Skip to content

Horizon CRecalls & reported problems

Hologic, Inc. · The Horizon® Dual-Energy X-Ray Absorptiometry (DXA) System is a computer-assisted X-Ray scanner used to aid a physician in measuring bone mineral density (BMD), determining fracture risk, and estimating the lean body mass and fat mass of non-osseous tissues. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jun 13, 2024Class IIOpen, Classified

    Non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601-1-2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA System exceeded the limit for acceptability.

    Z-2373-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

FDA clearance and classification

Cleared through Special 510(k) K023398 on Nov 8, 2002 (decision: Substantially Equivalent). Submitted by Hologic, Inc. as “Hologic Qdr X-Ray Bone Densitometer, Models 1000, 2000, 4500, Delphi, Discovery”. The FDA review took 30 days. FDA record ↗

The FDA regulates product code KGI (densitometer, bone) under 21 CFR 892.1170 as a Class II device. 21 CFR 892.1170 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Horizon C

Has the Horizon C been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2373-2024 (Class II), began on Jun 13, 2024: Non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601-1-2 for the safety and essential performance of medical electrical equipment, where… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the Horizon C FDA approved?

It was cleared through Special 510(k) K023398 on Nov 8, 2002. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Horizon C, and is it still sold?

The FDA's device database lists Hologic, Inc. as the company responsible for it (brand Horizon C, model Horizon C). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Densitometer, bone devices

All Densitometer, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page