Skip to content

Access Erythropoietin (EPO), Catalog Number - A16364 recallZ-1249-2025 · Beckman Coulter, Inc. · initiated Feb 3, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing.

FDA root cause category: Unknown/Undetermined by firm

What the company is doing

Firm began notifying customers February 6, 2025. Customers were instructed to discontinue using and discard all remaining reagent packs of the Access EPO reagent lot identified. Customers are advised to share this recall with their medical director and perform a retrospective review of patient test results generated with this lot and assess the need for retesting.

Recall details

Recall number
Z-1249-2025
Status
Open, Classified
Initiated by the company
Feb 3, 2025
Posted by the FDA
Feb 28, 2025
Recalling firm
Beckman Coulter, Inc.
Quantity
498 US; 1576 OUS
Product code
GGT
FDA submissions
K052223
Distribution
Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR PA TN TX UT WA WV. International distribution to Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Korea, Republic of, Netherlands, Philippines, Russian Federation, Saudi Arabia, Slovakia, Spain, United Kingdom of Great Britain and Northern Ireland

Product description

Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal factor regulating red blood cell production (erythropoiesis) in mammals. Renal production of EPO is regulated by changes in oxygen availability. Under conditions of hypoxia, the level of EPO in the circulation increases and this leads to increased production of red blood cells.

Code information (models, lots, serial numbers)

Catalog A16364 UDI-DI 15099590201838 Lot 439363

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page