Brevera Breast Biopsy System BREV09Recalls & reported problems
Hologic, Inc. · Single-use, disposable needle for acquiring biopsy specimens. · Class 2
All models of Brevera Breast Biopsy System (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 15420045507777), received between Apr 3, 2020 and May 27, 2021 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K163052); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Deformation2
- Failure to Cut1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 27, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING RETRACTING THE BIOPSY NEEDLE THE SLEEVE OF THE INTRODUCER DISTORTED AND A HOLE WAS NOTED IN THE SLEEVE. NO INJURY REPORTED. THE USER SUCCEEDED IN PLACING THE NEEDLE AND THE MARKER IN THE AREA THAT WAS BIOPSIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11893407 ↗InjuryMar 30, 2021Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE EQUIPMENT AND NEEDLE WERE SET UP AND TESTED CORRECTLY. AFTER TAKING 6 TISSUE SAMPLES THERE WERE NO CALCIFICATIONS VISIBLE. AFTER TAKING 18 TISSUE SAMPLES, ONLY 8 VERY SMALL SAMPLES WERE AVAILABLE. IT WAS UNKNOWN IF THE SAMPLES WOULD BE SUFFICIENT FOR DIAGNOSIS. ADDITIONAL INFORMATION OBTAINED FROM THE CUSTOMER CONFIRMED THAT THE PATIENT WILL NEED TO UNDERGO ANOTHER BIOPSY DUE TO INSUFFICIENT TISSUE SAMPLING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11591287 ↗MalfunctionApr 27, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE ON (B)(6) 2020 THE BREVERA SHEATH BEGAN TO TWIST AND CRUMPLE WHEN THE PHYSICIAN ATTEMPTED TO REMOVE THE NEEDLE FROM THE BIOPSY SITE. THERE WAS NO IMPACT OR HARM REPORTED TO THE PATIENT. NO ADDITIONAL DETAILS AT THIS TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10002772 ↗MalfunctionApr 3, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING THE PROCEDURE, "WHEN THE NEEDLE WAS DIALED DOWN INTO THE BREAST THE SHEATH WOULD SCRUNCH UP." NO KNOWN IMPACT TO PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9924261 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K163052. One submission often covers many models, so check the product and lot details in each record.
- Z-1314-2026 · Dec 15, 2025 · Open, ClassifiedBrevera Breast Biopsy System Disposable 9 Gauge Needle. Biopsy Instrument.
Other INSTRUMENT, BIOPSY devices
- MAX-CORE MC1616Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall4,258 reports
- ENCOR® ECP017GSenorx, Inc. · INSTRUMENT, BIOPSY1 recall4,186 reports
- MAX-CORE MC1820Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall2,404 reports
- MAX-CORE MC1816Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall1 death848 reports
- MAX-CORE MC1410Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall399 reports
- MAX-CORE MC1610Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall394 reports
- ENCOR® ECP0110GVSenorx, Inc. · INSTRUMENT, BIOPSY392 reports
- ENCOR® ECP0110GSenorx, Inc. · INSTRUMENT, BIOPSY390 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.