Bivona 60NFP40Recalls & reported problems
Icu Medical, Inc. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
These ICU Medical Bivona tracheostomy tubes are used to help keep the airway open. The 25 reports mostly describe problems disconnecting the tube, along with splitting, tearing, breaking, or suction resistance; there were no deaths and 15 injuries. One recall is open and began in May 2024 because the securement flange in specific lots may tear due to a manufacturing defect. Reports fell from 3 in the previous 12 months to 0 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 15
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 15021312014747), received between Apr 30, 2018 and Jul 10, 2025 — classified as 15 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Failure to Disconnect” (16 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K083641, K944178); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 29, 2024Open, Classified
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Z-2592-2024 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Failure to Disconnect16
- Material Split, Cut or Torn5
- Material Discolored2
- Break1
- Component Missing1
- Fracture1
- Material Puncture/Hole1
- Physical Resistance/Sticking1
Patient problems
- No Consequences Or Impact To Patient16
- No Clinical Signs, Symptoms or Conditions6
- Discomfort2
- Distress2
- Respiratory Insufficiency1
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 10, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT PARENT WAS PLANNING A TUBE CHANGE WHEN A SMALL DEFECT WAS NOTICED ON THE TUBE, AT THE END OF THE SHAFT. THE DEFECT WAS DESCRIBED AS ¿A SMALL CIRCULAR SHAPE, A HOLE, A BUBBLE¿ WITHIN THE SHAFT TUBING. THE REMAINING TUBES WERE NOT USED AS ALTERNATIVES. THE FAULT WAS NOTICED BY THE PARENTS BEFORE USE, WHEN BOX WAS OPENED. THERE WAS NO PATIENT INVOLVEMENT AND NO HUMAN HARM/ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22473361 ↗MalfunctionMar 13, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A SPLIT IN THE FLANGE OF THE TRACHEOSTOMY IN USE. THE EVENT DID OCCUR IN THE PATIENT'S HOME. THE PARENT NOTICED THE TEAR IN THE FLANGE. THERE WAS NO PATIENT HARM BECAUSE THE PARENT CHANGED THE TRACHEOSTOMY TUBE ONCE THE TEAR WAS NOTICED AND INFORMED THE CHILD'S NURSE. OTHER RELEVANT HISTORY, INCLUDING PRE-EXISTING CONDITIONS: THE PATIENT USES VENTILATION DUE TO CONGENITAL CENTRAL HYPOVENTILATION SYNDROME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21596138 ↗MalfunctionFeb 19, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT THE PATIENT EXPERIENCED RESISTANCE ON SUCTIONING WHILE THE TRACHEOSTOMY TUBE ON WAS PUT ON. THE SIMILAR SITUATION HAPPENED WHEN THEY SWITCHED TO A NEW DEVICE. PATIENT REPORTED TO HAVE THICK SECRETIONS. PARENTS REPORTED THAT THE SUCTION CATHETER INSERTED WITH EASE BUT AND THEY HAD TO FORCE THE CATHETER TO OVERCOME THE RESISTANCE. CORRECT SUCTION CATHETER SIZE WAS USED. THERE WERE NO HARMS REPORTED AS A RESULT OF THE EVENT.
