Smart Port CT H787CT75STSA0Recalls & reported problems
Angiodynamics, Inc. · Smart Port CT Single Titanium Port System with Attached 7.5F x 66cm Silicone Catheter and 8F Introducer Kit · Class 2 · Implant
All models of Smart Port CT (36)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15051684017992), received between Apr 22, 2025 and Feb 5, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (K062414); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Migration2
- Fracture1
Patient problems
- Foreign Body In Patient2
- Chest Pain1
- Dyspnea1
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2026Detachment of Device or Device Component
A PATIENT REPORTED AN ISSUE WITH THEIR SMARTPORT CT 7.5F SL TI STD FILLED ATT SIL. THE PATIENT HAD THEIR PORT PLACED ON (B)(6) 2024 AT (B)(6) HOSPITAL (B)(6), BY DR. (B)(6). WHILE THE PORT WAS IN PLACE; IT REQUIRED MANY FLUSHES. ON (B)(6) 2025, DR. (B)(6) ATTEMPTED TO REMOVE THE PORT AND A PORTION OF THE CATHETER BUT COMMENTED THAT IT SHOULD BE LONGER. IT WAS THEN REALIZED THAT A PORTION OF THE CATHETER MAY HAVE DETACHED, WITHIN THE PATIENT. ON (B)(6) 2025, THE PATIENT HAD A FOLLOW-UP APPOINTMENT AT (B)(6) IN (B)(6), WITH HER SURGEON DR. (B)(6), WHERE A CHEST X-RAY SHOWED THAT A FRAGMENT OF THE CATHETER WAS IN THE PATIENT'S PULMONARY ARTERY. THE FRAGMENT WAS REMOVED VIA A PROCEDURE ON (B)(6) 2025 AND THE PATIENT DID NOT EXPERIENCE ANY HARM OR ADVERSE EFFECTS AS A RESULT OF THIS MALFUNCTION.
Patient problems: Foreign Body In Patient
MAUDE report 24274586 ↗MalfunctionSep 17, 2025Fracture; Detachment of Device or Device Component; Migration; Fracture; Detachment of Device or Device Component; Migration
ON OR ABOUT (B)(6)2012, PLAINTIFF UNDERWENT RIGHT SUBCLAVIAN PLACEMENT OF AN ANGIODYNAMICS SMARTPORT PORT CATHETER PRODUCT, WITH MODEL NUMBER CT75STSA. THIS SMARTPORT PRODUCT IS COMPRISED OF A PORT BODY AND A SILICONE CATHETER. THE DEVICE WAS IMPLANTED BY DR. (B)(6) AT (B)(6) IN (B)(6) FOR THE PURPOSE OF CHEMOTHERAPY ADMINISTRATION TO TREAT PLAINTIFF'S BREAST CANCER. ON OR ABOUT EARLY (B)(6)2013, PLAINTIFF PRESENTED TO (B)(6) MEDICAL CENTER IN (B)(6), FOR A CHEMOTHERAPY TREATMENT, AND SUBSEQUENTLY HAD COMPLAINTS OF SIGNIFICANT PAIN IN HER RIGHT SHOULDER UPON ACCESSING AND FLUSHING THE PORT. PLAINTIFF UNDERWENT IMAGING WHICH DEMONSTRATED A FRACTURED CATHETER WITH EXTRAVASATION. ON OR ABOUT (B)(6)2013, PLAINTIFF PRESENTED TO DR. (B)(6) AT (B)(6) IN(B)(6),, FOR REMOVAL OF THE SMARTPORT, WHEREBY DR. (B)(6) REMOVED THE SMARTPORT BODY AND A SHORT SEGMENT OF THE DEVICE'S CATHETER. DR. (B)(6) WAS UNSUCCESSFUL IN REMOVING THE FRACTURED CATHETER FRAGMENT, WHICH REMAINED IN PLAINTIFF'S BODY. ON OR ABOUT (B)(6)2013, DR. (B)(6) REFERRED PLAINTIFF TO DR. (B)(6) A VASCULAR SURGEON, FOR EVALUATION OF THE RETAINED CATHETER FRAGMENT, WHICH WAS NOTED TO HAVE MIGRATED TO PLAINTIFF'S RIGHT VENTRICLE. D
Patient problems: Pain
MAUDE report 23075034 ↗MalfunctionApr 22, 2025Detachment of Device or Device Component; Migration; Detachment of Device or Device Component; Migration
ON OR ABOUT ON (B)(6) 2022, PLAINTIFF UNDERWENT PLACEMENT OF AN ANGIODYNAMICS SMARTPORT CATHETER PRODUCT, MODEL NUMBER: CT75STSA, LOT NUMBER: 5742657. THE SMARTPORT WAS IMPLANTED BY DR. (B)(6), M.D, AT (B)(6). ON OR ABOUT ON (B)(6) 2023, PLAINTIFF PRESENTED TO (B)(6) WITH COMPLAINTS OF CHEST PAIN AND DIFFICULTY BREATHING. PLAINTIFF UNDERWENT A CT SCAN OF THE PULMONARY ARTERIES, WHICH DEMONSTRATED THAT HER PORT HAD FRACTURED AND MIGRATED TO HER RIGHT PULMONARY ARTERY. ON OR ABOUT ON (B)(6) 2023, PLAINTIFF'S CATHETER FRAGMENT WAS REMOVED BY DR. (B)(6), M.D., AT (B)(6) HOSPITAL. ON OR ABOUT ON (B)(6) 2023, PLAINTIFF UNDERWENT AN ADDITIONAL PROCEDURE TO HAVE THE REMAINING PORT REMOVED BY DR. (B)(6), M.D., AT (B)(6) HOSPITAL.
Patient problems: Chest Pain; Dyspnea; Foreign Body In Patient
MAUDE report 21882414 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062414. One submission often covers many models, so check the product and lot details in each record.
- Z-1466-2021 · Feb 22, 2021 · TerminatedSmart Port CT-Implantable Port Catalogue #: CT96STSD-VI, CT96STSA-VI - Product Usage: indicated for any patient requiring repeated access of the vascular system, for delivery of me
- Z-0056-2012 · Aug 1, 2011 · Terminated8F Smart Port CT, Catalog number: CT80STPD and CT80STPD-NF; AngioDynamics, INC. The Smart Port CT Implantable Port is a Single Titanium Port System with Attachable 8.0 Fr. Polyur
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.