Skip to content

ATEC 0912-12Recalls & reported problems

Hologic, Inc. · 9G (3.7MM) X 12CM, 12MM HANDPIECE · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15420045504851), received between Jan 24, 2025 and Dec 10, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Failure to Obtain Sample” (1 report).

4 other recalls cover devices cleared under the same FDA submission (K042290); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Obtain Sample1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 10, 2025Failure to Obtain Sample; Failure to Obtain Sample

    IT WAS REPORTED THAT DURING AN ATEC PROCEDURE ON (B)(6) 2025, "THE SETUP, TEST AND LAVAGE MODE PASSED, BUT ALTHOUGH THE APERTURE WAS OPENING AND CLOSING, NO TISSUE WAS BEING COLLECTED." ADDITIONALLY, IT WAS REPORTED THAT, "THESE NEW NEEDLES ARE DELIVERED WITH AN EXCESSIVE AMOUNT OF GREASE ON THE HANDPIECE AREA, PARTICULARLY AROUND THE BASKET-TO-HANDPIECE CONNECTION. THIS DOES AFFECT THE SUCTION PERFORMANCE, ESPECIALLY WITH LARGER LESIONS OR WHEN A COMPLETE EXCISION IS INTENDED." IT IS REPORTED THAT THE PATIENT PROCEDURE WAS ABORTED. NO FURTHER INFORMATION IS CURRENTLY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23770489 ↗
  • MalfunctionJan 24, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT ON (B)(6) 2024 A PHYSICIAN REPORTED THAT SOMETIMES WHEN PERFORMING A BIOPSY WITH AN ATEC NEEDLE THEY NOTICED A PIECE OF MATERIAL PLASTIC IN BETWEEN THE SPECIMENS. PHYSICIAN PROVIDED A PICTURE IN WHICH IT CAN BE OBSERVED A PLASTIC PARTICLE MIXED ALONG WITH THE SAMPLES. THE PARTICLES WERE ONLY OBSERVED IN THE SAMPLES AND NOT IN THE PATIENTS. NO SAMPLES HAVE BEEN OBSERVED IN THE BREAST OF THE PATIENT. PHYSICIAN REPORTED THAT NO INJURY WAS REPORTED TO PATIENTS NOR USERS. NO OTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21233057 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042290. One submission often covers many models, so check the product and lot details in each record.

  • Z-2285-2010 · Aug 11, 2010 · Terminated
    Suros ATEC 0912-20 Breast Biopsy and Excision System Handpiece, REF ATEC0912-12, sterile, Hologic, Inc. 6100 Technology Center Drive Indianapolis, IN. The handpiece is inserted i
  • Z-1160-2010 · Feb 1, 2010 · Terminated
    ATEC 0909-12 Breast Biopsy and Excision System , 9 gauge, 9 centimeter handpiece (part number 0909-12), Rx, Sterile, Suros Surgical Systems, Inc. 6100 Technology Center DR. India
  • Z-0275-2009 · Sep 22, 2008 · Terminated
    ATEC Breast Biopsy and Excision System 12 gauge, 20 centimeter handpiece, sterile, Hologic, Inc., Indianapolis, IN; Model number ATEC 1212-20. Intended for partial or complete rem
  • Z-0362-2007 · Dec 15, 2006 · Terminated
    Suros ATEC Breast Biopsy System handpiece (9 gauge cannula 12 centimeters in length with a 20 millimeter aperture) for use with the ATEC Breast Biopsy and Excision System, disposab

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page