3Dimensions 3DM-SYS-INTL3DRecalls & reported problems
Hologic, Inc. · Selenia® Dimensions® digital mammography system is a mammographic system used for capturing digital x-ray images of the breast. · Class 3
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 15420045510579), received between Jun 24, 2024 and Jun 2, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Unintended System Motion” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (P080003); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 8, 2026Open, Classified
Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
Z-2168-2026 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Unintended System Motion5
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 2, 2025Unintended System Motion; Unintended System Motion
IT WAS REPORTED ON (B)(6) 2025, THAT THE LEFT C-ARM SIDE PANEL BUTTONS INTERMITTENTLY DO NOT WORK ON A 3DIMENSIONS MAMMOGRAPHY SYSTEM. A FIELD ENGINEER WAS DISPATCHED TO ASSESS THE EQUIPMENT AND FOUND THAT THE SWITCHES WOULD STICK, CAUSING UNCOMMANDED C-ARM MOTION. THE FIELD ENGINEER ORDERED NEW SWITCHES AND INDICATED THAT THEY WOULD BE INSTALLED WITH HOLOGIC TRAINED BIOMED UPON ARRIVAL. THE FIELD ENGINEER INDICATED THAT THE PROBLEM HAS NOW BEEN RESOLVED. NO PATIENT/USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22124320 ↗MalfunctionDec 24, 2024Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT DURING A 3DIMENSIONS PROCEDURE ON (B)(6) 2024, WHEN MOVING THE C-ARM TO 90° ANGLE, IT WAS REPORTED THAT THE C-ARM WAS RANDOMLY OVER ROTATING TO 92° WITHOUT ANY TYPE OF COMMAND. ERROR MESSAGE/CODE VTA (29:14) WAS DISPLAYED DURING THE INCIDENT (ISSUED WHEN THE MEASURED CARM/TOMO MOTION IS SLOW OR IN THE WRONG DIRECTION). THE INCIDENT OCCURRED PRIOR PROCEDURE WITH NO PATIENT IN THE ROOM. NO INJURY WAS REPORTED TO THE USER/STAFF. A FIELD ENGINEER EXAMINED THE EQUIPMENT AND IT WAS DETERMINED THAT THE ERROR COULD NOT BE REPRODUCED BUT THE ROT POTENTIOMETER WAS REPLACED AND THE BRAKES WERE ADDITIONALLY REPLACED AND THE ANTI-BACKLASH BRAKE PLATES. THE SYSTEM MET THE MANUFACTURER´S SPECIFICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21011459 ↗MalfunctionNov 4, 2024Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT ON THE (B)(6) 2024, OUR DEALER REPORTED AN INCIDENT WITH OUR 3DIMENSIONS MAMMOGRAPHY SYSTEM. IT WAS REPORTED THAT THE SYSTEM SHUT DOWN WHILE USING. DEALER ALSO REPORTED THAT SOME ERROR CODES WERE DISPLAYED. ON 10TH OF OCTOBER 2024, OUR HOLOGIC TECHNICAL SERVICE REVIEWED LOGS AND VIDEO PROVIDED BY THE DEALER. FROM THE LOGS, IT WAS CONFIRMED THAT ERROR CODE VTA 29:17 APPEARED AT THE TIME OF THE INCIDENT. THIS ERROR CODE VTA 29:17 IS TRIGGERED WHEN AN UNCOMMANDED VERTICAL MOTION OF THE C-ARM IS DETECTED. FROM A VIDEO, OUR HOLOGIC TECHNICAL SERVICE COULD CONFIRM THAT THE C-ARM SLIGHTLY MOVED DOWNWARDS, POTENTIALLY DUE TO THE BRAKE NOT ENGAGING. THE INCIDENT OCCURRED BEFORE PROCEDURE WITHOUT PATIENT IN THE ROOM. NO INJURY WAS REPORTED TO THE USER/STAFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20604353 ↗MalfunctionAug 30, 2024Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT ON AUGUST 14, WHEN THE RADIOLOGIST PRESSED THE C-ARM MOTION REQUEST OF THE SELENIA DIMENSION, THE MOVEMENT WAS SLIGHTLY IN THE OPPOSITE DIRECTION THAN WHAT WAS REQUESTED. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20112922 ↗MalfunctionJun 24, 2024Unintended System Motion; Unintended System Motion
IT WAS REPORTED ON (B)(6), THAT A 3DIMENSIONS WAS MOVING BY ITSELF DUE TO A SWITCH GETTING STUCK. THE FIELD ENGINEER EXAMINED THE DEVICE AND FOUND THAT A ROTARY SWITCH MIGHT HAVE MECHANICAL ISSUE. THE FIELD ENGINEER REPLACED THE DEFECTIVE SWITCH AND TESTED FUNCTIONALITY. THE SYSTEM IS FUNCTIONING PROPERLY AND MEETS ALL MANUFACTURER SPECIFICATIONS. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19601192 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P080003. One submission often covers many models, so check the product and lot details in each record.
- Z-3248-2024 · Aug 19, 2024 · Open, ClassifiedSelenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version Dimensions 1.12.0 and 3Dimensions 2.3.0 system software as follows Model/Catalog Number: 1. ASY-16824
- Z-1144-2023 · Jan 16, 2023 · Open, ClassifiedHologic 3Dimensions/Selenia Dimensions Models: RM-SDM-00001-3D SDM-00001-2D SDM-00001-3D SDM-05000-3D3 SDM-05000-2DC SDM-00001-M3D SDM-05000-3DC SDM-SYS-6000-3D SDM-05000-2D3 SDA-
- Z-1343-2020 · Jan 21, 2020 · TerminatedSelenia Dimensions Mammography System, Model number ASY-04160, Selenia Dimensions Gantry, UDI 15420045505636,
Other Digital breast tomosynthesis devices
- 3Dimensions 3DM-SYS-STDHologic, Inc. · Digital breast tomosynthesis1 recall57 reports
- Selenia Dimensions SDM-SYS-6000-3DHologic, Inc. · Digital breast tomosynthesis1 recall29 reports
- Selenia Dimensions SDM-SYS-9000-3DHologic, Inc. · Digital breast tomosynthesis1 recall16 reports
- 3Dimensions 3DM-SYS-INTL2D-MOBHologic, Inc. · Digital breast tomosynthesis1 recall5 reports
- Selenia Dimensions SDM-SYS-6000-2DHologic, Inc. · Digital breast tomosynthesis1 recall5 reports
- Selenia Dimensions SDA-SYS-3000-2DHologic, Inc. · Digital breast tomosynthesis1 recall4 reports
- Selenia Dimensions SDM-SYS-9000-2DHologic, Inc. · Digital breast tomosynthesis1 recall2 reports
- 3Dimensions ASY-10559Hologic, Inc. · Digital breast tomosynthesis1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page