Patient problems: Respiratory Insufficiency
MAUDE report 21415578 ↗MalfunctionJan 5, 2024Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE WAS BROKEN. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18457630 ↗MalfunctionJul 19, 2023Material Discolored; Material Split, Cut or Torn; Material Discolored; Material Split, Cut or Torn
IT WAS REPORTED THAT THE PARENT CONTACTED CONCERNED THAT THEY HAD DISCOVERED A SPLIT IN THE TUBING OF A TRACHEOSTOMY TUBE ON REMOVAL FROM THE CHILD. THE BACK UP TRACHEOSTOMY TUBE THEY WERE ABOUT TO INSERT WAS ALSO FOUND TO HAVE AN AREA THAT APPEARED SIMILARLY DISCOLORED. THE PLASTIC WAS DISCOLORED AND MAY ALSO HAVE A MICRO SPLIT THAT WAS NOT VISIBLE. REPORTED TO BE FOUND WHILE IN USE WITH PATIENT. NO PATIENT INJURY REPORTED. THE FAMILY HAD AN EXTRA SPARE TRACHEOSTOMY TUBE. THE CHILD WAS ABLE TO TOLERATE A DELAY IN THE REPLACEMENT OF HIS TRACHEOSTOMY TUBE. IT WAS REPORTED THAT THE CHILD WAS "UPSET/STRESSED AND REQUIRED ADDITIONAL SUPPORT SUCH AS INCREASED SUCTIONING". A1: ONE PATIENT, TWO PRODUCT MALFUNCTIONS. (B)(6) AND (B)(6) BOTH REPRESENT THE SAME PATIENT.
Patient problems: Discomfort; Distress
MAUDE report 17349407 ↗MalfunctionJul 19, 2023Material Discolored; Material Split, Cut or Torn; Material Discolored; Material Split, Cut or Torn
ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL. IT WAS NOTICED THERE WAS AN INCREASE IN SECRETIONS AND THE PATIENT HAS HAD A SUBSEQUENT CONSULTANT REVIEW AT THE SPECIALIST CENTER IN WHICH THE TRACHEOSTOMY BRAND WAS CHANGED.
Patient problems: Discomfort; Distress
MAUDE report 17349403 ↗MalfunctionApr 14, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE TWO TUBES THAT WERE COLLECTED APPEARED TO HAVE BEEN CUT BY THE PARENTS ON ACCIDENT. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16744270 ↗MalfunctionApr 13, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE TUBE SEEMED TO HAVE BEEN CUT BY THE PARENTS ON ACCIDENT. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16734412 ↗MalfunctionNov 18, 2022Component Missing; Component Missing
IT WAS REPORTED THAT THE UNIT IS MISSING A PART BUT CAN'T DETERMINE THE PART THAT IS MISSING. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15829051 ↗InjuryMay 1, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7476332 ↗InjuryMay 1, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7476319 ↗InjuryMay 1, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7476308 ↗InjuryMay 1, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7476293 ↗InjuryMay 1, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7476287 ↗InjuryMay 1, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7476282 ↗InjuryMay 1, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7476278 ↗InjuryApr 30, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7474092 ↗InjuryApr 30, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7474091 ↗InjuryApr 30, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7474085 ↗InjuryApr 30, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7474081 ↗InjuryApr 30, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7474078 ↗InjuryApr 30, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7474077 ↗InjuryApr 30, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7474076 ↗MalfunctionApr 30, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7474073 ↗InjuryApr 30, 2018Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE AIRWAY ADAPTER COULD NOT BE SEPARATED FROM THE NEONATAL/PEDIATRIC TRACHEOSTOMY TUBE. NO SERIOUS INJURY WAS REPORTED IN CONNECTION WITH THIS INCIDENT. THE PATIENT WAS HOWEVER MANUALLY "BAGGED" WHILE THE AIRWAY ADAPTER WAS REMOVED FROM THE TRACHEOSTOMY TUBE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7473345 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083641, K944178. One submission often covers many models, so check the product and lot details in each record.
- Z-0205-2024 · Aug 22, 2023 · Open, ClassifiedSmiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
- Z-1145-2021 · Jan 18, 2021 · TerminatedTube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck Flange 4.5mm - Product Us
- Z-0726-2021 · Dec 15, 2020 · TerminatedPORTEX Bivona Cuffless Neonatal/Pediatric FlexTend Tracheostomy Tube, REF 60PFSS45
- Z-2946-2020 · Aug 28, 2020 · TerminatedBivona Mid-Range Aire-Cuf Adult Tracheostomy Tube This tube is intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed up to
- Z-1465-2017 · Feb 9, 2017 · TerminatedBivona TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.