Beckman Coulter, Inc. · The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias. With the advent of the administration of recombinant erythropoietin as a biologic therapy to increase red blood cell mass, an erythropoietin assay may be used also to aid in the prediction and monitoring of response to recombinant erythropoietin treatment of anemias. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Feb 3, 2025Class IIOpen, Classified

    Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing.

    Z-1249-2025 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code GGT (assay, erythropoietin) under 21 CFR 864.7250 as a Class II device. 21 CFR 864.7250 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Access EPO A16364

Has the Access EPO A16364 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1249-2025 (Class II), began on Feb 3, 2025: Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Who makes the Access EPO A16364, and is it still sold?

The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand Access EPO, model A16364). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.