Skip to content

ATEC ATEC CanisterRecalls & reported problems

Hologic, Inc. · SUCTION CANISTER W/ LID · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15420045503182), received between Nov 20, 2024 and Nov 20, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Failure to Obtain Sample” (1 report).

4 other recalls cover devices cleared under the same FDA submission (K042290); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Obtain Sample1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 20, 2024Failure to Obtain Sample; Failure to Obtain Sample

    IT WAS REPORTED THAT ON A BREVERA PROCEDURE WITH AN ATEC CANNISTER ON (B)(6) 2024, THE TECH REPORTED THAT THE SYSTEM DISPLAYED AN ERROR RELATED TO VACUUM SYSTEM NOT AT REQUIRED PRESSURE WHEN THE NEEDLE WAS INSERTED INTO THE PATIENT´S BREAST AND FIRED. NO SAMPLES COULD BE OBTAINED, AND THE SYSTEM WAS UNABLE TO BE RECONNECTED AND THE PATIENT WAS RESCHEDULED FOR A NEW PROCEDURE. A FIELD ENGINEER EXAMINED THE EQUIPMENT AND IT WAS DETERMINED THAT THE CANISTER WAS NOT SEALING CORRECTLY AS EXPECTED WHICH CAUSED THE VACUUM ERROR ON THE SYSTEM. NO OTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20740054 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042290. One submission often covers many models, so check the product and lot details in each record.

  • Z-2285-2010 · Aug 11, 2010 · Terminated
    Suros ATEC 0912-20 Breast Biopsy and Excision System Handpiece, REF ATEC0912-12, sterile, Hologic, Inc. 6100 Technology Center Drive Indianapolis, IN. The handpiece is inserted i
  • Z-1160-2010 · Feb 1, 2010 · Terminated
    ATEC 0909-12 Breast Biopsy and Excision System , 9 gauge, 9 centimeter handpiece (part number 0909-12), Rx, Sterile, Suros Surgical Systems, Inc. 6100 Technology Center DR. India
  • Z-0275-2009 · Sep 22, 2008 · Terminated
    ATEC Breast Biopsy and Excision System 12 gauge, 20 centimeter handpiece, sterile, Hologic, Inc., Indianapolis, IN; Model number ATEC 1212-20. Intended for partial or complete rem
  • Z-0362-2007 · Dec 15, 2006 · Terminated
    Suros ATEC Breast Biopsy System handpiece (9 gauge cannula 12 centimeters in length with a 20 millimeter aperture) for use with the ATEC Breast Biopsy and Excision System, disposab

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page