AU OSR6178Recalls & reported problems
Beckman Coulter, Inc. · Creatinine · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15099590010737), received between Jan 21, 2020 and Jan 21, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Low Test Results” (1 report).
1 other recall cover devices cleared under the same FDA submission (K924692); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Low Test Results1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 21, 2020Low Test Results; Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF A FALSE LOW CREATININE RESULT FOR ONE (1) PATIENT SAMPLE ON THE AU5800 CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS PATIENT RESULT WAS REPORTED OUTSIDE THE LAB. DUE TO THE FALSE LOW CREATININE PATIENT RESULT, THE PATIENT WAS SCANNED AND RECEIVED AN INTRAVENOUS INJECTION OF IODINE CONTRAST AGENT. THERE WAS NO ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9613448 ↗
FDA clearance and classification
Cleared through 510(k) K924692 on Oct 19, 1992 (decision: Substantially Equivalent). Submitted by Olympus Corp. as “Olympus Creatinine Reagent”. The FDA review took 32 days. FDA record ↗
The FDA regulates product code CGX (alkaline picrate, colorimetry, creatinine) under 21 CFR 862.1225 as a Class II device. 21 CFR 862.1225 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K924692. One submission often covers many models, so check the product and lot details in each record.
- Z-0958-2013 · Dec 28, 2012 · TerminatedCreatinine, Part Numbers OSR6178, OSR6678 Product Usage: System reagent for the quantitative determination of Creatinine in human serum and urine on Beckman Coulter AU analyzer
Common questions about the AU OSR6178
Has the AU OSR6178 been recalled?
No FDA recall record in our data names this exact device identifier (DI 15099590010737), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the AU OSR6178?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jan 21, 2020. That report lists device problem “Low Test Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the AU OSR6178?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the AU OSR6178 FDA approved?
It was cleared through 510(k) K924692 on Oct 19, 1992. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the AU OSR6178, and is it still sold?
The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand AU, model OSR6178). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Alkaline Picrate, Colorimetry, Creatinine devices
- Dimension® Flex® reagent cartridge CRE2 10872079Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE1 recall9 reports
- ABX PENTRA CREATININE 120 CP 1220001933Horiba Abx Sas · Alkaline Picrate, Colorimetry, Creatinine1 recall0 reports
- ARCHITECT 3L81-22Abbott Laboratories · ALKALINE PICRATE, COLORIMETRY, CREATININE31 reports
- Alinity 07P9920Abbott GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE24 reports
- Atellica CH Crea_2 11097596Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE19 reports
- Creatinine Jaffe Gen.2 08057532190Roche Diagnostics GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE16 reports
- Alinity 04T9120Abbott Ireland · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
- ARCHITECT 3L81-32Abbott Laboratories · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page