APTIMA AC2, ROOM TEMP, 250T, READY MADERecalls & reported problems
Hologic, Inc. · Class 2
All models of Aptima (15)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 15420045514966), received between Sep 3, 2024 and Mar 24, 2025 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Difficult to Open or Close” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Difficult to Open or Close1
- Imprecise or Erratic Results1
- Off-Label Use1
Patient problems
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 24, 2025Off-Label Use; Off-Label Use
CUSTOMER QUESTIONED THEIR RESULTS WITH HOLOGIC¿S APTIMA COMBO 2 READY-MADE-REAGENTS ASSAY (MASTER LOT 909330) AS THEY HAD LOWER THAN EXPECTED RLUS FOR THEIR CHLAMYDIA (CT) POSITIVE CONTROLS ACROSS AT LEAST SIX PANTHER INSTRUMENTS (SERIAL NUMBERS: (B)(6), WITH THE LAST NOT REPORTED) WHERE THEY SUSPECTED A CONTAMINATION ISSUE. CUSTOMER PERFORMED CLEANING OF TOUCHPOINT AREAS WITH 50% BLEACH SOLUTION (WHICH IS OFF-LABEL AS THE HOLOGIC PACKAGE INSERT DETAILS A CONCENTRATION OF 2.5-3.5%) AND REPORTED THAT THE ISSUE PERSISTED. CUSTOMER REPORTED OUT THE RESULTS, WITH CUSTOMER LAST REPORTING THEY HAVE NOT AMENDED ANY RESULTS. CUSTOMER HAS NOT REPORTED THAT THEY PLAN TO AMEND RESULTS, AND NO RETESTING OF SAMPLES IS KNOWN. CUSTOMER DID NOT UPLOAD COMPREHENSIVE LOG FILES. THE FOLLOWING WORKLIST INFORMATION IS KNOWN: (B)(6) ¿ 6 CT POSITIVE SAMPLES (B)(6) ¿ 8 CT POSITIVE SAMPLES (B)(6) ¿ 13 CT POSITIVE SAMPLES (B)(6) ¿ 2 CT POSITIVE SAMPLES OF THE POSITIVES, NONE HAD RLUS ABOVE 1000. TECHNICAL SUPPORT REVIEWED UPLOADED INFORMATION AND FOUND THERE WAS POSSIBLE INHIBITION, INDICATING THE POTENTIAL FOR FALSE NEGATIVE RESULTS, AND ADVISED CUSTOMER TO DISCARD OPEN ITEMS, TO CLEAN ALL SAMPLE AND REAGE
Patient problems: Insufficient Information
MAUDE report 21684459 ↗InjuryMar 7, 2025Imprecise or Erratic Results; Imprecise or Erratic Results
ON FEBRUARY 10TH, 2025, CUSTOMER REPORTED A DISCREPANT RESULT USING APTIMA COMBO 2 ASSAY (ML 908987), WORKLIST 001320-20250205-09, ON PANTHER PLUS INSTRUMENT SERIAL NUMBER (B)(6). CUSTOMER NOTED THE SAMPLE WAS INITIALLY RAN ON WORKLIST 001222-20250204-34 ON PANTHER PLUS INSTRUMENT SERIAL NUMBER (B)(6) USING THE SAME ASSAY AND OBTAINED A CT NEGATIVE AND GC EQUIVOCAL RESULT. CUSTOMER RETESTED THE SAMPLE ON WORKLIST 001320-20250205-09 (AS PREVIOUSLY MENTIONED) AND OBTAINED A CT POSITIVE AND GC NEGATIVE RESULT. CUSTOMER RETESTED THE SAMPLE FOR A SECOND TIME ON WORKLIST 011015-20250205-41 ON PANTHER PLUS INSTRUMENT SERIAL NUMBER (B)(6) USING THE SAME ASSAY AND OBTAINED A CT POSITIVE AND GC NEGATIVE RESULT. CUSTOMER NOTED THAT THEY REPORTED OUT THE INITIAL CT NEGATIVE GC EQUIVOCAL RESULT TO THE PATIENT. CUSTOMER WAS NOT PROVIDED ANY TREATMENT INFORMATION OF THE PATIENT. HOLOGIC HAS NOT BEEN INFORMED OF ANY ADVERSE PATIENT OUTCOMES RELATED TO THIS SITUATION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21553354 ↗InjuryNov 6, 2024Difficult to Open or Close; Difficult to Open or Close
ON OCTOBER 9, 2024, CUSTOMER RELAYED TO HOLOGIC THAT TECHNICIANS HAD HURT THEIR HANDS, ALLEGEDLY BY UNCAPPING OF HOLOGIC AC2 READY-MADE-REAGENT (RMR) BOTTLES. UPON FURTHER INQUIRY, HOLOGIC BECAME AWARE THAT A TECHNICIAN HAD REPORTED WRIST STRAIN ON (B)(6) 2024, AND WAS NOT ABLE TO RETURN TO FULL WORK ACTIVITY UNTIL (B)(6) 2024. AS OF (B)(6)2024, THE TECHNICIAN WAS STILL GOING THROUGH MEDICAL CHECKS EVERY TWO WEEKS.
Patient problems: Insufficient Information
MAUDE report 20621111 ↗InjurySep 3, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ON AUGUST 6, 2024, CUSTOMER RELAYED TO HOLOGIC THAT IN JUNE 2024 TWO TECHNICIANS WENT TO URGENT CARE FOR HAND/WRIST PAIN FOR THE RIGHT HAND, ALLEGEDLY CAUSED BY THE UNCAPPING OF HOLOGIC READY-MADE-REAGENT (RMR) BOTTLES. THE TECHNICIANS SUBSEQUENTLY RECEIVED PHYSICAL THERAPY.
Patient problems: Insufficient Information
MAUDE report 20132104 ↗
Other Nucleic acid detection system for non-viral microorganism(s) causing sexually transmitted infections devices
- binx io V1 (CT/NG)Binx Health Limited · Nucleic Acid Detection System For Non-Viral Microorganism(S) Causing Sexually Transmitted Infections1 recall1 report
- Alinity m 09N17-095Abbott Molecular Inc. · Nucleic acid detection system for non-viral microorganism(s) causing sexually transmitted infections1 recall1 report
- Aptima Combo 2® KIT,APTIMACOMBO2,250T,REF,CEIVD-UPDATEDHologic, Inc. · Nucleic acid detection system for non-viral microorganism(s) causing sexually transmitted infections6 reports
- BD COR CTGCTV2 US 445435Becton, Dickinson And Company · Nucleic acid detection system for non-viral microorganism(s) causing sexually transmitted infections4 reports
- Aptima® KIT, BX 1 (REF) M GEN, 100T, AS, IVDHologic Ltd. · Mycoplasma Genitalium Nucleic Acid Detection System1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page