Skip to content

Bivona 67HA70Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Bivona (469)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 15021312013801), received between Apr 5, 2017 and Aug 10, 2026 — classified as 2 injuries, 11 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K081440); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak4
  • Leak/Splash3
  • Defective Component1
  • Device Markings/Labelling Problem1
  • Disconnection1
  • Fluid/Blood Leak1
  • Inflation Problem1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • No Consequences Or Impact To Patient3
  • Pulmonary Dysfunction1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 10, 2026Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50193513 ↗
  • MalfunctionJun 25, 2026Gas/Air Leak

    IT WAS REPORTED THAT DURING USE, THE LEAKAGE WAS OBSERVED ON THE CUFF. THIS ISSUE MAY RESULT IN INADEQUATE VENTILATION TO THE PATIENT AND COULD POTENTIALLY LEAD TO SERIOUS INJURY. THERE WAS PATIENT INVOLVEMENT, HOWEVER NO HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25625767 ↗
  • MalfunctionApr 6, 2026Gas/Air Leak

    IT WAS REPORTED THAT THE FIRST TRACHEOSTOMY TUBE WAS PLACED AND FOUND LEAKAGE AFTER 4 DAYS OF USE. THEN THE SECOND ONE WAS PLACED AND FOUND LEAKAGE ON THE 1ST DAYS OF USE. THE EVENT OCCURRED WHILE IN USE WITH PATIENT AT THE HOSPITAL. THERE WAS NO DELAY IN THERAPY THAT WAS CRITICAL TO THIS PATIENT. THIS PROBLEM WAS SOLVED BY REPLACING WITH A NEW ONE. THERE WAS A PATIENT INVOLVEMENT AND THERE WAS UNKNOWN PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791514 ↗
  • MalfunctionApr 4, 2025Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT AFTER FOUR HOURS REPLACING THE DEVICE, THE RESIDUAL AMOUNT OF WATER IN THE CUFF WAS 1ML. THIS PROBLEM WAS RESOLVED BY REPLACING WITH ANOTHER BRAND PVC TRACHEOSTOMY TUBE. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THERE WAS UNKNOWN PATIENT HARM OR ADVERSE EVENT REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21778499 ↗
  • MalfunctionDec 12, 2024Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole

    IT WAS REPORTED THAT LEAKAGE WAS FOUND ON THE INFLATION LINE DURING THE PRE-TEST. THIS PROBLEM WAS RESOLVED BY REPLACING THE DEVICE WITH A NEW ONE. THE EVENT OCCURRED IN THE HOSPITAL AND THE OPERATOR WAS A HEALTH PROFESSIONAL. NO MEDICAL INTERVENTION WAS REQUIRED. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20914252 ↗
  • MalfunctionAug 14, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT LEAKAGE WAS FOUND WHEN THE TRACHEOSTOMY TUBE WAS IN USE WITH PATIENT ON THE 16TH DAY OF USE. THIS PROBLEM WAS SOLVED BY REPLACING WITH A NEW BIVONA 67HA70. THE OPERATOR OF THE DEVICE WAS A HEALTH PROFESSIONAL. THERE WAS PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19982589 ↗
  • MalfunctionJun 13, 2023Defective Component; Defective Component

    IT WAS REPORTED, THAT THE DEVICE WAS FOUND WITHOUT A HOLE FOR SECURING THE LOCKING PIN. NO PATIENT HARM OR INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17121104 ↗
  • MalfunctionOct 13, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT UPON OPENING THE PRODUCT, THE NURSE READ THE PRODUCT INSTRUCTION FOR USE (IFU) AND COMPARED THE ENGLISH TEXT TO THE FINNISH TEXT AND NOTICED THAT THE TRANSLATION WAS NOT CORRECT, OBSERVED WITH FOUR LOT NUMBERS. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15598099 ↗
  • InjuryAug 19, 2022Disconnection; Disconnection

    IT WAS REPORTED THAT THE RING WAS WELL TIGHTENED BUT IT WAS SLIPPING WHICH MADE THE TRACHEOSTOMY COME OUT. THE PATIENT HAS A PULMONARY PATHOLOGY THAT HAS LED TO A FAILURE OF INTUBATION AND THEREFORE TO AN EMERGENCY TRACHEOSTOMY. PATIENT HAD TO BE SEDATED OVER THE LONG TERM TO AVOID DECANULATION. HAVING A GRANULOMA IN THE TRACHEA, WHEN THE TRACHEOSTOMY CAME OUT, IT CAME UP AGAINST THE GRANULOMA PREVENTING ANY VENTILATION. PATIENT HAD SEVERAL FIBROSCOPIES TO URGENTLY PUT THE CANNULA BACK IN PLACE. THE CANNULA HAD TO BE MAINTAINED BY ADHESIVES.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Pulmonary Dysfunction

    MAUDE report 15262625 ↗
  • MalfunctionJan 5, 2022Premature Separation; Premature Separation

    IT WAS REPORTED THAT THE PILOT BALLOON WAS FOUND DETACHED AFTER 6 DAYS OF USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13171396 ↗
  • MalfunctionJan 15, 2021Leak/Splash

    INFORMATION WAS RECEIVED THAT DEVICE CUFF WAS FOUND LEAKING AFTER TWO DAYS IN USE. A SECOND CUFF WAS INSERTED AND AFTER TWO DAYS USE, THE CUFF LEAKED AGAIN. NO PATIENT INJURY OCCURRED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11177036 ↗
  • InjuryDec 30, 2020Leak/Splash; Leak/Splash

    INFORMATION RECEIVED A SMITHS MEDICAL TRACHEOSTOMY/SILICONE - BIVONA TUBES ADULT HYPERFLEX TTS TUBE LEAKED AFTER TWO DAYS IN USE. A SECOND CUFF WAS INSERTED AND AFTER TWO DAYS IN USE THE SECOND CUFF WAS INSERTED LEAKED AGAIN. NO PATIENT ADVERSE EVENTS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11098798 ↗
  • MalfunctionApr 5, 2017Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE BIVONA® ADULT TTS¿ ADJUSTABLE NECK FLANGE HYPERFLEX¿ TRACHEOSTOMY TUBE WAS IN PATIENT USE FOR ONE WEEK WHEN IT WAS OBSERVED THAT THE CUFF WAS BROKEN. THE REPORTED DEVICE WAS REPLACED. THE EVENT WAS OBSERVED SPONTANEOUSLY BY HOSPITAL PERSONNEL DURING THE INITIAL USE OF THE DEVICE. ACCORDING TO THE REPORT, THE CUFF PATENCY WAS TESTED PRIOR TO PATIENT USE AND THE CUFF WAS FILLED WITH STERILE WATER. NO ADVERSE EFFECT TO PATIENT WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6461865 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081440. One submission often covers many models, so check the product and lot details in each record.

  • Z-2485-2020 · May 26, 2020 · Terminated
    Pediatric Adjustable Flange Tracheostomy Tube Cuffless

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →