026 — classified as 5 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (4 reports).

5 other recalls cover devices cleared under the same FDA submission (K083641, K914088); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash4
  • Gas/Air Leak3
  • Break1
  • Disconnection1
  • Inflation Problem1
  • Material Deformation1

Patient problems

  • Insufficient Information5
  • No Clinical Signs, Symptoms or Conditions2
  • Airway Obstruction1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 7, 2026Break; Disconnection; Gas/Air Leak

    IT WAS STATED THAT THE INFLATION LINE IS BROKEN AND DISCONNECTED. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24800704 ↗
  • MalfunctionSep 4, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THE CUFF OF THE CANNULA LOSES PRESSURE FROM 30 TO 10 WITHOUT ANY EXTERNAL INFLUENCE. THE CUFF LOSES PRESSURE. WITHOUT SPECIAL INCIDENTS, THE CUFF LOSES ITS EFFECT. IT HAD TO BE CONSTANTLY RE-BLOCKED. IN ORDER TO SOLVE THE ISSUE, A NEW CANNULA WAS USED. THERE WAS PATIENT INVOLVEMENT AND NO HUMAN HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20140602 ↗
  • MalfunctionSep 13, 2022Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE CUFF OF THE CANNULA IS DEFORMED. PATIENT HAS IRRITATED SKIN. ADDITIONAL INFORMATION WAS RECEIVED STATING THE LOT NUMBER IS UNKNOWN.

    Patient problems: Skin Inflammation/ Irritation

    MAUDE report 15414246 ↗
  • MalfunctionAug 3, 2021Inflation Problem; Inflation Problem

    INFORMATION RECEIVED A SMITH MEDICAL TRACHEOSTOMY/SILICONE - BIVONA TUBES ADULT AIRE-CUFF FAILED WHILE IN USE. NO INFORMATION ON PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 12269453 ↗
  • InjuryJul 30, 2021Leak/Splash; Leak/Splash

    INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL TRACHEOSTOMY TUBE FAILED DUE TO CUFF LEAKS. THERE WERE NO REPORTED ADVERSE EVENTS.

    Patient problems: Insufficient Information

    MAUDE report 12255024 ↗
  • InjuryJul 30, 2021Leak/Splash; Leak/Splash

    INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL TRACHEOSTOMY TUBE FAILED DUE TO CUFF LEAKS. THERE WERE NO REPORTED ADVERSE EVENTS.

    Patient problems: Insufficient Information

    MAUDE report 12255021 ↗
  • InjuryJul 30, 2021Leak/Splash; Leak/Splash

    INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL TRACHEOSTOMY TUBE FAILED DUE TO CUFF LEAKS. THERE WERE NO REPORTED ADVERSE EVENTS.

    Patient problems: Insufficient Information

    MAUDE report 12254997 ↗
  • InjuryJul 30, 2021Leak/Splash; Leak/Splash

    INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL TRACHEOSTOMY TUBE FAILED DUE TO CUFF LEAKS. THERE WERE NO REPORTED ADVERSE EVENTS.

    Patient problems: Insufficient Information

    MAUDE report 12253119 ↗
  • InjuryApr 9, 2021Gas/Air Leak; Gas/Air Leak

    INFORMATION RECEIVED A SMITHS MEDICAL TRACHEOSTOMY/SILICONE: BIVONA TUBES ADULT AIRE-CUFF WAS LEAKING. THE REPORT INDICATED ONGOING ISSUES WITH CUFF LEAKING WITH MULTIPLE FACTORS AND NOT EXCLUSIVE TO THE FAULT OF TUB. THE CUFF WAS REPORTED TO BE INSPECTED AND CHECKED PRIOR TO INSERTION. MEDICAL INTERVENTION OF A TRACHEOSTOMY CHANGE OUT WAS DONE.

    Patient problems: Airway Obstruction

    MAUDE report 11645629 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083641, K914088. One submission often covers many models, so check the product and lot details in each record.

  • Z-0205-2024 · Aug 22, 2023 · Open, Classified
    Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
  • Z-2946-2020 · Aug 28, 2020 · Terminated
    Bivona Mid-Range Aire-Cuf Adult Tracheostomy Tube This tube is intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed up to
  • Z-1465-2017 · Feb 9, 2017 · Terminated
    Bivona TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.
  • Z-0930-05 · Apr 26, 2005 · Terminated
    Bivona brand Pediatric TTS Tracheostomy Tube, 3.5 mm tracheostomy tubes with a 3.5mm obturator ; product code 67P035.
  • Z-0302-06 · Feb 17, 2005 · Terminated
    Bivona brand Neonatal TTS Tracheostomy Tubes, sterile, product code 67N040.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page