Horizon ARecalls & reported problems
Hologic, Inc. · The Horizon® Dual-Energy X-Ray Absorptiometry (DXA) System is a computer-assisted X-Ray scanner used to aid a physician in measuring bone mineral density (BMD), determining fracture risk, and estimating the lean body mass and fat mass of non-osseous tissues. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15420045505384), received between Nov 7, 2017 and Oct 21, 2025 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jun 13, 2024Open, Classified
Non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601-1-2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA System exceeded the limit for acceptability.
Z-2373-2024 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Device Displays Incorrect Message1
- Unintended Electrical Shock1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- Not Applicable1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 21, 2025Unintended Electrical Shock; Unintended Electrical Shock
IT WAS REPORTED BY THE USER SITE THAT THE HORIZON SYSTEM HAS HAD ISSUES WITH INTERMITTENT SHUTDOWNS AND ONE TECHNOLOGIST REPORTED RECEIVING AN ELECTROSTATIC DISCHARGE (ESD) WHILE PRESSING A CONTROL PANEL BUTTON ON (B)(6) 2025. THE USER SITE REPORTS THAT NO PATIENT WAS PRESENT AT THE TIME OF THE REPORTED ESD AND THE TECHNOLOGIST DID NOT REQUIRE ANY MEDICAL INTERVENTION. A HOLOGIC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EXAMINE THE SYSTEM AND COULD NOT REPRODUCE THE ESD BUT DID FIND THAT THE FRONT TABLE TAPE STRIP WAS NOT FUNCTIONING PROPERLY. THE FSE ORDERED A REPLACEMENT CONTROL PANEL AND TAPE STRIP. A SECOND HOLOGIC FSE WAS LATER DISPATCHED TO COMPLETE THE CONTROL PANEL AND TAPE STRIP REPLACEMENT AT WHICH POINT ADDITIONAL INFORMATION WAS OBTAINED FROM THE USER SITE. THE FSE REPORTS THAT THE USER INDICATED THAT THE INTERMITTENT SHUTDOWNS ARE OCCURRING WHEN PATIENTS GET ON/OFF THE TABLE, THE SYSTEM SHUTS DOWN FOR APPROXIMATELY 5 SECONDS AND THEN AUTOMATICALLY RESTARTS WITH NO ERROR CODES DISPLAYING AND NO ENTRIES IN THE ERROR LOG. THE FSE REPLACED THE CONTROL PANEL, TESTED THE SYSTEM AND DID NOT OBSERVE ANY SHUTDOWNS OR OTHER ANOMALIES. WHILE TESTING THE SYSTEM, THE FSE ALSO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23350701 ↗InjuryNov 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON OCTOBER 12TH, IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED LOSS OF SUCTION DURING A PROCEDURE, AND MULTIPLE ATTEMPTS TO REACH OUT FOR ADDITIONAL INFORMATION WERE MADE WITHOUT SUCCESS. THE CUSTOMER REPORTED INITIALLY THAT A PATIENT SAFETY REPORT WAS BEING SUBMITTED. NO ADDITIONAL INFORMATION ON THE SAFETY REPORT COULD BE OBTAINED. HOLOGIC IS SUBMITTING THIS REPORT IN AN ABUNDANCE OF CAUTION.
Patient problems: Insufficient Information
MAUDE report 18116091 ↗MalfunctionNov 7, 2017Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT AFTER SCAN THE FILE FAILED TO OPEN, CAUSING THE PATIENT TO BE RE-EXPOSED. RESTARTING THE SYSTEM RESOLVE THE ISSUE.
Patient problems: Not Applicable
MAUDE report 7006817 ↗
Other Densitometer, bone devices
- Horizon WHologic, Inc. · Densitometer, bone1 recall1 report
- Horizon CiHologic, Inc. · Densitometer, bone1 recall1 report
- Horizon WiHologic, Inc. · Densitometer, bone1 recall0 reports
- Horizon CHologic, Inc. · Densitometer, bone1 recall0 reports
- Discovery WiHologic, Inc. · Densitometer, bone5 reports
- Prodigy F2GE Medical Systems, Ultrasound & Primary Care Diagnostics LLC · Densitometer, bone1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.