Skip to content

EVIVA 0913-12Recalls & reported problems

Hologic, Inc. · PRODUCT, EVIVA, 9 GA, 13 CM LENGTH, 12MM APERTURE · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 15420045505322), received between Dec 28, 2017 and Jun 25, 2025 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem2
  • Break1
  • Defective Component1
  • Failure to Fire1
  • Failure to Obtain Sample1

Patient problems

  • No Consequences Or Impact To Patient2
  • Hematoma1
  • No Clinical Signs, Symptoms or Conditions1
  • No Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 25, 2025Failure to Fire; Failure to Fire

    IT IS REPORTED THAT ON (B)(6) 2025, DURING AN EVIVA PROCEDURE, THE USER REPORTED THE EVIVA PETITE BLUNT BIOPSY NEEDLE WOULD NOT FIRE. THE USER WAS GUIDED BY HOLOGIC TECHNICAL SUPPORT ON HOW TO INSTALL THE EVIVA PETITE BLUNT BIOPSY NEEDLE ONTO THE SYSTEM. ADDITIONALLY, IT IS REPORTED THAT THE USER WAS NOT UTILIZING THE STYLET HOLDER TO HOLD THE STYLET IN PLACE DURING THE INITIAL BIOPSY ATTEMPT. IT IS REPORTED THAT THE USER WAS ABLE TO UTILIZE THE STYLET HOLDER AND COMPLETE A SUCCESSFUL BIOPSY; HOWEVER, THE PATIENT REQUIRED A SECOND INCISION TO DO SO. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22329253 ↗
  • InjuryApr 30, 2025Failure to Obtain Sample; Failure to Obtain Sample

    IT WAS REPORTED THAT ON (B)(6) 2025, DURING AN EVIVA PROCEDURE, THE PHYSICIAN EXPERIENCED ISSUES WITH THE SUCTION AND VACUUM. THE NEEDLE STOPPED WORKING AFTER COLLECTING THE SAMPLES AND THE VACUUM ERROR WAS DISPLAYED. AT THIS POINT THE NEEDLE WAS EXCHANGED TO COMPLETED THE PROCEDURE BUT ANOTHER INCISION HAD TO BE PERFORMED. PATIENT WAS UNDER COMPRESSION FOR A LONG TIME AND PRESENTED WITH A BREAST HEMATOMA. UNKNOWN MEDICAL INTERVENTION WAS REQUIRED TO TREAT THE HEMATOMA AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO OTHER INFORMATION AVAILABLE.

    Patient problems: Hematoma

    MAUDE report 21936766 ↗
  • MalfunctionSep 26, 2018Break; Material Integrity Problem; Break; Material Integrity Problem

    IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, A PLASTIC INTRODUCER SHEETH OF THE DEVICE WAS SHREDDED. AS INDICATED, THE DEVICE TEST AND THE BIOPSY PROCEDURE WERE ALL SUCCESSFULLY PERFORMED. HOWEVER, WHEN THE PHYSICIAN WANTED TO REMOVE THE BIOPSY DEVICE AT THE END OF THE PROCEDURE, SHE FELT RESISTANCE. THEN BY PULLING IT OUT WITH POWER, THE PLASTIC INTRODUCER SHEETH WAS DAMAGED AND BROKEN AT ITS END. AS REPORTED, NO PATIENT HARM, INJURY OR IMPACT DID OCCUR.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7910206 ↗
  • MalfunctionSep 11, 2018Defective Component

    THE HOLOGIC BIOPSY PROBE, EVIVA STANDARD 9G DID NOT FUNCTION WHEN WE HAD THE PATIENT IN POSITION PRONE ON THE TABLE READY FOR BIOPSY. WE HAD TO DELAY THE START OF THE BIOPSY WHILE WE EXCHANGED THE DEFECTIVE PROBE FOR ONE THAT WORKED PROPERLY. THIS EXTENDED THE LENGTH OF TIME THE PATIENT WAS PRONE AND COMPRESSION.

    Patient problems: No Information

    MAUDE report 7863664 ↗
  • MalfunctionDec 28, 2017Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED A PHYSICIAN COMPLETED AN EVIVA BREAST BIOPSY ON (B)(6) 2017 AND HAD DIFFICULTY REMOVING THE DEVICE. WHEN THE PLASTIC INTRODUCER SHEATH WAS REMOVED THERE WERE SOME MISSING PLASTIC PARTS. THE PHYSICIAN THEN NOTED THE PLASTIC PARTS WERE LOCATED AT THE PATIENT'S INCISION HOLE. THERE WAS NO HARM TO THE PATIENT. THE PATIENT WAS DISCHARGED HOME.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7152413 ↗

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.