Skip to content

Aptima® KIT, BX 1 (REF) M GEN, 100T, AS, IVDRecalls & reported problems

Hologic Ltd. · Aptima® Mycoplasma genitalium Assay, 100 test, box 1 of 2 (2°C to 8°C box). · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15420045509146), received between Jun 5, 2026 and Jun 5, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Imprecise or Erratic Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Imprecise or Erratic Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 5, 2026Imprecise or Erratic Results

    ON (B)(6) 2026, CUSTOMER REPORTED RECEIVING DISCREPANT RESULTS WHILE USING THE APTIMA MYCOPLASMA GENITALIUM ASSAY (ML 920257) ON A PATIENT SAMPLE THAT WAS PART OF THEIR BI-ANNUAL INSTRUMENT COMPARISON STUDY. ON (B)(6) 2026, CUSTOMER TESTED SAMPLES ON THE PANTHER INSTRUMENT SN (B)(6) WITH WORKLIST (WL) ID (B)(4) AND RECEIVED A NEGATIVE RESULT FOR SAMPLE ID (B)(6). ON (B)(6) 2026, CUSTOMER TESTED SAMPLES ON THE PANTHER INSTRUMENT SN (B)(6) WITH WL ID (B)() AND RECEIVED A POSITIVE RESULT FOR SAMPLE ID (B)(6). ON (B)(6) 2026 CUSTOMER RETESTED SAMPLES ID (B)(6) ON A DIFFERENT PANTHER INSTRUMENT SN (B)(6) WITH WL ID (B)(4). A POSITIVE RESULT WAS RECEIVED FOR SAMPLE ID (B)(6) AND A NEGATIVE RESULT WAS RECEIVED FOR SAMPLE ID (B)(6). ON (B)(6) 2026, CUSTOMER RETESTED BOTH SAMPLES ONE MORE TIME ON THE PANTHER INSTRUMENT SN (B)(6) WITH WL ID (B)(4) AND RECEIVED THE SAME RESULTS, POSITIVE FOR SAMPLE ID (B)(6) AND NEGATIVE FOR SAMPLE ID (B)(6).CUSTOMER NOTED THAT THERE WAS A POSSIBILITY THAT THE SAMPLE IDS WERE SWAPPED DURING THE SECOND AND THIRD RETESTS. CUSTOMER DID NOT PROVIDE ANY INFORMATION ON THE PATIENT OR PATIENT TREATMENT. HOLOGIC HAS NOT BEEN INFORMED OF ANY ADVERSE PATIENT OUTCOMES R

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25411000 ↗

Other Mycoplasma Genitalium Nucleic Acid Detection System devices

All Mycoplasma Genitalium Nucleic Acid Detection System devices →

Sources and updates