Skip to content

Acessa ProVu System 7100 (UDI 15420045515307)Recalls & reported problems

Hologic, Inc. · The Acessa ProVu Console · Class 2

All models of Acessa ProVu System (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15420045515307), received between Dec 14, 2023 and Jan 6, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Computer Software Problem1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 6, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING AN ACESSA PROCEDURE ON (B)(6) 2025, THERE A TABLETOP GENERATOR INDICATOR (TTFG) THAT WOULD NOT STAY ON AFTER THE TTFG WAS PLUGGED INTO THE SYSTEM. THE STAFF CHECKED THE TTFG CABLE AND THE TTFG PLACEMENT AND CONFIRMED BOTH WERE CORRECT. THE SYSTEM POWER CYCLED AND AN ERROR WAS RECEIEVED WHEN IT POWERED BACK ON THAT STATED GUIDANCE FAILURE, NO TRACKER AND THERE WAS NO GREEN CHECK ON SCREEN TO INDICATE THE TTFG WAS CONNECTED. THE SYSTEM POWER CYCLED MULTIPLE TIMES AND THE TTFG CABLE WAS UNPLUGGED AND RE-PLUGGED FROM BOTH THE SYSTEM AND THE TTFG WHICH DID NOT RESOLVE. THE PHYSICIAN CHOSE TO ABORT THE PROCEDURE WHERE NO 3D GUIDANCE COULD BE USED. AFTER TESTING THE SYSTEM ON SITE WITH ADDITIONAL EQUIPMENT, IT WAS DISCOVERED THAT THE ISSUE WAS WITH THE GENERATOR. THE TESTING SUCCESSFULLY CONNECTED THE TTFG TO THE UUI SYSTEM WITH THE SAME TTFG CABLE WITH NO ERRORS AND CONNECTION ISSUES. THE SYSTEM WORKED AS INTENDED AFTER SWITCHING THE CONSOLE. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23977054 ↗
  • MalfunctionDec 29, 2025Computer Software Problem; Computer Software Problem

    IT WAS REPORTED THAT DURING AN ACESSA PROCEDURE ON (B)(6) 2025, THE PHYSICIAN REPORTED THAT AFTER 30 SECONDS INTO THE SYSTEM BEING POWERED ON, AN ERROR WAS SHOWN RELATED TO INSTRUMENT FAILURE. SYSTEM WAS POWERED CYCLED BUT THE PROCEDURE WAS ABORTED. AN INCISION WAS REPORTED AND THE HANDPIECE WAS INSIDE THE PATIENT WHEN THE ISSUE HAPPENED. NO MEDICAL INTERVENTION WAS REPORTED AND NO PATIENT INJURY REPORTED. NO OTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23907574 ↗
  • InjuryDec 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING AN ACESSA PROCEDURE ON (B)(4), 2023. THE PHYSICIAN EXPERIENCED ISSUES WITH THE FOOT PEDAL AND COULD NOT ACTIVATE THE ABLATION. DUE TO THE ISSUES EXPERIENCED WITH THE CONSOLE THE PROCEDURE HAD TO BE ABORTED, THE TROCARS HAD ALREADY BEEN PLACED. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18325871 ↗

FDA clearance and classification

The FDA regulates product code HFG (coagulator, laparoscopic, unipolar (and accessories)) under 21 CFR 884.4160 as a Class II device. 21 CFR 884.4160 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Acessa ProVu System 7100

Has the Acessa ProVu System 7100 been recalled?

No FDA recall record in our data names this exact device identifier (DI 15420045515307), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Acessa ProVu System 7100?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Dec 14, 2023 and Jan 6, 2026. In those reports, the most common specific device problem is “Computer Software Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Acessa ProVu System 7100?

None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Acessa ProVu System 7100, and is it still sold?

The FDA's device database lists Hologic, Inc. as the company responsible for it (brand Acessa ProVu System, model 7100). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other COAGULATOR, LAPAROSCOPIC, UNIPOLAR (AND ACCESSORIES) devices

All COAGULATOR, LAPAROSCOPIC, UNIPOLAR (AND ACCESSORIES) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page