Skip to content

Bivona 60PFPS40Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Bivona (469)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 15021312516838), received between Nov 23, 2021 and Nov 13, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K083641, K944178); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 29, 2024Open, Classified

    Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

    Z-2598-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information3
  • Defective Device1
  • Degraded1
  • Device Markings/Labelling Problem1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING ROUTINE CLEANING AND TAPE CHANGE, SUPPORT STAFF OBSERVED A SPLIT IN THE TRACHEOSTOMY TUBE AT THE POINT WHERE IT WIDENS TO CONNECT TO THE VENTILATOR, ABOVE THE FLANGE. THE VENTILATOR DID NOT ALARM TO INDICATE ANY ISSUE. THE TRACHEOSTOMY TUBE WAS IMMEDIATELY REPLACED WITHOUT COMPLICATION, AND NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23543262 ↗
  • MalfunctionJan 15, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE DEVICE WAS "DEFECTIVE". PATIENT INVOLVEMENT UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18514953 ↗
  • MalfunctionJan 12, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE DEVICE WAS "DEFECTIVE". TWO DEVICE LOT NUMBERS WERE REPORTED. BOTH DEVICES ARE UNUSED. THE SECOND INVOLVED DEVICE WAS DOCUMENTED ON (B)(4) AND REPRESENTS THE SAME PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18508537 ↗
  • MalfunctionJan 12, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE DEVICE WAS "DEFECTIVE". TWO DEVICE LOT NUMBERS WERE REPORTED. BOTH DEVICES ARE UNUSED. THE SECOND INVOLVED DEVICE WAS DOCUMENTED ON (B)(4) AND REPRESENTS THE SAME PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18508095 ↗
  • MalfunctionOct 31, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT ON THE OUTSIDE OF THE BOX, THE LOT NUMBER INDICATES THE FLANGE SHOULD BE STRAIGHT. WHEN THE TRACH WAS OPENED, IT WAS NOTED THAT THE FLANGE WAS V-SHAPED INSTEAD. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15698355 ↗
  • MalfunctionOct 31, 2022Defective Device; Defective Device

    IT WAS REPORTED THAT ON THE OUTSIDE OF THE BOX, THE LOT NUMBER INDICATES THE FLANGE SHOULD BE STRAIGHT. WHEN THE TRACH WAS OPENED, IT WAS NOTED THAT THE FLANGE WAS V-SHAPED INSTEAD. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15698330 ↗
  • MalfunctionNov 23, 2021Degraded; Degraded

    IT WAS REPORTED THAT THE TRACH TUBE SEEMS TO BE DETERIORATING. THE CUSTOMER FEELS LIKE THIS COULD POTENTIALLY BE LIFE THREATENING DUE TO THE POSSIBILITY OF THE PATIENT ASPIRATING DEBRIS INTO HER LUNGS. PHOTO ATTACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12863619 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083641, K944178. One submission often covers many models, so check the product and lot details in each record.

  • Z-0205-2024 · Aug 22, 2023 · Open, Classified
    Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
  • Z-1145-2021 · Jan 18, 2021 · Terminated
    Tube Tracheostomy and Tube Cuff, Bivona Cuffless Neonatal/Pediatric FlexTend" Tracheostomy Tube (model 60PFS45, lots 3991171, 3991172, and 3988103) V Neck Flange 4.5mm - Product Us
  • Z-0726-2021 · Dec 15, 2020 · Terminated
    PORTEX Bivona Cuffless Neonatal/Pediatric FlexTend Tracheostomy Tube, REF 60PFSS45
  • Z-2946-2020 · Aug 28, 2020 · Terminated
    Bivona Mid-Range Aire-Cuf Adult Tracheostomy Tube This tube is intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed up to
  • Z-1465-2017 · Feb 9, 2017 · Terminated
    Bivona TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page