Quick summary
This ICU Medical Bivona trach tube is a cuffed tracheostomy tube. Most of the 38 reports describe the securement flange splitting, cracking or tearing while in use, and an obturator that would not come out during tube changes, sometimes forcing an emergency tube change at home. There are no deaths, 18 injuries and 20 malfunctions. One open recall, started in May 2024, covers the same flange tearing from a manufacturing defect in certain lots.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 37
- Deaths reported
- 0
- Injuries reported
- 18
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 15021312516791), received between Nov 8, 2017 and Apr 6, 2026 — classified as 18 injuries, 19 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K083641, K944178); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 29, 2024Open, Classified
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Z-2601-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Break8
- Device Markings/Labelling Problem8
- Activation, Positioning or Separation Problem3
- Fracture3
- Corroded2
- Degraded2
- Difficult to Remove2
- Material Split, Cut or Torn2
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Consequences Or Impact To Patient9
- Insufficient Information5
- Distress4
- Irritation4
Recent adverse event reports
The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 6, 2026Difficult to Remove
IT WAS REPORTED THAT WHEN ATTEMPTING TO CHANGE THE PATIENT¿S TRACHEOSTOMY TUBE, BUT THE OBTURATOR DID NOT COME OUT. THIS CAUSED AN EMERGENCY REPLACEMENT, WHICH RESULTED IN THE SAME ISSUE OF THE OBTURATOR BECOMING STUCK. THEY WERE EVENTUALLY ABLE TO CHANGE THE TRACHEOSTOMY TUBE ON THE THIRD ATTEMPT USING AN OLDER TUBE WITHOUT A SHEATH. THERE WAS PATIENT INVOLVEMENT AND NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24789659 ↗MalfunctionApr 6, 2026Difficult to Remove
IT WAS REPORTED THAT WHEN ATTEMPTING TO CHANGE THE PATIENT¿S TRACHEOSTOMY TUBE, BUT THE OBTURATOR DID NOT COME OUT. THIS CAUSED AN EMERGENCY REPLACEMENT, WHICH RESULTED IN THE SAME ISSUE OF THE OBTURATOR BECOMING STUCK. THEY WERE EVENTUALLY ABLE TO CHANGE THE TRACHEOSTOMY TUBE ON THE THIRD ATTEMPT USING AN OLDER TUBE WITHOUT A SHEATH. THERE WAS PATIENT INVOLVEMENT AND NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24789467 ↗InjuryFeb 26, 2025Break; Crack; Break; Crack
IT WAS REPORTED THAT THE FLANGE OF TRACHEOSTOMY TUBE FULLY SPLIT AND CAUSED THE TUBE TO COME OUT WHICH REQUIRED AN EMERGENCY TRACHEOSTOMY CHANGE. THE EVENT OCCURRED AT THE PATIENT'S HOME. THERE WAS NO PATIENT HARM REPORTED, AND THE INCIDENT WAS RESOLVED WITH AN EMERGENCY TRACHEOSTOMY TUBE CHANGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21470138 ↗MalfunctionJan 13, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT A SPLIT WAS OBSERVED ON THE RIGHT SIDE OF THE FLANGE WHILE THE CHILD WAS USING IT. AS A RESULT, AN EMERGENCY TUBE CHANGE WAS PERFORMED. THE TUBE HAS BEEN USED THREE TIMES AND STERILIZED TWICE. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21139672 ↗InjuryOct 29, 2024Fracture; Fracture
IT WAS REPORTED THAT THE DEVICE TORE AT THE TRACH TUBE FLANGE EYELET. THE PATIENT HAD TO USE AN ALTERNATE TRACH TUBE. THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20561779 ↗InjuryJun 13, 2024Fracture; Fracture
IT WAS REPORTED THAT THERE WERE TWO TRACHS WITH BROKEN FLANGES. BOTH TRACHS FAILED WHILE IN USE WITH THE PATIENT WITH NO PATIENT HARM REPORTED. MOREOVER, IT WAS STATED THAT HOME NURSE HAD TO REPLACE BOTH BROKEN TRACHES WITH THEIR LAST PRIMARY BACK-UP ONE DUE TO BROKEN FLANGES ON BOTH UNITS. THERE WAS PATIENT INVOLVEMENT WITH NO PATIENT HARM/ADVERSE EVENT REPORTED. A1: ONE PATIENT, TWO PRODUCT MALFUNCTIONS. EMDR-25371 AND EMDR-25636 BOTH REPRESENT THE SAME PATIENT. THIS IS THE FIRST MALFUNCTION REPORT FOR THE RELATED COMPLAINTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19533137 ↗InjuryJun 13, 2024Fracture; Fracture
IT WAS REPORTED THAT THERE WERE TWO TRACHS WITH BROKEN FLANGES. BOTH TRACHS FAILED WHILE IN USE WITH THE PATIENT WITH NO PATIENT HARM REPORTED. MOREOVER, IT WAS STATED THAT HOME NURSE HAD TO REPLACE BOTH BROKEN TRACHES WITH THEIR LAST PRIMARY BACK-UP ONE DUE TO BROKEN FLANGES ON BOTH UNITS. THERE WAS PATIENT INVOLVEMENT WITH NO PATIENT HARM/ADVERSE EVENT REPORTED. ONE PATIENT, TWO PRODUCT MALFUNCTIONS. EMDR-25371 AND EMDR-25636 BOTH REPRESENT THE SAME PATIENT. THIS IS THE SECOND MALFUNCTION REPORT FOR THE RELATED COMPLAINTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19533090 ↗MalfunctionJun 15, 2023Break; Defective Device; Break; Defective Device
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE WAS BROKEN. NO PATIENT HARM OR INVOLVEMENT REPORTED.
