Panther Fusion® AMR KIT, PANTHER FUSION, 96T, IVDRecalls & reported problems
Hologic, Inc. · The Panther Fusion® AdV/hMPV/RV assay is a multiplex real-time PCR (RT-PCR) in vitro diagnostic test for the rapid and qualitative detection and differentiation of Adenovirus (AdV), human Metapneumovirus (hMPV), and Rhinovirus (RV). · Class 2
All models of Panther Fusion (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15420045509290), received between Oct 22, 2019 and Feb 3, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Imprecise or Erratic Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Imprecise or Erratic Results1
- Labelling, Instructions for Use or Training Problem1
Patient problems
- Insufficient Information1
- Respiratory Tract Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 3, 2023Imprecise or Erratic Results; Imprecise or Erratic Results
HOLOGIC BECAME AWARE OF A POTENTIAL FALSE POSITIVE HMPV RESULT SUBMITTED TO THE FDA (MDR REPORT KEY (B)(4). THE INCIDENT OCCURRED ON (B)(6) 2019, BUT HOLOGIC BECAME AWARE OF THE INCIDENT ON 01/05/2023. A PATIENT WAS TESTED FOR HUMAN METAPNEUMOVIRUS (HMPV) AND RHINOVIRUS (RV) USING THE PANTHER FUSION ADV/HMPV/RV ASSAY LOT: 247667. THE SAMPLE WAS RV-POSITIVE WITH A STRONG PCR SIGNAL AND HMPV-POSITIVE WITH A WEAK PCR SIGNAL. WHEN RETESTED, THE SAMPLE TEST RESULTS WERE STILL RV- AND HMPV-POSITIVE. THE FOLLOWING WEEK, TWO OTHER PATIENTS TEST POSITIVE FOR BOTH RV AND HMPV WITH THE SAME PCR SIGNAL RESULT, LEADING THE USER TO QUESTION THE INITIAL PATIENT¿S SAMPLE RESULT AS A POTENTIAL FALSE POSITIVE. THE INFORMATION REPORTED IN THE MDR IS INSUFFICIENT TO CHARACTERIZE ANY SAMPLE AS A CONFIRMED FALSE RESULT. HOLOGIC WAS NOT CONTACTED NOR MADE AWARE OF ANY RESULTS REPORTED, ANY AMENDMENTS TO RESULTS, OR ANY TREATMENT PROVIDED. USER CONTACTED THE VENDOR, WHO STATED THEY UNDERSTOOD THE CAUSE OF THIS ISSUE AND DID NOT ISSUE GUIDANCE TO CUSTOMERS. A PREVIOUS EVENT SEARCH PERFORMED ON (B)(6) 2023 FOUND NO SIMILAR COMPLAINTS ASSOCIATED WITH LN: 247667 SINCE ON (B)(6) 2019; THE ISSUE WITH ASSAY LOT:
Patient problems: Insufficient Information
MAUDE report 16296619 ↗InjuryOct 22, 2019Labelling, Instructions for Use or Training Problem; Labelling, Instructions for Use or Training Problem
CUSTOMER REPORTED THEY HAD DUAL POSITIVE SAMPLES (UNKNOWN NUMBER OF SAMPLES) FOR RV AND HMPV WHEN RUNNING THE ADV/HMPV/RV ASSAY ON THE PANTHER-FUSION. THESE DUAL POSITIVE SAMPLES HAD EARLY RV CYCLE TIMES ASSOCIATED WITH HIGH VIRAL CONTENT AND LATE HMPV CYCLE TIMES WITH A SLOW RAMP UP TO POSITIVITY. PER THE CUSTOMER'S PROCEDURE, THEY RETESTED THE DUAL POSITIVE RESULTS USING A NEW ALIQUOT OF VIRAL MEDIUM PLACED INTO A NEW STM TUBE. ONE SAMPLE THAT WAS ORIGINALLY DUAL RV AND HMPV POSITIVE THEN HMPV RETESTED AS NEGATIVE. HOLOGIC TECHNICAL SUPPORT (TS) REVIEWED THE CUSTOMER PANTHER LOGS. HOLOGIC DETERMINED THE CUSTOMER RESULTS SHOULD NOT BE COMPARED AS THEY ARE FROM DIFFERENT ALIQUOTS. THE RESULT DISCREPANCY WAS LIKELY TO BE DUE TO A LOW TARGET. TS INFORMED CUSTOMER THAT THERE IS A RISK OF HMPV OR ADV FALSE POSITIVE RESULT FOR SAMPLES CONTAINING A HIGH RHINOVIRUS VIRAL CONTENT. A RISK ASSESSMENT WAS COMPLETED, AND THE OVERALL RISKS ASSOCIATED WITH DUAL POSITIVE RESULTS IS THAT THE PROBABILITY OF INCORRECT RESULTS IS REMOTE AND THE CONSEQUENCES ARE NEITHER SEVERE NOR LIFE-THREATENING. ALTHOUGH THE POSSIBILITY OF DUAL POSITIVE RESULTS IS PART OF HOLOGIC'S LABELING, HOLOGIC UPDATE THE LABE
Patient problems: Respiratory Tract Infection
MAUDE report 9221135 ↗
Other Respiratory virus panel nucleic acid assay system devices
- ID NOW RSV 435-000Alere Scarborough, Inc. · Respiratory Virus Panel Nucleic Acid Assay System2 recalls0 reports
- Simplexa™ Flu A/B & RSV Direct Gen II MOL2655Diasorin Molecular LLC · Respiratory Virus Panel Nucleic Acid Assay System1 recall0 reports
- Alere™ i Influenza A & B 425-024Alere Scarborough, Inc. · Respiratory Virus Panel Nucleic Acid Assay System1 death1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page