Skip to content

Acessa ProVu System 7700Recalls & reported problems

Hologic, Inc. · The Acessa ProVu Transducer Sensor · Class 2

All models of Acessa ProVu System (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15420045515291), received between Feb 26, 2026 and May 13, 2026 — classified as 3 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • Insufficient Information2
  • Bladder Dysfunction1
  • Burn(s)1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2025, A PATIENT RECEIVED AN ACESSA PROCEDURE, THE PROCEDURE WAS REPORTED AS UNEVENTFUL AND THE SURGEON NOTICED A MARK ON THE BLADDER. AT THE TIME OF THE PROCEDURE, IT WAS REPORTED THAT A UROLOGY TEAM CONSISTING OF SURGEON, UROLOGIST AND SURGEON PROCTOR WERE PRESENT DURING THE PROCEDURE AND AGREED THAT IT WAS NOT LIKELY CAUSED BY THE ACESSA PROCEDURE. ON (B)(6) 2025, IT WAS REPORTED THAT THE PATIENT HAD REPORTED INCREASED URINARY FREQUENCY AND THAT NO FOLLOW UP TREATMENT WAS REQUIRED. ON APRIL 17TH, 2026, NEW INFORMATION WAS RECEIVED ON THE CASE. THE TREATING PHYSICIAN REPORTED THAT SHE ¨NOW BELIEVES THAT THE MARK OBSERVED HAD BEEN A THERMAL SPREAD AND THAT DURING A CYSTOSCOPY IT WAS OBSERVED SPREAD THROUGH THE BLADDER¨. PATIENT RECEIVED A FOLEY CATHETER FOR 1 WEEK AND WAS PLACED ON ANTIBIOTICS. PATIENT REQUIRED MULTIPLE VISITS FOR URINATION AND PAIN AND THEN HAD TO GO BACK TO THE UROLOGIST MONTHS AFTER FOR PAIN DURING URINATION. NO OTHER INFORMATION COULD BE OBTAINED.

    Patient problems: Bladder Dysfunction; Burn(s); Pain

    MAUDE report 25165259 ↗
  • InjuryFeb 26, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT RECEIVED AN ACESSA PROCEDURE ON (B)(6) 2026. DURING THE PROCEDURE THE CUSTOMER EXPERIENCED ISSUES WITH THE 3D GUIDANCE AND THE LINES IN THE IMAGE WERE POOR AND INTERMITTENT FOR WHICH THE ULTRASOUND HANDPIECE WAS EXCHANGED (2 ULTRASOUND DEVICES USED) BUT THIS DIDN´T FIX THE ISSUE. PHYSICIAN DECIDED TO CONTINUE WITH THE PROCEDURE WITHOUT THE 3D GUIDANCE. ADDITIONAL INFORMATION WAS RECEIVED IT WAS REPORTED THAT THE CUSTOMER HAD PREVIOUSLY DROPPED THE TABLE TOP FIELD GENERATOR (TTFG) ON A PREVIOUS PROCEDURE AND THAT THEY HAD USED IT DURING THIS PROCEDURE.THE PROCEDURE AT THE TIME HAD BEEN REPORTED TO BE COMPLETED WITHOUT PROBLEMS. THE PATIENT WAS READMITTED TO THE HOSPITAL ON A LATER DATE AND A BOWEL INJURY WAS FOUND. NO OTHER INFORMATION COULD BE OBTAINED.

    Patient problems: Insufficient Information

    MAUDE report 24461928 ↗
  • InjuryFeb 26, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT RECEIVED AN ACESSA PROCEDURE ON (B)(6) 2026. DURING THE PROCEDURE THE CUSTOMER EXPERIENCED ISSUES WITH THE 3D GUIDANCE AND THE LINES IN THE IMAGE WERE POOR AND INTERMITTENT FOR WHICH THE ULTRASOUND HANDPIECE WAS EXCHANGED (2 ULTRASOUND DEVICES USED) BUT THIS DIDN´T FIX THE ISSUE. PHYSICIAN DECIDED TO CONTINUE WITH THE PROCEDURE WITHOUT THE 3D GUIDANCE. ADDITIONAL INFORMATION WAS RECEIVED IT WAS REPORTED THAT THE CUSTOMER HAD PREVIOUSLY DROPPED THE TABLETOP FIELD GENERATOR (TTFG) ON A PREVIOUS PROCEDURE AND THAT THEY HAD USED IT DURING THIS PROCEDURE. THE PROCEDURE AT THE TIME HAD BEEN REPORTED TO BE COMPLETED WITHOUT PROBLEMS. THE PATIENT WAS READMITTED TO THE HOSPITAL ON A LATER DATE AND A BOWEL INJURY WAS FOUND. NO OTHER INFORMATION COULD BE OBTAINED.

    Patient problems: Insufficient Information

    MAUDE report 24461915 ↗

FDA clearance and classification

The FDA regulates product code HFG (coagulator, laparoscopic, unipolar (and accessories)) under 21 CFR 884.4160 as a Class II device. 21 CFR 884.4160 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Acessa ProVu System 7700

Has the Acessa ProVu System 7700 been recalled?

No FDA recall record in our data names this exact device identifier (DI 15420045515291), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Acessa ProVu System 7700?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Feb 26, 2026 and May 13, 2026. In those reports, the most common patient problems are “Bladder Dysfunction” (1), “Burn(s)” (1) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Acessa ProVu System 7700?

None of the reports are classified as deaths, 3 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Acessa ProVu System 7700, and is it still sold?

The FDA's device database lists Hologic, Inc. as the company responsible for it (brand Acessa ProVu System, model 7700). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other COAGULATOR, LAPAROSCOPIC, UNIPOLAR (AND ACCESSORIES) devices