Patient problems: Insufficient Information
MAUDE report 17133643 ↗MalfunctionJun 12, 2023Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE BIVONA TRACH WAS "DEFECTIVE" AND BROKE. PATIENT INVOLVEMENT UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 17113031 ↗MalfunctionMay 9, 2023Corroded; Degraded; Corroded; Degraded
ADDITIONAL INFORMATION RECEIVED VIA EMAIL: THE ISSUES WERE NOTICED FOLLOWING STERILIZATION AFTER USE BUT PRIOR TO ANY FURTHER USE AND REMOVED FROM USE IMMEDIATELY. NO INTERVENTION NEEDED BUT FURTHER TRACHEOSTOMY TUBE NEEDED IMMEDIATELY TO REPLACE AFFECTED TUBE. THE EVENT WAS RESOLVED BUT THIS WAS THE THIRD OCCASION THIS HAS HAPPENED WITH THIS PATIENT. NO RELEVANT TESTS/LABORATORY DATA. OTHER RELEVANT HISTORY, INCLUDING PRE-EXISTING CONDITIONS - THE PATIENT REQUIRES A TRACHEOSTOMY TUBE TO MAINTAIN THE AIRWAY DUE TO PHARYNGOMALACIA. HAS CHRONIC LUNG DISEASE FROM PREMATURITY WITH EMPHYSEMATOUS BRONCHOPULMONARY DYSPLASIA, AND SUBGLOTTIC CYSTS. WAS BORN AT 24+5 WEEKS GESTATION AND HAS PREVIOUSLY REQUIRED NON-INVASIVE VENTILATION AT HOME AND OXYGEN ? VENTILATION STOPPED 1 YEAR AGO, AND OXYGEN STOPPED RECENTLY. CONCOMITANT MEDICAL PRODUCTS AND THERAPY DATES (EXCLUDES TREATMENT OF EVENT) ? THE PATIENT USES TRACHEOSTOMY TUBES, SUCTION CATHETERS, SUCTION MACHINE, NG TUBE AND FEEDING PUMP. THE MEDICATIONS PRESCRIBED AT THE TIME OF INCIDENT WERE OMEPRAZOLE TWICE DAILY, LEVETIRACETAM TWICE DAILY AND AZITHROMYCIN EVERY OTHER DAY. PATIENT INFO WAS ADDED. OCCURRED IN FAMILY HOME. NO ADVERSE PATIEN
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16899312 ↗MalfunctionMay 9, 2023Corroded; Degraded; Corroded; Degraded
ADDITIONAL INFORMATION RECEIVED VIA EMAIL: THE ISSUES WERE NOTICED FOLLOWING STERILIZATION AFTER USE BUT PRIOR TO ANY FURTHER USE AND REMOVED FROM USE IMMEDIATELY. NO INTERVENTION NEEDED BUT FURTHER TRACHEOSTOMY TUBE NEEDED IMMEDIATELY TO REPLACE AFFECTED TUBE. THE EVENT WAS RESOLVED BUT THIS WAS THE THIRD OCCASION THIS HAS HAPPENED WITH THIS PATIENT. NO RELEVANT TESTS/LABORATORY DATA. OTHER RELEVANT HISTORY, INCLUDING PRE-EXISTING CONDITIONS - THE PATIENT REQUIRES A TRACHEOSTOMY TUBE TO MAINTAIN THE AIRWAY DUE TO PHARYNGOMALACIA. HAS CHRONIC LUNG DISEASE FROM PREMATURITY WITH EMPHYSEMATOUS BRONCHOPULMONARY DYSPLASIA, AND SUBGLOTTIC CYSTS. WAS BORN AT 24+5 WEEKS GESTATION AND HAS PREVIOUSLY REQUIRED NON-INVASIVE VENTILATION AT HOME AND OXYGEN ? VENTILATION STOPPED 1 YEAR AGO, AND OXYGEN STOPPED RECENTLY. CONCOMITANT MEDICAL PRODUCTS AND THERAPY DATES (EXCLUDES TREATMENT OF EVENT) ? THE PATIENT USES TRACHEOSTOMY TUBES, SUCTION CATHETERS, SUCTION MACHINE, NG TUBE AND FEEDING PUMP. THE MEDICATIONS PRESCRIBED AT THE TIME OF INCIDENT WERE OMEPRAZOLE TWICE DAILY, LEVETIRACETAM TWICE DAILY AND AZITHROMYCIN EVERY OTHER DAY. PATIENT INFO WAS ADDED. DATE OF EVENT WAS ON 18-APR-2023 AND OCCUR
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16899309 ↗MalfunctionApr 14, 2023Product Quality Problem; Product Quality Problem
IT WAS REPORTED THAT THE ITEM WAS "FAULTY, THE COILS ON DAMAGED FROM PRODUCTION". NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16744072 ↗MalfunctionDec 7, 2022Component Missing; Component Missing
