Selenia Dimensions SDM-SYS-6000-2DRecalls & reported problems
Hologic, Inc. · Selenia® Dimensions® digital mammography system is a mammographic system used for capturing digital x-ray images of the breast. · Class 3
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 15420045510753), received between Mar 8, 2022 and Jun 16, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Unintended System Motion” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (P080003); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 8, 2026Open, Classified
Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
Z-2167-2026 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Unintended System Motion4
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 16, 2025Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT DURING A SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM PROCEDURE ON (B)(6) 2025, THE USER WAS UTILIZING A SWITCH ON THE UNIT TO ROTATE THE C-ARM AND THE C-ARM CONTINUED ROTATING AFTER THE SWITCH WAS RELEASED. A FIELD ENGINEER WAS DISPATCHED TO ASSESS THE UNIT, ORDERED NEW C-ARM SWITCHES AND INSTALLED THE NEW SWITCHES ON THE SYSTEM. THE FIELD ENGINEER INDICATES THAT THE SYSTEM IS NOW WORKING IN ACCORDANCE WITH MANUFACTURER SPECIFICATIONS. NO PATIENT OR USER INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22214174 ↗MalfunctionMay 13, 2024Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT ON APRIL 16, DURING A PROCEDURE THE C-ARM MOVED INTERMITTENTLY WHEN THE SYSTEM WAS IN LATERAL POSITION WHEN THE TUBE HEAD WAS ON THE LEFT SIDE. A FIELD ENGINEER EXAMINED THE EQUIPMENT AND WAS UNABLE TO REPRODUCE THE FAILURE MODE. THE FIELD ENGINEER REPLACED THE VERTICAL DRAG BRAKE AND CONFIRMED THAT THE BACKLASH MEASUREMENTS ON THE MAIN GEAR WERE PASSING. THE SYSTEM IS FUNCTIONING PROPERLY AND MEETS ALL MANUFACTURER SPECIFICATIONS. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19305793 ↗MalfunctionFeb 15, 2024Unintended System Motion; Unintended System Motion
PENDING RESULTS OF INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18712815 ↗MalfunctionApr 4, 2023Break; Break
IT WAS REPORTED THAT DURING A PREVENTIVE MAINTENANCE FOR A SELENIA DIMENSIONS 2D ON (B)(6) 2023, A BROKEN BOLT FROM THE VTA ASSEMBLY WAS FOUND. THE FIELD ENGINEER REPLACED THE VTA ASSEMBLY PER PROCEDURES AND THE SERVICE MANUAL AND PERFORMED ALL REQUIRED CALIBRATIONS. ALL CALIBRATIONS AND TESTS PASSED AND THE SYSTEM MET THE MANUFACTURER´S SPECIFICATIONS. NO PATIENT INJURY OR STAFF REPORTED. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16674158 ↗MalfunctionMar 8, 2022Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT THE LEFT SIDE FOOT PEDAL SWITCH FOR VERTICAL TRAVEL WAS STICKING INTERMITTENTLY. NO INJURY REPORTED. A FIELD ENGINEER WAS DISPATCHED TO THE SITE AND DETERMINED THE FOOT PEDAL NEEDED TO BE REPLACED. ONCE THIS WAS COMPLETED THE SYSTEM WAS WORKING AS INTENDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13695882 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P080003. One submission often covers many models, so check the product and lot details in each record.
- Z-2168-2026 · Apr 8, 2026 · Open, Classified3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-INTL3D, 3DM-SYS-INTL3D-MOB, 3DM-SYS-INTL3D-NS, 3DM-SYS-STD, 3DM-SYS-ST
- Z-3248-2024 · Aug 19, 2024 · Open, ClassifiedSelenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version Dimensions 1.12.0 and 3Dimensions 2.3.0 system software as follows Model/Catalog Number: 1. ASY-16824
- Z-1144-2023 · Jan 16, 2023 · Open, ClassifiedHologic 3Dimensions/Selenia Dimensions Models: RM-SDM-00001-3D SDM-00001-2D SDM-00001-3D SDM-05000-3D3 SDM-05000-2DC SDM-00001-M3D SDM-05000-3DC SDM-SYS-6000-3D SDM-05000-2D3 SDA-
- Z-1343-2020 · Jan 21, 2020 · TerminatedSelenia Dimensions Mammography System, Model number ASY-04160, Selenia Dimensions Gantry, UDI 15420045505636,
Other Digital breast tomosynthesis devices
- 3Dimensions 3DM-SYS-STDHologic, Inc. · Digital breast tomosynthesis1 recall57 reports
- Selenia Dimensions SDM-SYS-6000-3DHologic, Inc. · Digital breast tomosynthesis1 recall29 reports
- Selenia Dimensions SDM-SYS-9000-3DHologic, Inc. · Digital breast tomosynthesis1 recall16 reports
- 3Dimensions 3DM-SYS-INTL2D-MOBHologic, Inc. · Digital breast tomosynthesis1 recall5 reports
- 3Dimensions 3DM-SYS-INTL3DHologic, Inc. · Digital breast tomosynthesis1 recall5 reports
- Selenia Dimensions SDA-SYS-3000-2DHologic, Inc. · Digital breast tomosynthesis1 recall4 reports
- Selenia Dimensions SDM-SYS-9000-2DHologic, Inc. · Digital breast tomosynthesis1 recall2 reports
- 3Dimensions ASY-10559Hologic, Inc. · Digital breast tomosynthesis1 recall0 reports