Skip to content

Smart Port CT H787CT96STSDNF0Recalls & reported problems

Angiodynamics, Inc. · Smart Port CT Single Titanium Port System with Attachable 9.6F x 66cm Silicone Catheter and 10 F Introducer Kit (with Non-Filled Suture Holes) · Class 2 · Implant

All models of Smart Port CT (36)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15051684018074), received between Mar 12, 2025 and Mar 30, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K062414); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1

Patient problems

  • Bacterial Infection1
  • Limb Pain1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem

    ON OR ABOUT (B)(6) 2015, PLAINTIFF UNDERWENT PLACEMENT OF AN ANGIODYNAMICS SMARTPORT, MODEL NUMBER CT96STSD-NF, LOT NUMBER 4820412. THE SMARTPORT WAS IMPLANTED BY DR. (B)(6), AT (B)(6). ON OR ABOUT (B)(6) 2023, PLAINTIFF PRESENTED TO (B)(6) WITH COMPLAINTS OF LEG PAIN AND A PAIN CRISIS. SHE UNDERWENT BLOOD CULTURES AFTER EXHIBITING SYMPTOMS OF INFECTION. THE BLOOD CULTURES GREW SERRATIA. PLAINTIFF'S MEDICAL TEAM DETERMINED THAT THE SMARTPORT WAS THE SOURCE OF THE INFECTION AND THAT IT REQUIRED REMOVAL. ON OR ABOUT (B)(6) 2023, PLAINTIFF UNDERWENT AN UNSUCCESSFUL ATTEMPT TO REMOVE THE SMARTPORT BY DR. (B)(6) THE PORT WAS SCARRED IN PLACE, AND DR. (B)(6) WAS UNABLE TO REMOVE THE DEVICE. ON OR ABOUT (B)(6) 2023, PLAINTIFF UNDERWENT AN ADDITIONAL PROCEDURE TO HAVE THE SMARTPORT REMOVED. THE DEFECTIVE DEVICE WAS REMOVED BY DR. (B)(6), AT (B)(6).

    Patient problems: Bacterial Infection; Limb Pain; Pain

    MAUDE report 24728320 ↗
  • MalfunctionMar 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    AN END USER REPORTED AN ISSUE WITH A CT VTX PORT W/SIL CATH NONFILLED HOLES. A PATIENT PRESENTED TO AN INFUSION CLINIC FOR THEIR FIRST INFUSION, POST IMPLANTATION. WHILE THE LINE WAS BEING FLUSHED WITH SALINE, THE PATIENT IMMEDIATELY COMPLAINED OF PAIN. THE PROCEDURE WAS STOPPED AND PATIENT RECEIVED A CONSULTATION WITH A PHYSICIAN. THE PHYSICIAN DETERMINED THAT AN EMERGENCY EXTRACTION OF THE DEVICE NEEDED TO TAKE PLACE; THEREFORE, THE PORT WAS REMOVED AND REPLACED WITH A PICC LINE. UPON INSPECTION, IT APPEARED THAT THE LINE DISCONNECTED FROM THE DEVICE, INSIDE THE STRAIN RELIEF, THAT IS LOCATED BELOW THE PORT. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21586639 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K062414. One submission often covers many models, so check the product and lot details in each record.

  • Z-1466-2021 · Feb 22, 2021 · Terminated
    Smart Port CT-Implantable Port Catalogue #: CT96STSD-VI, CT96STSA-VI - Product Usage: indicated for any patient requiring repeated access of the vascular system, for delivery of me
  • Z-0056-2012 · Aug 1, 2011 · Terminated
    8F Smart Port CT, Catalog number: CT80STPD and CT80STPD-NF; AngioDynamics, INC. The Smart Port CT Implantable Port is a Single Titanium Port System with Attachable 8.0 Fr. Polyur

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page