Bivona 60NFP30Recalls & reported problems
Icu Medical, Inc. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15021312014723), received between May 25, 2018 and May 2, 2023 — classified as 3 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K083641, K944178); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- May 29, 2024Open, Classified
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Z-2592-2024 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Kinked1
Patient problems
- Insufficient Information2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 2, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE DEVICE HAD JUST BEEN CHANGED THE DAY BEFORE. THE DEVICE WAS ON IT'S THIRD USE. THE DEVICE LOOKED TO BE INTACT ON INSPECTION PRIOR TO INSERTION, AND IT WAS FOUND DAMAGED WITHIN 24 HOURS. THE DEVICE WAS CHANGED AGAIN AS SOON AS THE DAMAGE WAS DISCOVERED. A PICTURE WAS SENT IN FROM THE CUSTOMER SHOWING POSSIBLE CHIPPING OF THE TRACH TIES. EXACT "DAMAGE" HAS NOT BEE COMMUNICATED.
Patient problems: Insufficient Information
MAUDE report 16857746 ↗InjuryMay 2, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE CUSTOMER REPORTED A DAMAGE TRACHEOSTOMY TUBE. THE TRACH HAD BEEN STERILIZED ONCE PREVIOUSLY AND WAS BEING CHANGED TO THE END OF ITS SECOND USE. HAD BEEN IN PLACE FOR ONE MONTH WHEN THE DAMAGE WAS NOTICED BY THE PATIENT'S MOTHER. THE DEVICE WAS CHANGED TO A TRACH THAT HAD BEEN STERILIZED TWICE, THIS WAS IT'S THIRD USE. THE TRACH WAS CHECKED PRIOR TO INSERTION AND FOUND TO BE SUFFICIENT FOR PLACEMENT. A PICTURE WAS SENT IN FROM THE CUSTOMER SHOWING POSSIBLE CHIPPING OF THE TRACH TIES. EXACT "DAMAGE" HAS NOT BEE COMMUNICATED.
Patient problems: Insufficient Information
MAUDE report 16856098 ↗InjuryMay 25, 2018Kinked; Kinked
IT WAS REPORTED THAT THE FLEXTEND PORTION OF THE TRACHEOSTOMY TUBE WAS FOUND TO BE KINKED. THE FAULT WAS NOTICED WHEN THERE WAS INCREASED RESISTANCE DURING SUCTIONING. ONCE THE TUBE WAS REMOVED, A "CLEAR PINCH" WAS NOTICED. THE TRACHEOSTOMY TUBE WAS REPLACED. NO PERMANENT INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7546707 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083641, K944178. One submission often covers many models, so check the product and lot details in each record.
- Z-0205-2024 · Aug 22, 2023 · Open, ClassifiedSmiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
- Z-1145-2021 · Jan 18, 2021 · TerminatedTube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck Flange 4.5mm - Product Us
- Z-0726-2021 · Dec 15, 2020 · TerminatedPORTEX Bivona Cuffless Neonatal/Pediatric FlexTend Tracheostomy Tube, REF 60PFSS45
- Z-2946-2020 · Aug 28, 2020 · TerminatedBivona Mid-Range Aire-Cuf Adult Tracheostomy Tube This tube is intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed up to
- Z-1465-2017 · Feb 9, 2017 · TerminatedBivona TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.