Skip to content

Bivona 60NFP35Recalls & reported problems

Icu Medical, Inc. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15021312014730), received between Mar 15, 2021 and Mar 16, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Suction Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K083641, K944178); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. May 29, 2024Open, Classified

    Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

    Z-2592-2024 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Suction Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 16, 2021Suction Problem; Suction Problem

    IT WAS REPORTED THAT THERE WAS A TINY HOLE IN THE BOTTOM OF THE FLEXTEND PORTION OF THE BIVONA TRACHEOSTOMY TUBE. THIS WAS NEAR THE OUTER HUB, AND WAS EXPOSING THE VIEWING OF THE WIRE IN THE FLEXTEND PORTION. THIS PRODUCT FAULT WAS NOTICED AFTER A WHISTLING SOUND WAS HEARD COMING FROM THE CHILD. THE VENTILATOR, WHICH WAS ALSO ON, WAS ALARMING. THE PATIENT'S MOTHER TRIED TO SUCTION THE CHILD'S TUBE BUT SHE WAS UNABLE TO INSERT THE SUCTION CATHETER TO THE NEEDED SUCTION DEPTH. THE MOTHER THOUGHT THAT THERE WAS A BLOCKAGE. AFTER SUSPECTING THIS, THE MOTHER PERFORMED A TRACHEOSTOMY TUBE CHANGE. NO PATIENT INJURY, OR FURTHER COMPLICATIONS WERE REPORTED IN RELATION TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11494541 ↗
  • MalfunctionMar 15, 2021Suction Problem; Suction Problem

    IT WAS REPORTED THAT THERE WAS A TINY HOLE IN THE BOTTOM OF THE FLEXTEND PORTION OF THE BIVONA TRACHEOSTOMY TUBE. THIS WAS NEAR THE OUTER HUB, AND WAS EXPOSING THE VIEWING OF THE WIRE IN THE FLEXTEND PORTION. THIS PRODUCT FAULT WAS NOTICED AFTER A WHISTLING SOUND WAS HEARD COMING FROM THE CHILD. THE VENTILATOR, WHICH WAS ALSO ON, WAS ALARMING. THE PATIENT'S MOTHER TRIED TO SUCTION THE CHILD'S TUBE BUT SHE WAS UNABLE TO INSERT THE SUCTION CATHETER TO THE NEEDED SUCTION DEPTH. THE MOTHER THOUGHT THAT THERE WAS A BLOCKAGE. AFTER SUSPECTING THIS, THE MOTHER PERFORMED A TRACHEOSTOMY TUBE CHANGE. NO PATIENT INJURY, OR FURTHER COMPLICATIONS WERE REPORTED IN RELATION TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11488493 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083641, K944178. One submission often covers many models, so check the product and lot details in each record.

  • Z-0205-2024 · Aug 22, 2023 · Open, Classified
    Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
  • Z-1145-2021 · Jan 18, 2021 · Terminated
    Tube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck Flange 4.5mm - Product Us
  • Z-0726-2021 · Dec 15, 2020 · Terminated
    PORTEX Bivona Cuffless Neonatal/Pediatric FlexTend Tracheostomy Tube, REF 60PFSS45
  • Z-2946-2020 · Aug 28, 2020 · Terminated
    Bivona Mid-Range Aire-Cuf Adult Tracheostomy Tube This tube is intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed up to
  • Z-1465-2017 · Feb 9, 2017 · Terminated
    Bivona TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.