IT WAS REPORTED THAT THE CORRECT TUBE SHOULD HAVE A STRAIGHT FLANGE BUT A V-SHAPE FLANGE WAS RECEIVED. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15937132 ↗MalfunctionSep 27, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT THE TUBE WAS MISLABELED AS 60PFSS45 BUT IS A V FLANGE AND NOT A TTS. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15492408 ↗MalfunctionSep 19, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT UPON OPENING THE PACKAGE, IT WAS DISCOVERED THAT PART NUMBER 67PFSS45 WAS FOUND INSIDE THE PACKAGE BUT PART NUMBER 60PFSS45 WAS ORDERED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15445137 ↗MalfunctionSep 16, 2022Device Misassembled During Manufacturing /Shipping
IT WAS REPORTED THAT THE THE CORRECT TUBE SHOULD HAVE A STRAIGHT FLANGE BUT A V-SHAPE FLANGE WAS RECEIVED. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15437947 ↗MalfunctionSep 16, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT UPON OPENING THE PACKAGE, THE LABEL LISTS A STRAIGHT FLANGE BUT THE TRACH RECEIVED HAD A V-FLANGE. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15437196 ↗MalfunctionAug 23, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT THE DEVICE WAS LABELLED AS STRAIGHT FLANGE BUT PRODUCT WAS A V FLANGE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15280308 ↗MalfunctionJun 3, 2022Defective Component; Defective Component
IT WAS REPORTED THAT THE INTRODUCER GOT LODGED IN THE DURING PRE-USE CHECK. A NEW TUBE WAS TESTED PRIOR TO INSERTION AND WAS USED TO RESOLVE THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14594898 ↗InjuryApr 2, 2021Break; Break
PARENTS NOTICED TEAR IN THE EYELET. TRACH CHANGE OUT DONE TO RESOLVE ISSUE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11611744 ↗InjuryApr 2, 2021Break; Break
PARENTS NOTICED TEAR STARTING AT FLANGE AREA OF TRACHEOSTOMY WHERE THE NECK TIES INSERTED. DECIDED TO CHANGE THE TRACHEOSTOMY TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11611743 ↗InjuryApr 2, 2021Break; Break
PARENTS NOTICED TEAR IN THE EYELET. TRACH CHANGE OUT DONE TO RESOLVE ISSUE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11611738 ↗InjuryApr 2, 2021Break; Break
PARENTS NOTICED TEAR IN THE EYELET. TRACH CHANGE OUT DONE TO RESOLVE ISSUE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 11611736 ↗MalfunctionApr 2, 2021Break; Break
IT WAS REPORTED THAT PARENTS NOTICED TEAR IN THE EYELET. TRACH CHANGE OUT DONE TO RESOLVE ISSUE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11611734 ↗MalfunctionNov 13, 2020Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
INVESTIGATION COMPLETED AND NO PRODUCT WAS RETURNED, HOWEVER A PICTURE OF DEVICE TORN WAS. H 10 ENGINEERING SUMMARY OF ANALYSIS .
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10834466 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083641, K944178. One submission often covers many models, so check the product and lot details in each record.
- Z-0205-2024 · Aug 22, 2023 · Open, ClassifiedSmiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
- Z-1145-2021 · Jan 18, 2021 · TerminatedTube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck Flange 4.5mm - Product Us
- Z-0726-2021 · Dec 15, 2020 · TerminatedPORTEX Bivona Cuffless Neonatal/Pediatric FlexTend Tracheostomy Tube, REF 60PFSS45
- Z-2946-2020 · Aug 28, 2020 · TerminatedBivona Mid-Range Aire-Cuf Adult Tracheostomy Tube This tube is intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed up to
- Z-1465-2017 · Feb 9, 2017 · TerminatedBivona TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page